Hanna Instruments HI935005 Instruction Manual

Hanna Instruments HI935005 Instruction Manual

Portable k-thermocouple thermometers
Hide thumbs Also See for HI935005:

Advertisement

Instruction Manual
HI935005 - HI935005N
HI935002 - HI935009
Portable
K-Thermocouple
Thermometers
www.hannainst.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI935005 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI935005

  • Page 1 Instruction Manual HI935005 - HI935005N HI935002 - HI935009 Portable K-Thermocouple Thermometers www.hannainst.com...
  • Page 2: Table Of Contents

    OPERATIONAL GUIDE ..............8 LCD, low battery detection, long battery life and two-year warranty. USER CALIBRATION ..............12 HI935005 has been designed for laboratory and field use, while FACTORY RECALIBRATION ............12 HI935005N is suitable for high accuracy (meter and probe can be BATTERY REPLACEMENT ..............13...
  • Page 3: Specifications Of Hi935005 And Hi935005N

    • HI and LO (Max and Min) temperature values always displayed on the Calibration mode. the LCD • Remaining battery life indication / low battery detection • Waterproof casing • Auto-off capability (HI935005 only) • Backlight feature (HI935005N only) • User calibration at 0 °C (HI935005N only)
  • Page 4: Specifications Of Hi935002 And Hi935009

    SPECIFICATIONS OF HI935002 & HI935009 HI935002 HI935009 SPECIFICATIONS -50.0 to 199.9 °C / 200 to 1350 °C Range (*) -58.0 to 399.9 °F / 400 to 2462 °F 0.1 °C (up to 199.9 °C) / 1 °C (outside) Resolution 0.1 °F (up to 399.9 °F) / 1 °F (outside) Accuracy ±0.2% full scale (@20 °C/68 °F)
  • Page 5: Operational Guide

    The measured temperature is held on the display until HOLD is pressed To access this feature, the HI935005 is provided with an internal button again. The "HOLD" tag blinks on the display while in HOLD mode. located in the battery compartment (see figure on page 10). Unscrew the 4 screws on the back of the meter and select the desired auto-off time period by pressing the button and viewing the value on the LCD.
  • Page 6 HI935009 allows the user to select the auto-off time period through 2-CHANNEL MODELS (HI935002 and HI935009) the front keyboard; press (ALT+) AOFF to enter the mode, then set HI935002 and HI935009 models can monitor two samples through the desired auto shut-off interval (8 min, 60 min, or disabled) with two independent temperature channels (probes).
  • Page 7: User Calibration

    User calibration is only performed on the current channel displayed (T1 or T2). FACTORY RECALIBRATION All Hanna Instruments thermometers have been accurately pre-calibrated at the factory. It is generally recommended to have all thermometers recalibrated at least once a year. For an accurate recalibration, contact your local...
  • Page 8: Accessories

    HI766E2 General purpose probe, max 900 °C/1650 °F If service is required, please contact your local Hanna Instruments Office. HI766F High temperature, flexible wire probe without handle, max If under warranty, report the model number, date of purchase, serial 1100 °C/2000 °F...
  • Page 9 World Headquarters Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com Local Office Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Phone: 800.426.6287 Fax: 401.765.7575 e-mail: tech@hannainst.com MANK199 07/18...

This manual is also suitable for:

Hi 935002Hi 935005nHi 935009

Table of Contents