Hanna Instruments HI 9043 Instruction Manual

Hanna Instruments HI 9043 Instruction Manual

Portable microprocessor k-type thermocouple thermometers

Advertisement

Quick Links

Instruction Manual
HI 9043 • HI 9044
Portable Microprocessor
K-Type Thermocouple
Thermometers
w w w . h a n n a i n s t . c o m

WARRANTY

All Hanna Instruments meters are warranted for two years against de-
fects in workmanship and materials when used for their intended purpose
and maintained according to instructions. Sensors and probes are war-
ranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed mainte-
nance are not covered.
If service is required, contact the dealer from whom you purchased the instru-
ment. If under warranty, report the model number, date of purchase, serial
number and the nature of the failure.
First obtain a Returned Goods Authorization number from the Customer Ser-
vice department, then return the instrument with the Authorization # included
along with shipment costs prepaid. If the repair is not covered by the war-
ranty, you will be notified of the charges.
When shipping any instrument, make sure it is properly packaged for com-
plete protection.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
Dear Customer,
Thank you for choosing a Hanna product.
This manual will provide you with the necessary information
for correct operation. Please read it carefully before using the
meter.
If you need additional technical information, do not hesitate
to e-mail us at tech@hannainst.com.
These instruments are in compliance with the
directives.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and exam-
ine it carefully to make sure that no damage has occurred
during shipping. If there is any noticeable damage, notify
your Dealer or the nearest Hanna Office.
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any defective item must
be returned in the original packaging together with
the supplied accessories.

GENERAL DESCRIPTION

HI 9043 and HI 9044 are microprocessor-based thermometers
which allow temperature measurements using K-type thermo-
couple probes. The non-linearity of the temperature probe is
linearized by the microprocessor.
Standard features include dual-level LCD for simultaneously
displaying of the highest and lowest measured temperatures,
display hold, selectable measurement unit (°C or °F) and resolution
which automatically switches from 0.1° to 1° for temperatures
above 200°C (400°F).
The meters are also provided with low battery detection and
BEPS (Battery Error Preventing System), which turns the unit
off when the battery is discharged avoiding erroneous read-
ings caused by low battery level.
• HI 9043 is supplied complete with a 9V battery and
instruction manual.
• HI 9044 is supplied complete with HI 766E2 fixed
temperature probe, a 9V battery and instruction manual.
FUNCTIONAL DESCRIPTION
1. Liquid Crystal Display
2. Measuring unit selection key
3. HI/LO values reset key
4. HOLD key
5. ON/OFF key
1) Current temperature value
2) HOLD indicator
3) Low Battery indicator
4) Minimum temperature value
5) Maximum temperature value
6) K-type probe indicator
7) Measuring unit, °C or °F
SPECIFICATIONS
Range (*)
-50.0 to 1350°C
-58.0 to 2462°F
Resolution
0.1°C (up to 199.9°C) / 1°C (outside)
0.1°F (up to 399.9°F) / 1°F (outside)
Accuracy
±0.2% F.S.
for 1 year, excluding probe error
Typical EMC
±3°C / ±6°F
Deviation
with HI 766 K-Thermocouple probe
Battery Type
1 x 9V (IEC 6LR61) battery
Life
approx. 500 hours of continuous use
Probe
HI 766 K-Thermocouple series
• optional for HI 9043
• HI 766E2 fixed for HI 9044
Environment
-10 to 50ºC (14 to 122ºF);
RH max 95% non-condensing
Dimensions
180 x 83 x 40 mm (7.1x3.3x1.6")
Weight
226 g (8 oz.)
(*) Range may be limited by probe. The range of HI 9044
is limited to 900°C/1650°F by the fixed probe.

FACTORY RECALIBRATION

All Hanna thermometers have been accurately pre-calibrated
at the factory. It is generally recommended to have all
thermometers recalibrated at least once a year.
For an accurate annual recalibration, contact your dealer or
the nearest Hanna Service Center.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 9043 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI 9043

  • Page 1 WARRANTY GENERAL DESCRIPTION FACTORY RECALIBRATION All Hanna Instruments meters are warranted for two years against de- HI 9043 and HI 9044 are microprocessor-based thermometers All Hanna thermometers have been accurately pre-calibrated fects in workmanship and materials when used for their intended purpose which allow temperature measurements using K-type thermo- at the factory.
  • Page 2 HIGH/LOW TEMPERATURES OPERATIONAL GUIDE CE DECLARATION OF CONFORMITY ACCESSORIES The maximum and minimum tempera- tures are continuously monitored and INITIAL PREPARATION K-TYPE THERMOCOUPLE PROBES displayed on the lower portion of the with integral handle, 1 m (3.3’) cable & mini-connector: Remove the battery cover on the back of the thermometer.