724
Users Manual
Using Source Mode
In SOURCE mode, the calibrator generates calibrated
signals for testing and calibrating process instruments,
supplies voltages and resistances, and simulates the
electrical output of RTD and thermocouple temperature
sensors.
Sourcing Electrical Parameters
Volts or ohms are sourced and shown in the lower
display.
V
+
–
26
Red
Black
Figure 12. Electrical Sourcing Connections
To select an electrical sourcing function, proceed as
follows:
1.
Connect the test leads as shown in Figure 12,
depending on the source function.
2.
If necessary, press
V
3.
Press
for dc voltage, or
4.
Enter the desired output value by pressing
keys. Press
and
change.
30V MAX ALL TERMINALS
SOURCE / MEASURE
V
TC
3W
RTD
4W
COM
for SOURCE mode.
U
for resistance.
and
to select a different digit to
MEASURE
V
mA
LOOP
COM
Common
zi09f.eps