Series 4060
2. The MAIN MENU and SUB MENU screens have a flashing
arrow on the left side indicating which item is being pointed
to.
3. MODIFIABLE ITEM screens have a flashing value
somewhere in the line.
4.6.4.1 M
M
AIN
The MAIN MENU consists of 15 functions you can use to customize
and check the operation of the analyzer. They are listed here with brief
descriptions:
Set up and/or start a span calibration
SPAN:
ALT-SPAN
MODEL:
version of the instrument.
SELF-TEST:
routine to check the integrity of the power supply, output
boards, cell and amplifiers.
ALARMS:
whether each alarm will be active or defeated, HI or LO
acting, and failsafe or not.
RANGE:
switched manually.
 ANALOG-OUT ADJUST: Adjust the analog output with
offset and scaling.
LINEARIZATION
CHANGE STREAM
switch for using alternate gases. Has no effect during span
mode (span gas relay setting used at that time).
AUTO-CAL:
calibration sequence. AUTOCAL is an optional feature not
included in the standard configurations of the 4060.
: Low level application specific timing parameters.
TIMING
Factory set. Do not change these.
Teledyne Analytical Instruments
ENUS
: Set up and/or start an alternate span calibration.
Displays Manufacturer, Model, and Software
The instrument performs a self-diagnostic
Used to set the alarm setpoints and determine
Used to set up three analysis ranges that can be
: Linearization option.
: (optional feature) Controls a relay
Used to define and/or start an automatic
Operation
49