The fuse block, at the right of the power cord receptacle, accepts US
or European size fuses. A jumper replaces the fuse in whichever fuse
receptacle is not used.
3.3.2.3 50-P
IN
Figure 3-2 shows the pin layout of the Equipment Interface
connector. The arrangement is shown as seen when the viewer faces the
rear panel of the analyzer. The pin numbers for each input/output
function are given where each function is described in the paragraphs
below.

Figure 3-2: Equipment Interface Connector Pin Arrangement

3.3.2.4 A
NALOG
There are four DC output signal pins—two pins per output. For
polarity, see Table 3-1. The outputs are:
0–1 VDC % of Range:
0–1 VDC Range ID:
4–20 mA DC % Range: Current increases linearly with
4–20 mA dc Range ID:
Teledyne Analytical Instruments
30
E
I
QUIPMENT
NTERFACE
O
UTPUT
Voltage rises linearly with increasing
concentration, from 0 V at 0
concentration to 1 V at full scale. (Full
scale = 100% of programmable range.)
0.20 V = Low Range
0.5 V = Medium Range
0.80 V = High Range
concentration, from 4 mA at 0
concentration to 20 mA at full scale.
(Full scale = 100% of programmable
range.)
6.8 mA = Range 1
12.0 mA = Range 2
16.8 mA = Range 3
Series 4060