Instructions for use SONICflex quick 2008 LS - 1.007.0540 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Technical condition......................7 2.3 Accessories and combination with other equipment ............8 2.4 Qualification of personnel....................8 2.5 Service and repair......................
Sirona is a registered brand of Dentsply Sirona GmbH, Bensheim. All other trademarks are property of their respective owners. KaVo Original Factory Repair In the event of a repair, please ship your product to the KaVo Original Factory Repair using www.kavobox.com. KaVo Technical service...
Page 5
Instructions for use SONICflex quick 2008 LS - 1.007.0540 1 User instructions Serial number Manufacturer Manufacturing date Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code UDI symbol CE mark for medical devices Medical device, labelling of medical devices...
Instructions for use SONICflex quick 2008 LS - 1.007.0540 1 User instructions Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material damage. The warning notes are designated as shown below: HAZARD In cases which –...
4 Place the torque wrench on the tip when the handpiece is in its holder. 2.2 Technical condition A damaged product or damaged or NOT KaVo original components could injure patients, users or third parties. 4 Only operate devices or components if they show no signs of damage on the outside.
Repairs, servicing and safety checks may only be performed by trained service personnel. The following persons are authorised to do this: ▪ Service technicians of KaVo branches after the appropriate product training ▪ Service technicians of KaVo authorised dealers after the appropriate product...
Instructions for use SONICflex quick 2008 LS - 1.007.0540 2 Safety | 2.6 Protective equipment As a result of the use of NON-KaVo original spare parts during the repair, parts may become detached and injure patients, users or third parties. This may res- ult in aspiration, swallowing of parts and possibly even a risk of suffocation.
The SONICflex is a dental handpiece in accordance with DIN EN ISO 18397. The vibrations are generated by a rotating steel sleeve. In combination with the different KaVo tips, an oscillating elliptical tip motion is generated for the re- spective application. The internal water cooling (spray cooling) prevents the treatment site from heating up, and keeps the treatment surface clean.
NOTE KaVo recommends the start in the maximum setting of the foot control. The intensity can then be regulated with the foot control. 3.3 Technical Specifications Drive pressure 2.5 - 3.5 bar (36-51 psi)
Page 12
Instructions for use SONICflex quick 2008 LS - 1.007.0540 3 Description of the product | 3.4 Transportation and storage conditions Relative humidity: 5% RH to 85% RH absence of condensation Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Dispose of and recycle the packaging appropriately in accordance with current packaging law, employing waste management / recycling companies. Comply with the comprehensive return system. KaVo has had its packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
Instructions for use SONICflex quick 2008 LS - 1.007.0540 4 Startup and shut down | 4.3 Check the O-rings of the Sirona coupling 4 Insert the Sirona Prüfstern pressure gauge between the Sirona quick coup- ling and the medical device and check according to technical specifications in chapter 3.3.
Instructions for use SONICflex quick 2008 LS - 1.007.0540 5 Operation | 5.1 Attaching the medical device 5 Operation 5.1 Attaching the medical device CAUTION Inexact coupling. Reduces the service life of the lamp. 4 Avoid inexact coupling. 4 Check the secure fit of the (LUX) handpieces on the coupling by pulling on them.
Instructions for use SONICflex quick 2008 LS - 1.007.0540 5 Operation | 5.4 Removing the SONICflex tip 4 Use the thin region of the handle ① for insertion. 4 Use the thick region of the handle ② for tightening or detachment.
Page 17
Instructions for use SONICflex quick 2008 LS - 1.007.0540 5 Operation | 5.5 Setting the power on the SONICflex 4 Set the power level on the control ring of the SONICflex. 17 / 28...
Malfunction and material damage. 4 Reprocess it in a washer disinfector only. KaVo recommends the following products based on the compatibility of the ma- terials. The microbiological efficacy must be ensured by the disinfectant manu- facturer and proven by an expert opinion.
Instructions for use SONICflex quick 2008 LS - 1.007.0540 6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.2 Manual reprocessing 6.2 Manual reprocessing WARNING Incomplete disinfection. Patients, users or third parties could be infected by contaminated medical devices.
6.2.3 Manual drying Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Hold the can vertically.
NOTE Please comply with the instructions for use of the washer disinfector. 4 In order not to affect the KaVo medical device, make sure that the product is dry on the inside and outside after completion of the cycle. 4 Remove any residual liquids with KaVoDRYspray.
6.4.2 Servicing SONICflex with KaVo Spray 4 Remove the tip from the medical device. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Press the spray key for 1 to 2 seconds.
Instructions for use SONICflex quick 2008 LS - 1.007.0540 6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.6 Sterilisation 6.6 Sterilisation Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060 / EN ISO 17665-1 CAUTION Improper service and care.
Instructions for use SONICflex quick 2008 LS - 1.007.0540 7 Other products 7 Other products Available from specialised dental dealers. Consumables Material summary Mat.No. INTRA Instrument stand 3.005.5204 Insert for SONICflex 3.005.3089 Cellulose pad 100 units 0.411.9862 Torque wrench 1.000.4887 Silicon paste 1.000.6403...
24 months from the date of the invoice, subject to the following conditions: With regard to justified complaints KaVo grants warranty in the form of a free of charge repair or delivery of a replacement. Other claims of any kind whatso- ever, in particular with respect to compensation, are excluded.