Measuring Temperature; Using Thermocouples - Fluke 725 User Manual

Multifunction process calibrator
Hide thumbs Also See for 725:
Table of Contents

Advertisement

725
Users Manual

Measuring Temperature

Using Thermocouples

The calibrator supports ten standard thermocouples,
including type E, N, J, K, T, B, R, S, L, or U. Table 5
summarizes the ranges and characteristics of the
supported thermocouples.
To measure temperature using a thermocouple, proceed
as follows:
1.
Attach the thermocouple leads to the appropriate TC
miniplug, then to the TC input/output as shown in
Figure 10. One pin is wider than the other. Do not try
to force a miniplug in the wrong polarization.
20
If the calibrator and the thermocouple plug are at
different temperatures, wait one minute or more
for the connector temperature to stabilize after
you plug the miniplug into the TC input/output.
2.
If necessary, press
3.
Press
for the TC display. If desired, continue
pressing this key to select the desired thermocouple
type.
If necessary, you can toggle between °C or °F
temperature units by pressing
Note
for MEASURE mode.
.

Hide quick links:

Advertisement

Table of Contents
loading