2-Ohm Output Current Test - Fluke 150 Series Calibration Manual

Insulation testers
Hide thumbs Also See for 150 Series:
Table of Contents

Advertisement

150X Series
Calibration Manual
Step
1.
2.
3.
4.

2-Ohm Output Current Test

26
Downloaded from
Elcodis.com
electronic components distributor
Table 8. Earth Bond Resistance Tests
Meter Range
2000 Ω
2000 Ω
20.00 KΩ
Remove calibrator and apply a separate 2 Ω resistor for the following step.
20.00 Ω
The following procedure verifies the minimum current for a 2-Ω Load in continuity
function.
1. Connect a 2-Ω resistor and DMM to the UUT as shown in Figure 9.
2. Set the DMM rotary function switch to the mVDC function.
3. Set the UUT rotary function switch to the Ω function.
4. Press the T key and record the DMM voltage reading.
5. Using Ohms Law (V = I x R where: V = Voltage, I = Current, and R =
Resistance), calculate the current through the 2-Ω resistor.
6. The current should be greater the 200 mA.
1507
INSULATION TESTER
Figure 9. 2-Ω Output Current Test
Applied
Display Units
810.0 Ω
990.0 Ω
18.0 kΩ
kΩ
2.0 Ω
I = E
R
2
Display Reading
Low Limit
High Limit
795
825
972
1008
17.70
18.30
1.94
2.06
mV
ecz14f.eps