Calibrating a Transmitter

Use the measurement (upper display) and source (lower
display) modes to calibrate a transmitter. The following
example shows how to calibrate a temperature
transmitter.
Connect the calibrator to the instrument under test as
shown in Figure 15. Proceed as follows to calibrate a
transmitter:
K
1.
Press
for current measurement with loop power.
2.
Press
(lower display). If desired, continue
pressing this key to select the desired thermocouple
type.
3.
If necessary, press
32
for SOURCE mode.
4.
Set your zero and span parameters by pressing
and
keys. Enter these parameters by pressing and
holding
and
setting parameters, see "Setting 0 % and 100 %"
earlier in this manual.
5.
Perform test checks at 0-25-50-75-100 % points by
pressing
or
necessary.
. For more information on
. Adjust the transmitter as