Substitute Leakage Current Test (I
Operational Error Measurement Range ... 0.25 to 19.00 mA
Operational Error ......................................................... 10%
Accuracy ................................................. ± (5% + 5 counts)
Display Range ......................................... 0 to 19.99 mA ac
Resolution.............................................................. 0.01 mA
Test Voltage................................................. 35 V ac ± 20%
(at nominal mains input voltage)
Touch Current Test (I
TOUCH
Operational Error Measurement Range ....... 0.1 to 1.99 mA
Operational Error ........................................................ 6.0%
Accuracy ................................................. ± (4% + 2 counts)
Display Range ........................................... 0 to 1.99 mA ac
Resolution.............................................................. 0.01 mA
Internal Resistance (via probe) ....................................2 kΩ
Measuring method .....................................................Probe
The appliance under test is energized at mains potential.
32
)
SUB
)

Load/ Leakage Test: Load Current

Display Range ......................................................0 to 13 A
Accuracy ................................................. ± (4% + 2 counts)
Resolution...................................................................0.1 A
The appliance under test is energized at mains potential.

Load/Leakage Test: Load Power

Display Range ................................................. 0 to 999 VA
Accuracy ................................................. ± (5% + 3 counts)
Resolution ............................................ 1 VA (0 to 999 VA)
The appliance under test is energized at mains potential.
Load/Leakage Test: Leakage Current (I
Operational Error Measurement Range ... 0.25 to 19.00 mA
Operational error....................................................... 12.0%
Accuracy ................................................. ± (4% + 5 counts)
Display Range ......................................... 0.25 to 19.99 mA
Resolution..............................................................0.01 mA
The appliance under test is energized at mains potential.
1.0 kVA to 3.2 kVA
0.1 kVA (1.0 kVA to 3.2 kVA)
)