Stepping and Ramping the Output Value

Two features are available for adjusting the value of
source functions, except pressure, which requires
that you use an external pressure source:
Stepping the output manually with the
keys, or in automatic mode.
Ramping the output with optional continuity or V
trip detect.

Using Manual Step

The manual
feature allows you to select a step
Step
size in engineering units (mV, V, mA, °C, etc.) or %
of scale. Stepping the output in % of scale is useful
for quickly jumping between 0% and 100% (set step
size = 100%) or 0-50-100% (set step size = 50%).
Stepping works in SOURCE and in
MEASURE/SOURCE modes. Proceed as follows to
select a step size:
58
1. Refer to the appropriate "Using Source Mode"
subheading earlier in this manual (e.g.,
"Sourcing Electrical Parameters") and connect
the calibrator to the circuit to be tested.
2. If necessary, press
3. Set the calibrator for the desired source value.
4. If you want to step the source value in % of
scale, set the % of scale value as described
and
previously under "Sourcing in Percent of Scale."
5. Press the
6. Use the numeric keypad to enter the step size in
the units shown on the display.
7. Press the
8. Now you can adjust the output in steps by
pressing the
for SOURCE mode.
softkey.
Step
softkey.
Done
and
keys.