Operation
This menu allows the operator to set the time the analyzer should
AUTO
be held in the
span mode, after the readings of the analyzer settle.
Five minutes is the default, but it could be adjusted anywhere from 1 to
60 minutes by using the ▲▼ keys.
Press Enter to move to the next screen.
Use the ▲▼ arrow keys to enter the oxygen-concentration mode.
Use the ◄► arrow keys to blink the digit you are going to modify. Use
the ▲▼ arrow keys again to change the value of the selected digit.
When you have finished typing in the concentration of the span gas you
are using, press Enter to begin the Span calibration.
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates
information on Slope. Spanning automatically ends when the span
output corresponds, within tolerance, to the value of the span gas
concentration. Then the instrument automatically returns to the Analyze
mode.
4.4.2.2 M
M
ANUAL
ODE
Press Span to start the Span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually.
Use the ▲▼ keys to toggle between
MAN
Stop when
appears blinking on the display. Press Enter to move to
the next screen.
Press Enter to move to the next screen.
This menu allows the operator to set the time the analyzer should
be held in the AUTO span mode. It does not have any effect in the
MANual mode. Just press Enter key to continue.
Teledyne Analytical Instruments
Span Val: 000008.00
<ENT>Span <UP>Mod #
####
ppm Span
Slope=####
S
PANNING
Span: Settling:MAN
<ENT> For Next
AUTO
Calib. Holding time
Cal hold: 5 min
3000TA-XL-EU
ppm/s
MAN
and
span settling.
52