VIBRATION CONTROLLER/MONITOR Model : PVB-8219 Your purchase of this VIBRATION CONTROLLER/ MONITOR marks a step forward for you into the field of precision measurement. Although this vibration controller is a complex and delicate instrument, its durable structure developed. Please read the following instructions carefully and always keep this manual within easy reach.
Caution Symbol Caution : * Risk of electric shock ! Caution : * Do not use fingers or any tool to touch the Wire Terminals. * Do not apply the relay contact load current > 0.5 Amp. * The instrument contains no user serviceable parts and should not be opened by the user.
1. FEATURES * Applications for industrial vibration monitoring : All industrial machinery vibrates. The level of vibration is a useful guide to machine condition. Poor balance,misalignment & looseness of the structure will cause the vibration level increase, it is a sure sign that the * Frequency range 10 Hz - 1 kHz, sensitivity relative meet ISO 2954.
2. SPECIFICATIONS 2-1 General Specifications Circuit Custom one-chip of microprocessor LSI circuit. Display 4 digits red LED, digit size : 14 mm. Measurement Velocity, Acceleration, Displacement Function Acceleration, Velocity : RMS Displacement : p-p ( peak-peak ) Display Unit Measurement Metric Imperial Acceleration...
Page 6
Relay outputs Number 2 relays Function Relay 1 : Control relay. Relay 2 : High/Low alarm relay. Max load 0.5 ACA/250 ACV 0.5 DCA/24 DCV * Do not apply the relay contact load current > 0.5 A, other wise the relay may be damaged permanently without warranty.
4. MEASURING PROCEDURE 4-1 Indicator Description A. Metric units are red Indicators: 1) Acceleration: m/s² 2) Velocity: mm/s 3) Displacement: mm B. Inch units are yellow. Indicators 1) Acceleration: ft/s² 2) Velocity: inch/s 3) Displacement: inch C. Setting Indicators: 1) PROCESS VALUE(PV) 2) Lit during normal test will be light.
4-3 Terminal connection 1) Input the ACV power ( 90 to 260 ACV ) to T1, T2. Do not input the over voltage to the AC input terminals. 2) Connect the " Alarm Relay " output from T3, T4. Connect the " Control Relay " output from T5, T6. 4-4 VIBRATION Measurement A.
B. Unit select Press and hold the SET Button(3-4, Fig. 1) > 2 sec. into setting then short press SET Button (3-4, Fig. 1) 5 times,The screen will stay in Unit use " ▲(3-6,Fig.1)"、"▼(3-5,Fig.1)" Button select Metric or Imperial then short press "SET Button " once, wii be save select unit, short press "SET Button "...
Page 13
Under Display show " LoLt ", if " Displacement indicator ( 3-12, Fig. 1 ) " is lit, meter is ready for " Displacement Low Limit " setting. The function of " Low Limit value " setting, refer to page 13, Fig. 3. B .
Under Display show " ALLo ", if " Acceleration indicator " ( 3-10, Fig. 1 ) is lit, meter is ready for " Acceleration Low Limit " setting. Under Display show " ALLo ", if " Velocity indicator " ( 3-11, Fig. 1 ) is lit, meter is ready for "...
Page 15
A. Digital Filter Value 1) Press the " Set Button " ( 3-4, Fig. 1 ) continuously at least two seconds, the " Display " will show " FiLt ", now the meter is ready for the Digital Filter Value Setting. 2) Press the "...
Page 16
CtHy CtHy CtHy CtHy CtLo CtHI CtLo CtHI CtLo CtHI Fig.3 For example : Control High limit value : 39.0 Control Low limit value : 10.0 Control Hysteresis value : 0.5 a. When measurement value ≧39.0 The Control relay will On ,when measuring value ≦...
Page 17
ALHy ALHy ALHy ALHy ALLo ALHI ALLo ALHI ALLo ALHI Fig.4 For example : Alarm High limit value : 10.0 Alarm Low limit value : 2.0 Alarm Hysteresis value : 0.5 a. When measurement value ≧10.0 The alarml relay will On ,when measuring value ≦...
2) In " Gain " screen, short press "Function Button"(3-7, Fig. 1) to select Acceleration or Velocity or Displacement Function . 3) Press the " ▲(3-6,Fig.1)"or"▼(3-5,Fig.1)" Button once, display will show digital value ,then use " ▲(3-6,Fig.1)"and"▼(3-5,Fig.1)" Button to adjustment Gain Value, then short press the "SET Button "...
Page 19
The 16 digits data stream will be displayed in the following format : D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the following status : Start Word When send the upper display data = 1 When send the lower display data = 2 D12 &...
RS232 setting Baud rate 9600 Parity No parity Data bit no. 8 Data bits Stop bit 1 Stop bit 6. SYSTEM RESET Power on the meter, use the two fingers to press " Set Button " ( 3-4, Fig. 1 ) and " ▼ Button " ( 3-5, Fig. 1 ) continuously more than 2 seconds until the Display show the text "...