Setting the SPD-M30A
■ Setting the IP address of SPD-M30A
Connect to SPD-M30A using Internet Explorer to change
the IP address.
1)
Open the setting screen from IE.
Input "http://192.168.200.98" when newly installed.
2)
Log in as Administrator.
Input user ID and password, and press down the
Login button. When newly installed, input User ID:
Admin and Password: Admin.
3)
Change the IP address and Subnet mask
Input IP address and Subnet mask from the
Configuration tub.
Click on the Register button.
Changing any field in network settings and clicking
*
the [Register] button causes the settings on the
instrument to be changed and the connection to be lost.
In order to continue configuring settings other than
the network settings, enter the newly set IP address
into the web browser and log in again.
■ Setting Group name and Master name
When starting analysis synchronously with the
CBM-20A(lite), change the Group Name and Master
Name.
Perform setting from the Configuration tub.
- Set the same Group Name as CBM.
- Set System name of CBM to the Master Server Name.
Click on the Register button.
(Setting Example)
Setting of
CBM
Group Name
HPLC1
System Name
CBM1
Master Name
* It is possible to set IP address, Subnet mask, Group name and Master name from SPD-M30A Utility.
For details, see "5.3.9 Configuring Network Setting (SPD-M30A Utility)" in "Nexera SPD-M30A HELP".
Setting of
SPD-M30A
HPLC1
M30A1
CBM1