Manual Mode Zeroing; Cell Failure - Teledyne 3000TB Instruction Manual

Trace oxygen analyzer
Hide thumbs Also See for 3000TB:
Table of Contents

Advertisement

Trace Oxygen Analyzer
The zeroing process will automatically conclude when the output is
within the acceptable range for a good zero. Then the analyzer automatically
returns to the Analyze mode.
Press Zero to enter the Zero function. The screen that appears allows
you to select between automatic or manual zero calibration. Use the Δ∇ keys
to toggle between AUTO and MAN zero settling. Stop when MAN appears,
blinking, on the display.
Press Enter to begin the zero calibration. After a few seconds the first
of five zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate. It calculates the differences between successive samplings and
displays the rate of change as Slope= a value in parts per million per second
(ppm/s).
NOTE:It takes several seconds for the true Slope value to display.
Wait about 10 seconds. Then, wait until Slope is sufficiently
close to zero before pressing Enter to finish zeroing .
Generally, you have a good zero when Slope is less than 0.05 ppm/s
for about 30 seconds. When Slope is close enough to zero, press Enter. In a
few seconds, the screen will update.
Once span settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the Analyze
mode.
Cell failure in the 3000TB is usually associated with inability to zero
the instrument down to a satisfactorily low ppm reading. When this occurs,
the 3000TB system alarm trips, and the LCD displays a failure message.
####
4 Left=###
4.4.1.2

Manual Mode Zeroing

Zero: Settling: Man
<ENT> To Begin
####
ppm Zero
Slope=####
4.4.1.3

Cell Failure

#.#
ppm Anlz
Teledyne Analytical Instruments
PPM
Zero
ppm/s
ppm/s
Operation 4
4-13

Advertisement

Table of Contents
loading

Table of Contents