724
Users Manual
Testing an Output Device
Use the source functions to test and calibrate actuators,
recording, and indicating devices. Proceed as follows:
1.
Connect the test leads to the instrument under test
as shown in Figure 16.
2.
Press
for dc voltage, or
(lower display).
3.
If necessary, press
34
for resistance
for SOURCE mode.
Red
0 to 1 V dc
Input
Black
Figure 16. Calibrating a Chart Recorder
724
TEMPERATURE CALIBRATOR
V
mA
LOOP
V
˚
C
MEAS
SOURCE
˚
F
TC
RTD
STORE
100%
SETUP
25%
RECALL
25%
0%
zi13f.eps