3. Replace the black glass with the Instrument Standard (White Tile) and press MEASURE.
4. Replace the Instrument Standard and place the Diagnostics Check Tile (Green Tile) and
Press MEASURE. The Diagnostics Check Tile values were entered at the factory and are
also listed on the back of the tile.
Optional: To skip the Diagnostics Check Tile test, tap SKIP CHECK TILE. Skipping may impact
instrument performance, and a warning message will be displayed.
Note: If the Diagnostics Check Tile fails, clean the instrument White Reference Tile,
Check Tile, and/or Black Glass, and run the test again. Contact HunterLab Support
Reading a Sample
To measure a sample:
1. Place the sample on the port .
2. Press MEASURE. The first reading will be treated as a standard since the 'CIELAB
[D65/10]' default WorkSpace uses the Ad hoc/working standard type.
3. To set another sample as the standard in this job, measure the sample, tap its name,
and select SET AS STANDARD.
4. In the MEASUREMENT DETAILS dialog, click on the associated line to edit the SAMPLE
NAME, PRODUCT ID, and ENTER NOTES. Use the DELETE to remove a sample
measurement.
14
User's Manual for ColorFlex L2 and EasyMatch Essentials V. 1.0
if the issue persists.
Figure 2. Set a Sample as Standard