Summary of Contents for Woodpecker Mini Ray Dental X-Ray Generator
Page 1
Mini Ray 牙科X射线机英文说明书V1.0(CE 130X190mm)231220 14.02.21.010.indd 2 Mini Ray 牙科X射线机英文说明书V1.0(CE 130X190mm)231220 14.02.21.010.indd 2 2024/1/4 17:24:13 2024/1/4 17:24:13...
Page 2
Preface Thank you for purchasing the Dental X-ray Generator produced by Guilin Woodpecker Medical Instrument Co., Ltd. Woodpecker is a high-tech enter- prise researching, developing, producing, and dental products sales company. Please read the full text of the instruction manual carefully to ensure that you can use the device correctly and safely Mini Ray 牙科X射线机英文说明书V1.0(CE 130X190mm)231220 14.02.21.010.indd 3...
1 Product introduction 1.1 Product introduction This device is a portable dental X-ray generator used to photograph teeth and obtain dental image information. Features of this device: 1) Small, light, easy for doctors to carry. 2) High quality and efficient user interface, making exposure easier. 3) Low radiation and high efficiency, providing good user experience.
Page 5
longs to internal power supply ME device. 3. Degree of protection against electric shock: The Beam limiter is B Type applied part. 4. Degree of protection against harmful ingress of water: Ordinary device (IPX0) 5. Degree of safety application in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide: Device can’t be used in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
16. Reference LOADING conditions: 70kV, 2mA, 2s, loading interval 1s/60s, 59 exposures per hour. 17. Radiation output stated: the variation coefficient under the function less than 5%. 18. Under the reference condition (70kV, 2mA, 2s) of this device, the leak- age radiation value of the accessible surface (excluding the X-ray outlet) is not exceed 0.05mGy/h.
Page 7
Woodpecker acces- sories and operated according to the operating instructions.Please do not attempt to disassemble or modify the device.
5) Please contact the manufacturer to check the device, and do not attempt to disassemble it for repair. 6) Please use an image receiver that meets the Woodpecker operational re- quirements. If an image receiver with a low resolution or does not meet the relevant requirements of local medical devices is used, the image quality may be affected, resulting in blurred images, affecting clinical judgment.
Page 9
Please use the provided power adapt- er to charge the device. Using other non-matching power sources may lead to device damage due to differences in voltage. And do not position me device to make it difficult to plug off the adapter Power adapter Using a Connection cable of hand brake and hand brake, the device can be con-...
Page 10
Note: The Hand brake interface can only be connected to the hand brake using a hand brake cable and cannot be connected to other USB devices. Figure 1 Schematic diagram of Dental X-Ray Generator When remotely controlling the device’s exposure via a hand brake cable, a fixed bracket can be used to secure the X-ray device.
Page 11
should be greater than 3KG. 4.2.1 Installation area Take out all the parts from the packing box. Be careful not to drop or dam- age the device. 4.2.2 Power adapter installation Take out the power adapter and power connector from the packing box and connect them as shown in the figure.
Page 12
Figure 4 Exposure hand brake connecting cable Connect the other end of the hand brake connecting cable to the host: Figure 5 Exposure handbrake connection device Note: ① wired connection mode: Long press the switch of exposure hand brake, the exposure starts light on. 4.2.5 Functions of the control panel See Table 1 for the functions of the icons on the control panel.
Page 13
Lock indication. Press and hold the “▲” and “▲” buttons at the same time to unlock/lock the generator. The generator cannot be operated in the locked state. Time setting of X-ray exposure, “▼” decreases time of exposure , “▲” increases time of exposure When you click the Calibration button, the displayed Angle re- turns to zero.
not be less than 200cm. 5 Operating instructions The user of the medical generator must comply with the requirements of the relevant operating regulations and relevant regulations of the medical de- partment, and is limited to the use of trained doctors or technicians using in hospital environment.
Page 15
values are as follows (The following values are only reference values, the operator can adjust the angle according to the actual situation to ensure that the beam limiting tube and the image receiving device remain verti- cal.): Table 3 Tooth position X - ray tilt direction Angle of tilt Maxillary incisor position...
Page 16
Figure 7 Figure 8 2) Human Body Mode After selecting the generator mode, select the human mode. First, click the human body selection icon indicated by the arrow at the first position in Figure 9, and button back and forth between adult and child modes at the second position.
Page 17
sponding to the tooth position. Figure 10 5.4.2 Setting of exposure time If there is required to change the exposure time, Click the “▲” or “▼” but- ton and the exposure time can be adjusted from 0.02 seconds to 2 seconds. 5.4.3 Exposure location icon When exposure, the operator should hold the generator, aim the output port of the beam limiting cylinder at the part of the patient that needs to shoot...
Page 18
Country Brand Model IP image plates France ACTEON IMAGING PLATE Dental image plates scanner China Woodpecker i-Scan Digital intraoral X-ray imag- China Woodpecker i-Sensor H2 ing system 5.5 Using and Maintaining The Battery 5.5.1 Using The Battery (1) The battery level indicator for remaining power is displayed on the right side of the control panel.
Page 19
Figure 12 (2) Charging a. Connect one end of the charger to the charging port of the device and the other end to the power supply of the network (100-240V, 50/60Hz). b. When charging, the device displays the charging icon, and when charg- ing is finished, the battery is charged fully.
Page 20
saving mode is activated and the generator automatically powers off. All displays are turned off on the control panel and the generator must be pow- ered back on to resume normal operation in power saving mode. 5.5.2 Battery Maintenance 1) When the generator is not in use, the power button should be turned off to save electrical energy.
Table 8 Page number reference in Instruc- Testing information tions for use Image receiving generator requirment P.15 Radiation and Protection P.27 6 Cleaning and Maintenance [Note]Maintenance and servicing of the device is prohibited during use. The power adapter, power cord, exposure handbrake and handbrake connect- ing wire, and battery of this device are replaceable parts.
Page 22
Maximum anode heat dissipation: 110W Overall dimension and wiring: as shown in Figure 13 Maximum rated value: as shown in Figure 14 Thermal characteristics: see Figure 15 Filament and emission characteristics: see Figure 16 Figure 13 Mechanical dimension generator wiring Figure 14 Maximum rating diagram Mini Ray 牙科X射线机英文说明书V1.0(CE 130X190mm)231220 14.02.21.010.indd 19 Mini Ray 牙科X射线机英文说明书V1.0(CE 130X190mm)231220 14.02.21.010.indd 19...
Figure 15 X-ray tube anode heating and cooling curve Figure 16 Filament and emission characteristic curve 8 Storage, maintenance, and transportation 8.1 Storage/ maintenance 1) The device shall be handled with care, away from an uneven or vibrat- ing source, and shall be installed or stored in a cool, dry and ventilated place.
10% -93%, atmospheric pressure of 70kPa ~ 106kPa and temperature of -20℃ ~ + 55℃. Inspect the device for scratches, wear and other mechanical scratches or damage after each use. 8.2 Transportation 1) Avoid excessive shock and vibration during transportation, and handle with care to avoid inversion.
ranty period and scope. This product does not contain self-maintained parts, and the maintenance of this device should be carried out by designated pro- fessionals or special repair shops. At the same time, the maintenance of the circuit diagram, parts list and other information are provided by the company to help maintenance personnel to repair parts;...
Page 26
RF emission Group B CISPR 11 The dental X-ray generator is suitable for used in Harmonic emission all establishments, including domestic establish- Group A lEC 61000-3-2 ments and establishments directly connected to the public low-voltage power supply network that Voltage fluctuation/ supplies buildings used for domestic purposes.
Page 27
Mains power quali- <5% U <5% U ty should be that of a (>95% dip in U (>95% dip in U typical commercial or Voltage dips, for 0.5 cycle for 0.5 cycle hospital environment. short interrup- <5%U <5% U If the user of the dental tions and voltage (>95% dip in U (>95% dip in U...
Page 28
2:2014+A1:2020) 2:2014+A1:2020) LEVEL in V/m. Notes: Without the explicit consent of Woodpecker, unauthorized changes or modi- fications to the device may cause electromagnetic compatibility problems of this device or other device. 11.4 Recommended separation distances between RF communications equipment and dental imaging plate scanner.
Separation distance according to frequency of transmitter/ m Rated maximum output power of 150kHz ~ 80MHz 80MHz ~ 800MHz 800MHz ~ 2.5GHz transmitter /W d =1.2×P d =1.2×P d=2.3×P 0.01 0.12 0.12 0.23 0.38 0.38 0.73 For the rated maximum output power of transmitters not listed in the above table, the recommended separation distance “d”...
WEEE mark Please deal with the waste disposal Temperature limit for produced by the device storage: -20°C ~ +55°C following relevant laws and regulations. Humidity limit for stor- Electrostatic Sensitive age: 10% ~ 93% Atmospheric pressure for Danger! High voltage storage: 70kPa ~ 106kPa Avoid sun exposure Date of manufacture...
of local medical devices are used, the image quality may be affected, re- sulting in blurred images, etc., affecting clinical judgment. 13.6 If taken from one room to another, it is necessary to hold the equipment with both hands to prevent it from falling. 14 Radiation and Protection 14.1 Radiation 1) dose area product...
Page 32
the focal point of the X-ray tube is 24cm, and the dose area product tost is- carried out. When choosing the corresponding exposure time, the deviaa- tion of the measured dose area product does not exceed 50% of the valuin the table.
Figure 18 Stray radiation distribution diagram When measuring the scatter radiation, the focal point is 1.15 meter above the ground and 0.1 meter outside the middle of limited occupied area. 2) Physical protection During operation, it is recommended to wear protective clothing and pro- tective glasses to reduce radiation hazards.
Page 34
Mini Ray 牙科X射线机英文说明书V1.0(CE 130X190mm)231220 14.02.21.010.indd 31 Mini Ray 牙科X射线机英文说明书V1.0(CE 130X190mm)231220 14.02.21.010.indd 31 2024/1/4 17:24:15 2024/1/4 17:24:15...