1.
Connect a pressure module to the calibrator as
shown in Figure 18. The threads on the pressure
modules accept standard ¼ NPT pipe fittings. Use the
supplied ¼ NPT to ¼ ISO adapter if necessary.
2.
Press
(lower display). The calibrator
automatically senses which pressure module is
attached and sets its range accordingly.
3.
Zero the pressure module as described in the
module's Instruction Sheet. Modules vary in zeroing
procedures depending on module type.
4.
Pressurize the pressure line with the pressure source
to the desired level as shown on the display.
If desired, continue pressing
display units to psi, mmHg, inHg, cmH
cmH
O@20 C, inH
O@4 C, inH
2
2
2
bar, kg/cm
, or kPa.
36
to change pressure
O@4 C,
2
O@20 C, mbar,
2
725
MULTIFUNCTION CALIBRATOR
V mA
LOOP
V mA
MEAS
SOURCE
TC
RTD
STORE
SETUP
RECALL
Figure 18. Connections for Sourcing Pressure
ZERO
Hz
˚
C
˚
F
100%
25%
25%
0%
sh19f.eps