Hanna Instruments HI 93510N Instruction Manual

Hanna Instruments HI 93510N Instruction Manual

Portable microprocessor thermistor thermometers

Advertisement

Instruction Manual
HI 93510 - HI 93510N
HI 93512 - HI 93522
Portable Microprocessor
Thermistor Thermometers
w w w . h a n n a i n s t . c o m

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 93510N and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 93510N

  • Page 1 Instruction Manual HI 93510 - HI 93510N HI 93512 - HI 93522 Portable Microprocessor Thermistor Thermometers w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    • 3 x 1.5V AA alkaline batteries WARRANTY • Instruction manual. All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained Note: Save all packing material until you are sure that the according to instructions.
  • Page 3: Specifications Of Hi 93510 And Hi 93510N

    SPECIFICATIONS OF HI 93510 & HI 93510N HI 93510 HI 93510N SPECIFICATIONS Range -50.0 to 150.0°C -58.0 to 302.0°F Resolution 0.1°C 0.1°F (-58.0 to 230.0°F) / 0.2°F (outside) Accuracy ±0.4°C / ±0.8°F for one year, excluding probe error Typical EMC ±0.5°C / ±1°F Deviation with HI 762BL probe...
  • Page 4: Specifications Of Hi 93512 And Hi 93522

    SPECIFICATIONS OF HI 93512 & HI 93522 HI 93512 HI 93522 SPECIFICATIONS Range -50.0 to 150.0°C -58.0 to 302.0°F Resolution 0.1°C 0.1°F (-58.0 to 230.0°F) / 0.2°F (outside) Accuracy ±0.4°C / ±0.8°F for one year, excluding probe error Typical EMC ±0.5°C / ±1°F Deviation with HI 762BL probe...
  • Page 5: Operational Guide

    Note: Although the display is frozen, internally the meter continues OPERATIONAL GUIDE measuring and updating Hi/Lo and relative values. In HI 93522 only, the held value is also stored in non-volatile To switch the instrument ON, press the ON/OFF key. memory and can be recalled by pressing (ALT+) MR key.
  • Page 6 HI 93522 allows the user to select the auto-off time period through 2-CHANNEL MODELS (HI 93512 and HI 93522) the keyboard; press (ALT+) AOFF to enter the mode, then set the HI 93512 and HI 93522 models can monitor two samples through desired auto shut-off interval (8 min, 60 min, or disabled) with the two independent temperature channels (probes).
  • Page 7: User Calibration

    USER CALIBRATION BATTERY REPLACEMENT The HI 93510N and HI 93522 models can be calibrated at 0°C by When the battery level is below 10%, a warning using an ice bath. symbol will blink on the LCD to indicate a low •...
  • Page 8: Ce Declaration Of Conformity

    HI 762W Wire probe without handle for hard-to-reach places w/1 m (3.3') cable Hanna Instruments reserves the right to modify the design, con- HI 762W/10 Wire probe, without handle for hard-to-reach places struction and appearance of its products without advance notice.
  • Page 9 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel.