Questions and answers

Angela Joyner
February 3, 2025

We purchased two HI93501 thermometers to use for checking our pasteurizer temperatures. The calibration certificate provided used temperatures of 32 and 122F but the temperature range we used goes up to 160F or greater. Is there an option to get a calibration certificate at a higher temperature and is the thermometer accurate at these higher temperatures?

1 comments:
Mr. Anderson
February 10, 2025

The Hanna Instruments HI93501 thermometer has a temperature range of -50.0 to 150.0 °C (-58.0 to 302.0 °F) and includes a Calibration Check feature. However, there is no mention of user calibration for higher temperatures. It has an accuracy of ±0.4 °C (±0.7 °F), which applies across its measurement range, including temperatures above 122 °F.

This answer is automatically generated

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI93501

  • Page 1 Instruction Manual HI93501 HI935001 HI935007 HI935004 HI935008 Portable Microprocessor Thermometers w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid.
  • Page 3: Preliminary Examination

    HI93501 is a NTC10K-thermistor thermometer. It offers a wide temperature range with a resolution of 0.1 °C (0.1 °F). HI935001 and HI935007 is a K-thermocouple thermometer. It offers a wide temperature range, automatic change of resolution from 0.1 °C (0.1 °F) to 1 °C.
  • Page 4: Specifications Of Hi93501

    SPECIFICATIONS OF HI93501 Main features: • FC762PW NTC1OK-thermistor probes series • Range -50.0 to 150.0 °C (-58.0 to 302.0 °F) • Calibration Check feature • Remaining battery life indication / low battery detection • Auto-off capability • Waterproof casing Range -50.0 to 150.0 °C...
  • Page 5: Specifications Of Hi935001 & Hi935007

    SPECIFICATIONS OF HI935001 AND HI935007 Main features: • FC766PW K-thermocouple probe for HI935001 • Fixed K-thermocouple probe for HI935007 • Range -50.0 to 1350.0 °C (-58.0 to 2462.0 °F) • Remaining battery life indication / low battery detection • Auto-off capability •...
  • Page 6: Specifications Of Hi935004 & Hi935008

    SPECIFICATIONS OF HI935004 AND HI935008 Main features: • FC767PW T-thermocouple probe for HI935004 • Fixed T-thermocouple probe for HI935008 • Range -50.0 to 400.0 °C (-58.0 to 752.0 °F) • Remaining battery life indication / low battery detection • Auto-off capability •...
  • Page 7: Operational Guide

    At start-up the thermometer displays all LCD segments for a few seconds, followed by the percentage indication of the remaining battery life; then it enters Cal Check mode and then enters measure- ment mode. HI93501 HI935001 and HI935007 HI935004 and HI935008...
  • Page 8 Note: The full display can be frozen by keeping the ON/OFF button pressed. If a temperature probe is plugged in, the meter displays the measured temperature. If no temperature probe is plugged in or the reading is over-range, “NO PROBE” message is the display shows blinking dashes and displayed (see WARNING MESSAGES section) .
  • Page 9 If the meter is calibrated, the display will stabilize and show “0.0 °C” (±0.3 °C) in a few seconds. WARNING MESSAGES NO PROBE message If the probe is not connected or is interrupted and “NO PROBE” message is displayed. HI93501 HI935001 and HI935007 HI935004 and HI935008...
  • Page 10 CALIBRATION CHECK ERROR MESSAGE If during CALIBRATION CHECK the meter display more than ±0.3 °C, then a “Err” message is dispayed . Contact your dealer or your nearest Hanna Service Center for recalibration. DAMAGED PROBE message When the readings are very far from the reading range, the probe can be damaged, “NO PROBE”...
  • Page 11: Meter Setup

    METER SETUP Enter Setup to change the tempera- ture unit ( °C or °F) or to adjust the auto shut off time. To enter the Setup MODE, press the SET button located inside the battery compartment. • Use the ON/OFF button to switch between “°C”...
  • Page 12: Battery Replacement

    BATTERY REPLACEMENT When the remaining battery life is less than 10% the battery tag blinks on the display to warn the user. If the battery level is 0% enough to cause erroneous readings, the Battery Error Prevention System (BEPS) turns the meter off. Immediately replace the batteries with new ones.
  • Page 13: Temperature Probes

    Hanna probes are also available in different handle colors to avoid cross contamination during testing. Note: To clean the meters, do not use aggressive detergents. It is recommended to use water. FOR HI93501 FC762PW (*) General purpose penetration probe w/1 m (3.3') white cable & white handle...
  • Page 14 24 VAC or 60 VDC. To avoid damage or burns, do not perform any measurement in microwave ovens. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.
  • Page 16 Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.com Local Sales and Customer Service Office Printed in ROMANIA MAN93501Y 12/16...