Thermal Conductivity Analyzer
3. Replace fuse by reversing process in step 1.
4.4
System Self Diagnostic Test
1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to
3. Use the < > arrow keys to move to
4. Observe the error-code readings on the VFD Display screen, and
check Table 4-1, below, to interpret the codes.
Table 4-1: Self Test Failure Codes
Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK - > 0 means gains of the amplifier have a
high offset, e.g. 204 means the highest
gain has a high offset it is common on
sealed air reference.
Teledyne Analytical Instruments
Part II: Analysis Unit
, and press Enter .
, and press Enter .
Part II: 4-3