4 Operation
Use < > arrows to blink
Span/Zero set appears.
Press < > arrows to blink
won't be able to set OFF to ON if a zero interval is entered.) A Span
Every ... (or Zero Every ...) screen appears.
Use Δ Δ Δ Δ Δ ∇ arrows to set an interval value, then use < > arrows to move to
the start-time value. Use Δ Δ Δ Δ Δ ∇ arrows to set a start-time value.
To turn ON the Span and/or Zero cycles (to activate Auto-Cal): Press
System again, choose
Zero values screen appears, use the < > arrows to blink the Span (or Zero)
OFF/ON field. Use Δ Δ Δ Δ Δ ∇ arrows to set the OFF/ON field to ON. You can
now turn these fields ON because there is a nonzero span interval defined.

4.3.3 Password Protection

If a password is assigned, then setting the following system parameters
can be done only after the password is entered: span and zero settings,
alarm setpoints, analysis range definitions, switching between autoranging
and manual override, setting up an auto-cal, and assigning a new password.
However, the instrument can still be used for analysis or for initiating a self-
test without entering the password.
If you have decided not to employ password security, use the default
password TETAI. This password will be displayed automatically by the
microprocessor. The operator just presses the Enter key to be allowed total
access to the instrument's features.
Note: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
4.3.3.1
To install a new password or change a previously installed password,
you must key in and ENTER the old password first. If the default password
is in effect, pressing the ENTER button will enter the default TETAI pass-
word for you.
4-6
, and press Enter. A new screen for
Auto—Cal
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Span
(or
Zero
Span Every 0 d
Start 0 h from now
and press Enter again. When the Span/
Auto—Cal,

Entering the Password

Teledyne Analytical Instruments
), then press Enter again. (You
Model 3000TB