OBJECT SCAN 4 How to use RAYSCAN Object Scan
5
6
7
8
On the Object CT modality screen, select
the desired mode (protocol). When
scanning
calibration data.
Caution: When scanning impression, if
impression material and calibration data
do not match, there may be a problem in
STL converting.
Press the [ready] button on the console
PC or touch monitor to rotate the device
to the position for acquisition. Selecting
the [cancel] button cancels acquisition
and moves to the standby screen.
When the exposure switch is lit green,
press the button until the acquisition is
complete.
Note: Push the button until you hear a
beep. If you release the button in the
middle of exposure, the acquisition might
stop. In case of an emergency, take your
hand off the exposure switch.
When exposure is finished, select either
[Confirm/Retake/Reject] buttons.
Confirm: Save the image and move to
the acquisition list screen and standby.
Retake: Automatically moves to the
patient information screen and resumes.
Reject: Saves images with delection
information, moves to shooting list
screen, and waits.