Disinfection guide
This guide provides the reprocessing instructions for the Sleep Lab Mask (SLM) variants of the AirTouch N30i, AirFit N30i, and AirFit P30i masks only, which are intended for multi-patient reuse in
a sleep lab, clinic or hospital. The Sleep Lab Mask variant is identified by the "SLM" acronym on the packaging label and a Unique Device Identifier labeling on the mask. If you use the mask as a
single user in the home, refer to the User Guide for cleaning instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed's validated procedures.
Potable water with hardness ≤150 mg/L is recommended for all cleaning, rinsing and disinfection steps.
Mask component
1
Cushion (AirTouch N30i)
Cushion (AirFit N30i)
Cushion (AirFit P30i)
Multi-hole vent
Frame (SLM)²
Elbow body/ elbow clip/ elbow ring
Headgear
QuietAir
vent
™
3
1
Masks and components may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Product Guide on ResMed.com.
2
The frame of the Sleep Lab Mask (SLM) does not contain fabric sleeves.
3
Replace with new parts between patients
9
4
8
7
6
5
Cleaning
Mild alkaline, anionic detergent (e.g. Alconox)
Thermal Disinfection - Manual
167°F (75°C) for 30 minutes
194-199°F (90-93°C) for 1 to 10 minutes
Not intended for multi-patient use
10
1
2
3
Validated number of cycles
30
-
1
Elbow body
2
Elbow ring
3
Headgear
4
Frame (SLM)
5
AirTouch / AirFit N30i cushion
6
AirFit P30i cushion
7
Multi-hole vent
8
QuietAir vent
9
Elbow vent
10
Elbow clip
Clinical use only
Additional disassembly
a
b