Calibrating a Pressure Transmitter
The following example shows how to calibrate a pressure
transmitter.
Connect the calibrator to the instrument under test as
shown in Figure 20. Proceed as follows:
1.
Press
for current (upper display). If required,
press
again to activate loop power.
2.
Press
(lower display).
3.
If necessary, press
4.
Zero the pressure module.
5.
Perform checks at 0 % and 100 % of span and adjust
the transmitter as necessary.
for SOURCE mode.
Calibrating a Pressure Transmitter