Teledyne 9060 Operating Instructions Manual page 7

Zirconia oxygen analyzer
Table of Contents

Advertisement

SPECIFICATIONS
Inputs
• Zirconia oxygen probe, heated or unheated
• Furnace, kiln or flue thermocouple, field selectable as type K or R.
• Main flame established safety interlock (for heated probes only)
• Purge pressure switch
• Dual Fuel selector
Remote alarm accept
Outputs
• Two linearised 4 to 20 mA DC outputs, max. load 1000Ω
• Common alarm relay
Three other alarm relays with selectable functions
Computer
• RS 232-C or RS 485 for connection of a computer terminal or printer for diagnostics of
the transmitter, probe, sensor or combustion process. This connection is suitable for
network connection to computers, DCSs or PLCs using
Range of Output 1
Field selectable from the following:
Output Selection
Linear, Probe 1
Linear, Probe 1 and 2 averaged 0 to 1% oxygen to 0 to 100 % oxygen
(If 2 probes are used)
Log
Reducing
Reducing
Linear, probe 1, very low range 0 to 0.001% to 0 to 2.0 % oxygen (10ppm to 20,000ppm)
Range of Output 2
Field selectable from the following:
Output
Sensor EMF
Carbon Dioxide
Oxygen Deficiency
Aux Temperature
Log Oxygen
Reducing Oxygen
Linear Oxygen, probe 2
Combustibles %, Probe 1
Linear, Probe 1 and 2 averaged
(If 2 probes are used)
7
Range
0 to 1% oxygen to 0 to 100 % oxygen
0.1 to 20 % oxygen, fixed
100 % to 10 -4 oxygen, fixed
10 -1 to 10 -25 % oxygen, fixed
Zero Range
0 to 1100 mV in 100 mV steps
0 to 10 %
0 to 20% O 2 deficiency
0 to 100°C (32 to 210°F) in 1 degree steps
0.1% O 2 Fixed
10 +2 (100%) to 10 -10 % oxygen in one
decade steps, non-overlapping
0% oxygen, fixed
0% combustibles fixed
0% oxygen, fixed
7
9060-Zirconia Analyser
MODBUS protocol.
Span Range
1000 to 1300 mV in 100
mV steps
2 to 20 %
0 to 100% O 2 excess
100 to 1400°C (210 to
2550°F) in 100 degree
steps
20% O 2 Fixed
10 -3 to 10 -30 % oxygen
in one decade steps. Min
1 to 100%
0.5 to 2.0 %
1 to 100%
span two decades.