Agilent Technologies U1401B User's Manual And Service Manual page 55

Handheld multi-function calibrator/meter
Hide thumbs Also See for U1401B:
Table of Contents

Advertisement

U1401B User's and Service Guide
Defining the autoscan parameters in the memory
Press and hold MODE for more than one second to enter the
autoscan adjustment mode. A total of 16 steps with
individually definable time interval and amplitude are
available.
When the instrument is in the autoscan adjustment mode,
the secondary display shows the amplitude. The first two
digits of the primary display are used to indicate which step
is being adjusted. The last two digits of the primary display
are used to indicate time interval.
1 Press MODE to cycle through step, time interval, and
amplitude adjustments. The digit to be adjusted will flash
on the display.
• For amplitude adjustment, press
the digit to be adjusted, and then press
adjust the value of the selected digit. The amplitude
can be set to any value within the selected output
range (±1.5 V or ±15 V for constant voltage output, ±25
mA for constant current output).
• For time interval adjustment, press
select the digit to be adjusted, and then press
to adjust the value of the selected digit. The time
interval can be set within the range of 0 to 99 seconds.
• Press
for more than one second to directly reset
the time interval and amplitude of the present step to
zero.
Calibrator Output Operations
and
to select
and
and
2
to
to
and
35