Cell Failure; Span Cal; Auto Mode Spanning - Teledyne 3000TA-XL-EU Operating Instructions Manual

Trace oxygen analyzer
Hide thumbs Also See for 3000TA-XL-EU:
Table of Contents

Advertisement

Trace Oxygen Analyzer
4.4.1.3 C
F
ELL
AILURE
Cell failure in the 3000TA-XL-EU is usually associated with
inability to zero the instrument down to a satisfactorily low ppm reading.
When this occurs, the instrument will still Zero, but the 3000TA-XL-EU
system alarm trips, and the LCD displays a failure message.
Before replacing the cell:
1. Check your span gas to make sure it is within specifications.
2. Check for leaks up-stream from the cell, where oxygen may
be leaking into the system.
If there are no leaks and the span gas is within specification,
replace the cell as described in Chapter 6, Maintenance.

4.4.2 Span Cal

The Span button on the front panel is used to span calibrate the
analyzer. Span calibration can be performed using the automatic mode,
where an internal algorithm compares consecutive readings from the
sensor to determine when the output matches the span gas concentration.
Span calibration can also be performed in manual mode, where the
operator determines when the span concentration reading is acceptable
and manually exits the function.
4.4.2.1 A
M
S
UTO
ODE
Press Span to enter the span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually. Use the ▲▼ arrow keys to toggle between
AUTO
MAN
and
span settling. Stop when
display.
Press Enter to move to the next screen.
Teledyne Analytical Instruments
#.#
ppm Anlz
CELL FAIL/ ZERO HIGH
PANNING
AUTO
Span: Settling: AUTO
<ENT> For Next
Calib. Holding time
Cal hold: 5 min
Operation
appears blinking on the
51