IMPORTANT
1. Do not turn off the power to the transmitter
immediately after completion of the change in
the LRV (and/or URV) setting(s). Note that
powering off within thirty seconds after setting
will ca use a return to the previous settings.
2. Changing LRV automatically changes URV to
the following value.
URV
previous URV
3. If the range-setting push-button and external
zero-adjustment screw are not touched during
a range-change operation, the transmitter
automatically switches back to the normal
operation mode.
Integral indicator
Note: Use a thin bar which has a blunt tip, e.g.,
a hexagonal wrench, to press the range-
setting push-button.
Figure 1.2.1 Range-setting Switch
(new LRV
previous LRV)
Range-setting switch
(Push-button)
F0106.EPS