ResMed AirFit N30i Disinfection Manual
Hide thumbs Also See for AirFit N30i:

Advertisement

Quick Links

Disinfection guide
This guide is intended for multi-patient re-use of the AirFit N30i nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning
instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed's validated procedures.
Thermal Disinfection - Manual
AirFit N30i mask component
1
194-199°F (90-93°C) for 1 to 10 minutes
• Cushion
• Frame (SLM)
2
• Elbow body / elbow clip / elbow ring
• Headgear
1
This mask may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Product Guide on ResMed.com.
2
The frame of the Sleep Lab Mask (SLM) does not contain sleeves.
1
D
E
1
Cleaning - Alconox
Validated number of cycles
F
A
B
C
30
30
30
30
A
Elbow body
B
Elbow ring
C
Headgear
D
Frame (SLM)
E
Cushion
F
Elbow clip
1
Vent
Clinical use only
English
1

Advertisement

loading

Summary of Contents for ResMed AirFit N30i

  • Page 1 Clinical use only This guide is intended for multi-patient re-use of the AirFit N30i nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning instructions.
  • Page 2 When using detergents or disinfectants, always follow the manufacturer’s instructions. In the event of conflict, this guide takes precedence. ResMed Ltd See ResMed.com for other ResMed locations worldwide. AirFit is a trademark of the ResMed family of companies. Alconox is a trademark of Alconox Inc. For patent and other intellectual property information, see 1 Elizabeth Macarthur Drive Resmed.com/ip.