Teledyne 3000TB Instruction Manual page 39

Trace oxygen analyzer
Hide thumbs Also See for 3000TB:
Table of Contents

Advertisement

Trace Oxygen Analyzer
2. Hold the fitting body steady with a backup wrench, and with
SAMPLE IN: In the standard model, gas connections are made at the
SAMPLE IN
Tee'd into the Sample inlet with appropriate valves.
The gas pressure in should be reasonably regulated. Pressures between
3 and 40 psig are acceptable as long as the pressure, once established, will
keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4
SLPM). Exact figures will depend on your process.
If greater flow is required for improved response time, install a bypass
in the sampling system upstream of the analyzer input.
Note: If the unit is for vacuum service, the above numbers apply
instead to the vacuum at the EXHAUST OUT connector, de-
scribed below, with minus signs before the pressure readings.
ZERO IN and SPAN IN: These are additional ports for inputting span
gas and zero gas. There are electrically operated valves inside for automatic
switching between sample and calibration gases. These valves are com-
pletely under control of the 3000TB Electronics. They can be externally
controlled only indirectly through the Remote Cal Inputs, described below.
Pressure, flow, and safety considerations are the same as prescribed for
the SAMPLE IN inlet, above.
another wrench rotate the nut another 1
Figure 3-10: Gas Connector Panel
and
EXHAUST OUT
Teledyne Analytical Instruments
1
/
turns.
4
connections. Calibration gases must be
Installation 3
3-13