3 Installation
EXHAUST OUT: Exhaust connections must be consistent with the
hazard level of the constituent gases. Check Local, State, and Federal laws,
and ensure that the exhaust stream vents to an appropriately controlled area if
required.
Note: If the unit is for vacuum service, see Sample In, above, for gas
pressure/flow considerations.
3.6

Testing the System

Before plugging the instrument into the power source:
Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.
Check that the voltage selector switch on the Electrical
Connector Panel is in the appropriate position for your power
source.
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test as described in chapter 4, section
4.3.5.
3-14
Teledyne Analytical Instruments
Model 3000TB