Recommended Test Equipment; Table 7-1 Recommended Test Equipment - Agilent Technologies U1401B User's Manual And Service Manual

Handheld multi-function calibrator/meter
Hide thumbs Also See for U1401B:
Table of Contents

Advertisement

Recommended Test Equipment

Table 7-1 Recommended test equipment

Application
DC Voltage
DC Current
Resistance
AC Voltage
AC Current
Frequency
Temperature
Duty Cycle
Diode
U1401B User's and Service Guide
The test equipment recommended for the performance
verification and adjustment procedures are listed below. If
the exact instrument is not available, substitute with another
calibration standard of equivalent accuracy.
Recommended equipment
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Agilent 33250A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Performance Tests and Calibration
Recommended accuracy
requirements
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
7
131

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents