Table 7-5
Calibration tests
Step
Test function
b.
Turn the rotary switch to mV.
Press
to select AC.
Turn the rotary switch to
. Press
AC.
U1401B User's and Service Guide
50 mV @ 1 kHz
500 mV @ 1 kHz
5 V @ 1 kHz
to select
50 V @ 1 kHz
500 V @ 1 kHz
750 V @ 1 kHz
Performance Tests and Calibration
Cal range
5 mV
45 mV
50 mV
450 mV
0.5 V
4.5 V
5 V
45 V
25 V
250 V
75 V
750 V
Input
U1401B
5 mV
5.000 mV
45 mV
45.000 mV
50 mV
50.00 mV
450 mV
450.00 mV
0.5 V
0.5000 V
4.5 V
4.5000 V
5 V
5.000 V
45 V
45.000 V
25 V
25.00 V
250 V
250.00 V
75 V
75.0 V
750 V
750.0 V
7