Download Print this page

HP 740B Operating And Service Manual page 53

Dc standard/differential voltmeter

Advertisement

Section V
Model 740B
Section V
Model 740B
FRONT PANEL TROUBLESHOOTING TREE
IIIOLATDlG THE TROUBLE TO A llECTION
Figure 5-14.
Front Panel Troubleshooting Tree
5-16
NOTES
01794-1
Table 5-8. Preliminary Checks
SYMPTOM
POSSIBLE CAUSE/CORRECTIVE ACTION
Inoperative (any or all
1. Check AC Fuses.
modes)
2. Make sure Internal Alignment Switch (S3)
is in "Operate" position.
3. Check Power Supply Voltages (Figure 7-3,
Sheet 2 of 4).
Erratic Operation or
Switches may be dirty. Turn instrument off;
Poor Repeatability
rotate all switches several times (including
Internal Alignment Switch, 83). Switches
are self-cleaning (Paragraph 5-108).
OVERLOAD Light On
Check DC Fuse.
If
blown, replace.
If
fuse
(all modes)
blows again, Q1 or Q2 may be shorted.
Proceed to Main Loop Troubleshooting Tree,
step 34.
OVERLOAD Light On
Normal indication. In STD mode, mV Ranges
(STD mode, mV Ranges)
are inoperative (Paragraph 3 -15, step h).
Output voltage inaccurate
1. Make sure SENSE terminals are -connected
correctly (Paragraph 3 -17). Remote
Sensing must be used for remote loads.
2. Make sure ZERO control is set properly
(Paragraph 3-31).
~
VM mode inaccurate
1. Make sure INPUT Z Switch is set to
00
position.
2. Make sure ZERO control is set properly
(Paragraph 3-46, 3-47).
Unable to zero Input or
1. Recheck Internal Zero Adjustment (Para-
Output with ZERO control
graph 5-44), Balance Adjustment (Para-
graph 5-46) and Meter Zero Adjustment
(Paragraph 5-50).
2. Check A10BT1 (1. 3 V minimum).
OUTPUT Indicator lighted
Normal indication. OUTPUT Indicator lights
when Output Cable is not
when depressed even when Output Cable As-
connected
sembly is not connected. The Output Rotary
Switch (K1), however, does not energize.
Internal Buzzing when
This is normal.
The noise comes from the
changing Ranges or ex-
Internal Current Limit Circuit.
ceeding Current Limit
Noisy meter on 1 mV Range,
Normal indication.
(Paragraph 3-59, step g)
X10 4 Sensitivity
5-16
54.
_rument
{
::~~
50. Perform STD SlId
.0.
VM
{
Accuracy Tests. _ru-
ment paue8.
44.
Select VIiI,
1
m V Range.
Apply I mV input. Instru-
55. See Text.
{
ment gives correct indica-
tion.
38. Select VM. 1000 V
.
51. Troubleohoot A3
RaDp
Range 1000 V input. In-
SwltchlDg.
strument gives correct
indication.
45. Troubleshool Range
Switching (used in
VM).
48. Troubleshoot M.in Loop.
37. Remove 50 kllioad. O u t {
put monitor indicates
1000 V. OVERLOAD.
52. Troubleshoot Range
light is
off.
{SwItChing (used in STD).
47. Select VM. 1000 V Range.
Apply 1000 V input.
in-
strument gives correct
1nd1cation.
53. Troubleshoot Main Loop.
48. Troubleshoot Function
SwllchlDg (.o.VM).
38. Select VM and check
meter indication on all
sensitivities. Meter
gives correct indications.
I V.
Loop.
39. Troubleshoot Func.tion
Switching ("VM).
33.
34. Select 10 V Range, XI0
{400
i.::leshoqt Meter
SENSITIVITY. Meter
gives correct indication.
41. Troubleshoot Function
80rllcblllc
(VIiI).
32. Select STD, 1000 V
Range. 1000 V Out-
put. Coonect a 50 til
20 W load across Out-
IUt
terminals. Inatru-
ment supplies
1000 V
output.
27. Output monitor indicates
IV.
26. Select "'VM. I V Range.
Apply I V in!"t. Instru-
ment gives correct
indi-
cation.
28. Troubleshoot STANDARD
{
VERNIER circuit.
21. Select VM. 10 V Range,
XIO SENSITIVITY. Apply
1 V
input.
Meter indica
s
+1.
O.
29. Troubleshoot Meter Loop.
20. Select VM,
I
V Range.
Apply I V input. Meter
gives correct indication.
01794-1
30. Troubleshoot Function
{
Switching
('ITD).
22. Select .6.VM. Instrument
{
gives correct indication.
31. Troubleshoot Reference
17. Select
VM.
Apply I V
L90p.
input. OJtput monitor
indicates
1 V.
23. TrotKJleshoot Main Loop.
16. Observe meter. Meter
indicates
-+
1. O.
8. Select STD, I V Range,
{
I V output. Output moni-
tor indicates l' V.
4. Select .o.VM. I V Range,
XI0 4 SENSlT1VITY. Ad-
joet ZERO. Meter Indi-
catea
0 V.
Noise
<
2 J,J.V
pop (I minor division).
9. Troubleshoot Function
Switching (.o.VM).
10. Troubleshoot Function
{
Switching (STD).
5.
~1~~~Arri~~d-
just ZERO. Meter indi-
cates
0 V.
Noise
<
2 JJ.V
pop.
11. Troubleshoot Reference
L90p.
6. Select .o.VM, I V Ranse.{12'
~~t:I~~~.unCtion
Xl0 4 SENSITIVITY. Ad-
just ZERO. Meter indi-
cates
0 V.
Noise
<
2 IlV
pop (I minor division).
13. Troubleshoot Meter
L90p.
{ 1 8 '
~::h~f'ncllon
14. Select .o.VM, I V Range,
XI04SENSITIVITY. Ad-
{
~':~~~~'N~ie::~~:~
[24'
~~::~e~:~~PUt
Pro-
pop.
7. select STD. 1 V Range.
19 Connect jwnper from
+
o
V output. Adjust ZERO.
. INPUT
to
AI7Vl (redwire).
C:::~::;:l~rlJ~c_~.es
Noiae
and
offset disappear.
15. Troubleshoot M.in L90p.
25. Troubleshoot Main L90p.
2. Select STD, 1 V Range, OV
output. Adjust ZERO.
Output monitor indicates 0
V. Noi.. < 2
~V
pop.
3.
(l)aerve output monitor.
Adjust ZERO. Output monl-
tor iDcUcates 0 V. Noise on
output
<
2
~
V pop.
FAILS
7408-0-30019
PASSES
t
2. Refer to Functional DIagram, Figure 7-3. to ob-
serve overall instrument operation.
I.
Uae.
sensitive dc voltmeter (-hp- Model 419A)
connected to Output terminals to monitor output
voltage and noise.
3. All check. (except No. 19) should be performed
with
instrument covers and shields
in
place.
4. Procedure.ssume.boardpluu•. cl.... Slldmaldac
good contact and power supplies are workiDc. Refer
10
Fleure
7 -3 lor
.-er
supply voltage. SlId lest
points.
I. Select
VM,
1 V Range, In-
pulohorted.
xl04
SENSl-
TJVITY.
Adjust ZERO.
Meter indicates O. Meter
noise <
2
It
V p-p (l
minor
division).

Advertisement

loading
Need help?

Need help?

Do you have a question about the 740B and is the answer not in the manual?