Teledyne 9060 Operating Instructions Manual page 84

Zirconia oxygen analyzer
Table of Contents

Advertisement

ZIRCONIA OXYGEN SENSOR OUTPUT (mV)
PROBE TYPE 1231, SENSOR TYPE 1234
% OXYGEN
___________________________________________________
21.0
20.5
20.0
19.5
19.0
18.5
18.0
17.5
17.0
16.5
16.0
15.5
15.0
14.5
14.0
13.5
13.0
12.5
12.0
11.5
11.0
10.5
10.0
9.5
9.0
8.5
8.0
7.5
7.0
6.5
6.0
5.5
5.0
4.5
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
0.2
_________________________________
'K' TC mV
These tables are based on the Nernst equation:
Sensor e.m.f. = 0.02154 x T x 1n x 20.95 / % oxygen, where T = ° K (° C + 273), e.m.f. is in mV's
mV at 720°C (1320°F)
0.00
0.46
0.99
1.53
2.09
2.66
3.25
3.85
4.47
5.11
5.77
6.45
7.15
7.87
8.62
9.40
10.21
11.05
11.92
12.83
13.78
14.78
15.82
16.92
18.08
19.30
20.60
21.98
23.45
25.04
26.75
28.61
30.65
32.90
35.42
38.28
41.58
45.48
50.25
56.41
65.08
79.91
99.51
29.212 at 720°C (1320°F)