Chroma 19071 Manual

Sentry hipot tester
Hide thumbs Also See for 19071:

Advertisement

Quick Links

www.valuetronics.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 19071 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Chroma 19071

  • Page 1 www.valuetronics.com...
  • Page 2: Legal Notices

    The information in this document is subject to change without notice. Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3: Warranty

    Any recommendations made by Chroma for use of its products are based upon tests believed to be reliable, but Chroma makes no warranty of the results to be obtained. This warranty is in lieu of all other warranties, expressed or implied, and no representative or person is authorized to represent or assume for Chroma any liability in connection with the sale of our products other than set forth herein.
  • Page 4 www.valuetronics.com...
  • Page 5 Storage. Freight. Maintenance. Disposal Storage When don’t use the device, please pack it properly and store under a good environment. (The packing is no needed when the device under appropriate environment.) Freight Please use the original packing material when move the device. If the packing material is missing, please use the equivalent buffer material to pack and mark it fragile and waterproof etc to avoid the device damage during movement.
  • Page 6: Revision History

    Revision History The following lists the additions, deletions and modifications in this manual at each revision. Date Version Revised Sections Jan. 2003 Complete this manual. Modify ”Specifications (18°C ∼ 28°C RH ≤ 70%)” April 2003 1.1 “Standard Accessory” “Rear Panel” “RS-485 Interface”...
  • Page 7: Table Of Contents

    HIPOT Tester 19071/19072/19073 User’s Manual Table of Contents Introduction .......................1-1 An Overview of Function..................1-1 Product Features......................1-1 Specifications (18ºC ∼ 28ºC RH ≤ 70%) ..............1-3 Standard Accessory ....................1-5 Optional Accessory ....................1-5 Operation and Setting....................2-1 Front Panel ......................2-1 2.1.1 Display Area....................... 2-1 2.1.2...
  • Page 8 HIPOT Tester 19071/19072/19073 User’s Manual IRR Calibration (IR Mode) ...................4-10 4.5.1 Connection Diagram ....................4-10 4.5.2 Insulation Resistance Calibration................4-10 Ground Continue Calibration ................4-12 4.6.1 Connection Diagram ....................4-12 4.6.2 Description of Ground Continue Calibration ............4-12 LCD Contrast Calibration ..................4-13 Complete Calibration ....................4-14...
  • Page 9: Introduction

    Introduction 1. Introduction 1.1 An Overview of Function www.valuetronics.com...
  • Page 10 HIPOT Tester 19071/19072/19073 User’s Manual for the outputted test voltage from zero rising to the specified value. www.valuetronics.com...
  • Page 11: Specifications (18ºc ~ 28ºc Rh ≤ 70%)

    Introduction 1.3 Specifications (18ºC ∼ 28ºC RH ≤ 70%) Withstanding Voltage Test AC: 0.05∼5kV / DC: 0.05∼6kV Constant Voltage (19072, 19073 Test Voltage only) ≤ 1%+5V, Rated Load Voltage Regulation V-display Accuracy ± (1% of reading + 5 counts), 2V resolution AC: 0.1mA ~ 20mA, DC: 0.01mA ~ 5mA (Note1), 0.1uAdc Cutoff Current (Note1) resolution...
  • Page 12 HIPOT Tester 19071/19072/19073 User’s Manual Indication, Alarm GO: (Short Sound) NG: W-Arc, W-Hi, W-Lo, IR-Lo, IR-Hi, GFI (Long Sound) Remote Connector Rear Panel 9 Pin D-type Input: Start, Stop, Interrupt Connector Output: Under test, Pass, Fail TEST/RESET Control Low - active control (24V open voltage typical).
  • Page 13: Standard Accessory

    Introduction 1.4 Standard Accessory Item Part Number Q’ty Description Power Cord W12 010130 Line cord 3P – 2P Adapter N31 000039 Power cord adapter Test Cable (1) W38 001940 Withstanding test cable for HIGH terminal Test Cable (2) W38 001760 Withstanding test cable for LOW terminal Test Cable (3) W38 001930...
  • Page 14 www.valuetronics.com...
  • Page 15: Operation And Setting

    Operation and Setting 2. Operation and Setting 2.1 Front Panel The front panel is divided into several function areas. This section explains each control and the information displays on LCD. Button Area Button Area Stop Button Display Area Function Key Display Area Function Keys Power Switch Status Line Pass Indicator Test Status Indicator Common Test Terminal...
  • Page 16: Button Area

    Start Button [START]: Press this button the instrument will be in test state, which means there are outputs at the test terminal and all judgment functions are activated. Calibration Switch: It is used by Chroma for calibration before shipping the product. Nonprofessional is prohibited to use this function to avoid causing any damage.
  • Page 17: Rear Panel

    2.2 Rear Panel The rear panel is divided into several areas. This section describes the function of each area. The standard rear panel of 19071/19072/19073: (2) (8) The rear panel of link test model of 19073 + RS485 is shown as below: (10) www.valuetronics.com...
  • Page 18 HIPOT Tester 19071/19072/19073 User’s Manual (1) REMOTE CONTROL www.valuetronics.com...
  • Page 19 Operation and Setting (4) EARTH GROUND TERMINAL: Terminal Grounding for Safety Use appropriate tools to ground this terminal surely. If it is not grounded surely, when the power circuit or any other devices connection short-circuited with the ground terminal, the instrument chassis may contain high voltage. This is very dangerous as it may cause electric shock if anyone touches it under the situation;...
  • Page 20: Operation

    HIPOT Tester 19071/19072/19073 User’s Manual 2.3 Operation [Standby Menu]: The screen appears as below after powered on. STEP 1/ 1 AC LOW :OFF PROGRAM ARC :OFF 0.000KV RAMP:OFF PRESET FALL:OFF 0.500mA MENU 3.0s MORE.. STANDBY LOCK OFST ERR 2.3.1 Memory Process...
  • Page 21: System Setup

    Operation and Setting (4) At this time use Function Key [F1] and [F2] to move the highlight to the memory for processing, then press Function Key [F3] will prompt a confirmation window. Press Function Key [ENTER] to confirm and Function Key [UP] or [DOWN] to select stored character of file name.
  • Page 22: Preset Parameter Setting

    HIPOT Tester 19071/19072/19073 User’s Manual (3) System Parameter Setting Table Item Range Default Description CONTRAST 1∼15 Adjust the LCD brightness BUZZER VOLUME HIGH/MEDIUM HIGH Adjust the buzzer volume /LOW/OFF EN50191 ON/OFF The output is meet to EN50191 requirement (AC Max 3mA,DC Max...
  • Page 23: Program Parameter Setting

    Operation and Setting 2.3.4 PROGRAM Parameter Setting www.valuetronics.com...
  • Page 24 HIPOT Tester 19071/19072/19073 User’s Manual (1) AC HiPot Test Mode STEP 1/1 LOW : OFF ARC : OFF VOLT:0.000kV RAMP: OFF DOWN HIGH:0.500mA FALL: OFF TIME: 3.0s NEXT EXIT SELECT MODE LOCK OFST VOLT : Set the required voltage for HiPot test.
  • Page 25 Operation and Setting (3) Insulation Resistance Test Mode STEP 1/1 HIGH : OFF RAMP : OFF VOLT :0.000kV DWELL: OFF DOWN LOW : 1.0MΩ FALL : OFF TIME : 3.0s NEXT EXIT SELECT MODE LOCK OFST ERR VOLT : Set the required voltage for insulation resistance test. : Set the low limit for insulation resistance.
  • Page 26: How To Perform Tests

    HIPOT Tester 19071/19072/19073 User’s Manual 2.4 How to Perform Tests www.valuetronics.com...
  • Page 27 Operation and Setting “Line 4” indicates the testing time set under STANDBY state; or the remaining testing time under TEST state. “Line 5” is the “Status Line” that shows the current state or test results. (3) Press [STOP] to enable it standby for test and the “Status Line” shows “STANDBY”. (4) Press [START] to begin the test.
  • Page 28: Other User Setting Function

    HIPOT Tester 19071/19072/19073 User’s Manual Other User Setting Function 2.5.1 KEY LOCK Function www.valuetronics.com...
  • Page 29: Auto Range

    Operation and Setting 2.5.2 Auto Range (1) Set WV-AUTO RANGE as ON under PRESET function list. (2) The current set to high range as below shown. STEP 1/1 AC LOW : OFF PROGRAM ARC : OFF 0.050kV RAMP : OFF PRESET FALL : OFF 20.00mA...
  • Page 30: Remote Control

    HIPOT Tester 19071/19072/19073 User’s Manual CHANGE PASSWORD NEW PASSWORD:**** ENTER RETURN 1-10 CHAR. LOCK OFST ERR (3) Use Function Key [A] and [B] to input the NEW PASSWORD (maximum10 characters), and press Function Key [ENTER] will display the “CONFIRM” window.
  • Page 31 Operation and Setting www.valuetronics.com...
  • Page 32 HIPOT Tester 19071/19072/19073 User’s Manual to control the external signal. The connecting point specification is 115V AC, current less than 0.3A. The action time is from the DUT is passed until it is stopped. www.valuetronics.com...
  • Page 33: Installation And Maintenance

    Installation and Maintenance 3. Installation and Maintenance 3.1 Notices for Installation This instrument generates high voltage output up to 6KV for external test. Any incorrect or mistaken procedures in using this instrument may cause injury or death. Thus for your safety, be sure to read through the notices described in this chapter and memorize them to prevent any accidents from happening.
  • Page 34 HIPOT Tester 19071/19072/19073 User’s Manual probe, and output terminal when the instrument is in test mode. DO NOT touch the alligator clip as shown below on the test cable when the system is in test mode as the rubber insulation is not good enough and may cause danger if touched.
  • Page 35 Installation and Maintenance <Formula> –t/RC Test Voltage * e = Residual Voltage –t/RC Ex.: 1000V * e = 30V –t/RC = ln 0.03 - t / RC = -3.5 t = 3.5 sec www.valuetronics.com...
  • Page 36 HIPOT Tester 19071/19072/19073 User’s Manual power cord. Do not use it again as the failure is very dangerous. Send it back Chroma or dealer for repair and services. www.valuetronics.com...
  • Page 37 Installation and Maintenance if over the range. The storage temperature is -10ºC~60°C, ≤ 80% RH. If you are not planning to use it for a long period of time, pack it with the original box for storage. For the sake of correct test and safety of this instrument, make sure not to store it in a place with direct sunlight or high temperature, also away from shaky, damp and dusty area.
  • Page 38: Troubleshooting

    Besides replacing the fuse (see 3.1.3 Solving Problems) or other notified in the manual, there is no troubleshooting to be performed by user. Please contact Chroma or sales agent when any abnormal condition occurs in this device. Do not perform maintenance by yourself to avoid causing any unnecessary hazards that may damage the device.
  • Page 39: Calibration Procedure

    Calibration Procedure 4. Calibration Procedure Before performing this section the HiPot tester must warm up at least for 30 minutes. Take off the calibration label at the front panel and press the lock switch. After completed the calibration, please press the lock switch again for opening the hardware data backup protection circuit to avoid calibration data loss.
  • Page 40 HIPOT Tester 19071/19072/19073 User’s Manual LCD Contrast Calibration CAL. CONTRAST (7) ; LCD contrast calibration 4.1 Calibration www.valuetronics.com...
  • Page 41: Dcv Calibration (Dc/Ir Mode)

    Calibration Procedure [ENTER] for saving. (5) Press [STOP] to stop ACV voltage offset calibration. (6) Press [UP] to change to the next calibration item. (7) Press [STOP][START] to read the high voltage meter value, ex. 4.052kV. (8) Press [INC.] or [DEC.] until shows 4.052 kV and press [ENTER] for saving. (9) Press [STOP] to stop ACV voltage full scale calibration.
  • Page 42: Irv Calibration (Dc/Ir Mode)

    HIPOT Tester 19071/19072/19073 User’s Manual (8) Press [INC.] or [DEC.] until shows 4.052 kV and then press [ENTER] for saving. (9) Press [STOP] to stop DCV voltage full scale calibration. 4.2.4 IRV Calibration (DC/IR Mode) (1) Press [UP] or [DOWN] to select IRV offset calibration.
  • Page 43: Current Calibration

    Calibration Procedure Current Calibration 4.3.1 Connection Diagram Caution The dummy load connected to HiPot terminal contains high voltage. Be very careful with it as it may cause hazard. 4.3.2 ACA Calibration (AC MODE) www.valuetronics.com...
  • Page 44 HIPOT Tester 19071/19072/19073 User’s Manual (6) Change the dummy load resistor to 500kΩ 10watt or higher. (7) Press [STOP][START] to read the ammeter value, ex. 2.450mA. (8) Press [INC.] or [DEC.] until shows 2.450mA and then press [ENTER] for saving.
  • Page 45: Dca Calibration (Dc/Ir Mode)

    Calibration Procedure 4.3.3 DCA Calibration (DC/IR MODE) www.valuetronics.com...
  • Page 46: Arc Calibration

    HIPOT Tester 19071/19072/19073 User’s Manual (12) Press [INC.] or [DEC.] until shows 2.45mA and then press [ENTER] for saving. (13) Press [STOP] to stop DCA 5.0mA range offset calibration. (14) Press [UP] to change to the next calibration item. (15) Change the dummy load resistor to 250kΩ 20watt or higher.
  • Page 47: Ac Arc Calibration (Ac Mode)

    Calibration Procedure Caution 1. ARC calibration is very special, the high voltage terminal are in outside. Be careful with it as it may cause hazard. 2. Please contact your local agent for more detailed descriptions. 4.4.2 AC ARC Calibration (AC MODE) www.valuetronics.com...
  • Page 48: Irr Calibration (Ir Mode)

    HIPOT Tester 19071/19072/19073 User’s Manual Press [RETURN] twice to exit ARC reading setting. Press [STOP] [START] for starting test. Press [INC.] or [DEC.] to adjust the low terminal of HiPot tester get closer to another terminal of the dummy load resistor gradually, and generate sparking between these two.
  • Page 49 Calibration Procedure Press [STOP] to stop IRR range 1 calibration. Press [UP] to change to the next calibration item. Change the standard dummy load resistor to 100MΩ. Press [STOP] [START]. Press [INC.] or [DEC.] until shows the standard resistance value, ex. 100MΩ and then press [ENTER] for saving.
  • Page 50: Ground Continue Calibration

    HIPOT Tester 19071/19072/19073 User’s Manual press [ENTER] for saving. (19) Press [STOP] to stop IRR range 4 calibration. Ground Continue Calibration 4.6.1 Connection Diagram 4.6.2 Description of Ground Continue Calibration www.valuetronics.com...
  • Page 51: Lcd Contrast Calibration

    Calibration Procedure (5) Press [ENTER] for saving the calibration value. (6) Press [STOP] to stop GC 5Ω Offset calibration. Change the standard resistor to 3Ω. (1) Press [UP]. (2) Connect resistor to ground and CONT.CHK OPTION terminal of rear panel. (3) Press [STOP][START] to read GC resistance value 3Ω.
  • Page 52: Complete Calibration

    HIPOT Tester 19071/19072/19073 User’s Manual 4.8 Complete Calibration www.valuetronics.com...
  • Page 53: Rs485 Interface (For Link Model Of 19073 + Rs485)

    RS485 Interface (For Link Model of 19073 + RS485) RS485 Interface (For Link Model of 19073 + RS485) If link model of 19073 + RS485 is purchased, please refer this chapter. 5.1 Description of Function RS485 interface can connect multi-device, and it up to multi-device test synchronously through Master operation.
  • Page 54: Terminal And Pin Signal Connection

    HIPOT Tester 19071/19072/19073 User’s Manual signal meanings of Master are as the following: UNDER TEST signal: It means Slave in testing. PASS signal: It means all Slave connections are normal and the test completed. FAIL signal: It means Slave not complete the test, may be the connections abnormal.
  • Page 55: Communication Protocol

    RS485 Interface (For Link Model of 19073 + RS485) 5.4 Communication Protocol This interface uses half duplex of 2-wire system non-synchronized transmission mode. Master Slave 1 Slave n Before the bus control is transferred, please wait the time of two characters. Master Slave1 Master...
  • Page 56: Command List

    HIPOT Tester 19071/19072/19073 User’s Manual Command Code Parameter 1 byte 0 ~ n-1 bytes Checksum: Summing these values into two supplements, included data are DA + SA + Length + Data Field 5.5 Command List Commands Code (Hex) *IDN? 0x90...
  • Page 57 0x01 = Source Address 0x16 = Data Field Length 0x90 = Command Code 0x43 0x48 0x52 0x4F 0x4E 0x4D 0x41 = “CHROMA” Company Name (return data) 0x2C = “,” (return data) 0x31 0x39 0x30 0x37 0x33 = “19073” Device Name (return data) 0x2C = “,”...
  • Page 58 HIPOT Tester 19071/19072/19073 User’s Manual 0x00 = Reply Message Return Data 0x0E = Checksum Stop Description: Stop test Command code: 0x21 Parameter: None Return data: Reply Message Example: Master (0x70): 0xAB 0x01 0x70 0x01 0x21 0x6D Slave (0x01): 0xAB 0x70 0x01 0x02 0x7F 0x00 0x0E...
  • Page 59 RS485 Interface (For Link Model of 19073 + RS485) 0x01 = Source Address 0x02 = Data Field Length 0x7F = Reply Message Command Code 0x00 = Reply Message Return Data 0x0E = Checksum Offset Get/Off Description: It switches OFFSET status. Command code: 0x23 Parameter:...
  • Page 60 HIPOT Tester 19071/19072/19073 User’s Manual The reading meanings of returning HEX code are as the following: 0xAB = Header 0x70 = Destination Address 0x01 = Source Address 0x02 = Data Field Length 0xA3 = Command Code 0x00 = 0: Off (Return Data)
  • Page 61 RS485 Interface (For Link Model of 19073 + RS485) Reserved Reserved Reserved Reserved GC Mode Name Size (byte) Unit Range Description Step index 1~10 Must less or equal original step number + 1 Mode GC mode Source 0/100 Reserved Reserved Dwell Time 100mS 1~10...
  • Page 62 HIPOT Tester 19071/19072/19073 User’s Manual 0x0E = Checksum The reading meanings of returning HEX code are as the following: 0xAB = Header 0x70 = Destination Address 0x01 = Source Address 0x02 = Data Field Length 0x7F = Reply Message Command Code...
  • Page 63 RS485 Interface (For Link Model of 19073 + RS485) 0x0B = Checksum Preset Parameters Description: It sets all parameters of Preset. Command code: 0x25 Parameter: 6 bytes Name Size Unit Range Description (byte) AC Frequency 50/60 Software AGC 0:OFF, 1:ON WV Auto Range 0:OFF, 1:ON IR Auto Range...
  • Page 64: Store Memory

    HIPOT Tester 19071/19072/19073 User’s Manual Return data: 6 bytes Example: Master (0x70): 0xAB 0x01 0x70 0x01 0xA5 0xE9 Slave (0x01): 0xAB 0x70 0x01 0x07 0xA5 0x3C 0x01 0x00 0x01 0x01 0x00 0xA4 Descriptions of example The setting meanings of HEX inputs are as the following:...
  • Page 65: Recall Memory

    RS485 Interface (For Link Model of 19073 + RS485) 0x43 0x48 0x52 0x4F 0x4D 0x41 = ”CHROMA” Memory Name (Command Parameter) 0xA6 = Checksum The reading meanings of returning HEX code are as the following: 0xAB = Header 0x70 = Destination Address...
  • Page 66: System Setting

    HIPOT Tester 19071/19072/19073 User’s Manual Descriptions of example The setting meanings of HEX inputs are as the following: 0xAB = Header 0x01 = Destination Address 0x70 = Source Address 0x02 = Data Field Length 0x28 = Command Code 0x01 = “01” Memory S/N (Command Parameter)
  • Page 67 RS485 Interface (For Link Model of 19073 + RS485) The reading meanings of returning HEX code are as the following: 0xAB = Header 0x70 = Destination Address 0x01 = Source Address 0x02 = Data Field Length 0x7F = Reply Message Command Code 0x00 = Reply Message Return Data 0x0E = Checksum System Setting?
  • Page 68 HIPOT Tester 19071/19072/19073 User’s Manual Descriptions of example The setting meanings of HEX inputs are as the following: 0xAB = Header 0x01 = Destination Address 0x70 = Source Address 0x02 = Data Field Length 0x2A = Command Code 0x01 = “01”key board lock ON, recall key lock (Command Parameter)
  • Page 69 RS485 Interface (For Link Model of 19073 + RS485) Initialize All Steps Parameters Description: It sets all steps to the initial setting, this command will delete all steps. Command code: 0x2C Parameter: None Return data: Reply Message Example: Master (0x70): 0xAB 0x01 0x70 0x01 0x2C 0x62 Slave (0x01): 0xAB 0x70 0x01 0x02 0x7F 0x00 0x0E Descriptions of example...
  • Page 70: Remote Status

    HIPOT Tester 19071/19072/19073 User’s Manual 0x05 = “05” the step numbers of being set (Return Data) 0xDB = Checksum Remote/Local Description: It switches the device to remote or local control. Command code: 0x2E Parameter: 1 byte 0: Go to Local.
  • Page 71 RS485 Interface (For Link Model of 19073 + RS485) 0x01 = Data Field Length 0xAB = Command Code 0xE0 = Checksum The reading meanings of returning HEX code are as the following: 0xAB = Header 0x70 = Destination Address 0x01 = Source Address 0x02 = Data Field Length 0xAE = Command Code 0x00 = “1”...
  • Page 72 HIPOT Tester 19071/19072/19073 User’s Manual Meter 3 Reserved Ramp Time 100mS 30000: Maximum, 31000: Not Value Dwell Time 100mS 30000: Maximum, 31000: Not Value Test Time 100mS 30000: Maximum, 31000: Not Value Fall Time 100mS 30000: Maximum, 31000: Not Value...
  • Page 73 RS485 Interface (For Link Model of 19073 + RS485) Return test result codes: Mode Code HEX DEC HEX DEC HEX DEC HEX DEC HEX DEC STOP USER INTERRUPT CAN NOT TEST TESTING PASS SKIP GFI FAIL HIGH FAIL LOW FAIL ARC FAIL I/O FAIL NO OUTPUT...
  • Page 74: Reply Message

    HIPOT Tester 19071/19072/19073 User’s Manual 0x01 = Step1 (Return Data) 0x74 = Test result code ”PASS” (Return Data) 0xD7 = Item number (Return Data) 0x01 = AC mode (Return Data) 0x63 0x00 = Display Volt 0.100kV (Return Data) 0x5A 0x00 0x00 0x00 = Curr 0.009mA (Return Data) 0xOF 0x00 = ramp time 1.5sec (Return Data)
  • Page 75: Flow Diagram

    RS485 Interface (For Link Model of 19073 + RS485) 5.7 Flow Diagram 1. Confirm if all of Slaves’ communication is normal or not: www.valuetronics.com...
  • Page 76 HIPOT Tester 19071/19072/19073 User’s Manual Upload the test steps from one slave then download to all slaves. www.valuetronics.com...
  • Page 77 RS485 Interface (For Link Model of 19073 + RS485) Read the measurement data: www.valuetronics.com...
  • Page 78 www.valuetronics.com...
  • Page 79: Maintenance

    Taiwan please contact your local Chroma distributor. 6.2 Instrument Return Before returning an instrument to Chroma for service please call our Service Department at 886-2-22983855 for return material authorization. It will be necessary to include a purchase order number to ensure expedient processing, although units found to be in warranty period will be repaired at no-charge.

This manual is also suitable for:

1907219073

Table of Contents