This guide is intended for multi-patient re-use of the AirFit N30i and AirFit P30i masks in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for
cleaning instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed's validated procedures.
Thermal Disinfection - Manual
Mask component
1
194-199°F (90-93°C) for 1 to 10 minutes
• AirFit N30i cushion
• AirFit P30i cushion
• Multi-hole vent
• Frame (SLM)
2
• Elbow body / elbow clip / elbow ring
• Headgear
This mask may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Product Guide on ResMed.com.
1
2
The frame of the Sleep Lab Mask (SLM) does not contain sleeves.
Page 1
This guide is intended for multi-patient re-use of the AirFit N30i and AirFit P30i masks in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for...
Page 2
When using detergents or disinfectants, always follow the manufacturer’s instructions. In the event of conflict, this guide takes precedence. ResMed Pty Ltd See ResMed.com for other ResMed locations worldwide. AirFit is a trademark of the ResMed family of companies. Alconox is a trademark of Alconox Inc. For patent and other intellectual property information, see 1 Elizabeth Macarthur Drive Resmed.com/ip.