Agilent Technologies U1401B User's Manual And Service Manual page 156

Handheld multi-function calibrator/meter
Hide thumbs Also See for U1401B:
Table of Contents

Advertisement

7
Performance Tests and Calibration
Table 7-3
Input performance verification tests (continued)
Step
Test Function
8
Turn the rotary switch to
.
9
Turn the rotary switch to mV.
Press and hold
more than 1 second.
136
Connection to 5520A
Connect a diode to the
U1401B input terminals in
forward bias position.
Connect K-type
for
thermocouple to the
U1401B input terminals.
Range
5520A output
2 V
1.9 V
–40 °C to
0 °C
1372 °C
–40 °F to
32 °F
2502 °F
U1401B User's and Service Guide
Error from
nominal 1
year
±1.45 mV
±3 °C
±6.096 °F