Trace Oxygen Analyzer
4.1
Once the analyzer has been installed, it can be configured for your
application. To do this you will:
Before you configure your 3000TB these default values are in effect:
Ranges: LO = 100 ppm, MED = 1000 ppm, HI = 10,000 ppm.
Auto Ranging: ON
Alarm Relays: Defeated, 1000 ppm, HI, Not failsafe, Not latching.
Zero: Auto, every 0 days at 0 hours.
Span: Auto, at 000008.00 ppm, every 0 days at 0 hours.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.

Operation

Introduction

Set system parameters:
Establish a security password, if desired, requiring Operator
to log in.
Establish and start an automatic calibration cycle, if desired.
Calibrate the instrument.
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
Teledyne Analytical Instruments
Operation 4
4-1