Configuration and Setup Addresses
Holding
Reg.
Function
716
Probe #1 offset
717
Probe #2 offset
Purge control related variables
754
Purge enable
Calibration checking gas related variables
759
Gas calibration check
204
Probe #1 EMF
8
205
Probe #2 EMF
0
205
Probe #1 OXYGEN
2
205
Probe #2 OXYGEN
4
205
Probe #1, Impedance
6
205
Probe #2, Impedance
8
206
Probe #1 TC mV
0
206
Probe #2 TC mV
2
206
Probe #1 temperature
4
206
Probe #2 temperature
6
206
ALRM-ARRAY
8
208
ALRM-TIMES
4
Alarm array order -
1. Heater 1 fail
2. Sensor1 fail (Impedance too high)
3. Probe 1 filter blocked
4. Probe 1 thermocouple open circuit
5. Reference air Pump fail
6. Battery backed RAM fail
7. Mains frequency measurement fail
8. ADC warning (outside normal specifications, but still accurate)
9. DAC warning (outside normal specifications, but still accurate)
9060-Zirconia Analyser
Description
10 = 1.0mV
10 = 1.0mV
0= off, 1= on
0= off, 1= 1 gas, 2= 2 gasses
100,000 = 100.000 mV
100,000 = 100.000 mV
100,000,000 = 100.0%
100,000,000 = 100.0%
1,000 = 1 k Ω
1,000 = 1 k Ω
100,000 = 100.000 mV
100,000 = 100.000 mV
700 = 700 degC
700 = 700 degC
Array of current alarm status. See below
Array of timestamp of alarms