Appendix
3000TA-XL-EU
Power: Universal power supply 85-250 VAC, at
47-63 Hz.
Operating Temperature: 5-35°C (41-95°F)
Accuracy: ±2% of full scale at constant temperature
on all ranges except 0-1 ppm.
±5% of full scale over operating
temperature range, except 0-10 ppm and
0-1 ppm analysis range, once thermal
equilibrium is reached.
±0.1 ppm on 0-1 ppm analysis range, once
thermal equilibrium is reached.
Analog outputs: 0-1 VDC percent-of-range,
0-1 VDC range ID.
4-20 mA DC (isolated) percent-of-range,
4-20 mA DC (isolated) range ID.
Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep
(6.96 in high, 8.7 in wide, 12.2 in deep).