  • Page 1 Installation Guide HP 70909A and HP 70910A RF Section ABCDE HP Part No. 70909-90002 Printed in USA April 1993 www.valuetronics.com...
  • Page 2 Notice The information contained in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and tness for a particular purpose.
  • Page 3: Certi Cation

    Certi cation Hewlett-Packard Company certi es that this product met its published speci cations at the time of shipment from the factory. Hewlett-Packard further certi es that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
  • Page 4: Safety Symbols

    Safety Symbols The following safety symbols are used throughout this manual. Familiarize yourself with each of the symbols and its meaning before operating this instrument. Caution caution sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in damage to or destruction caution of the instrument.
  • Page 5: General Safety Considerations

    General Safety Considerations Warning , make sure it has been properly Before this instrument is switched on grounded through the protective conductor of the ac power cable to a socket outlet provided with protective earth contact. Any interruption of the protective (grounding) conductor , inside or outside the instrument, or disconnection of the protective earth terminal can result in personal injury.
  • Page 6: Installation At A Glance

    Installation at a Glance www.valuetronics.com...
  • Page 7 Electrical components are easily damaged by small amounts of static electricity. If possible, work at a static-safe work station. Installation overview The installation procedure for the HP 70909A or HP 70910A RF section is straightforward. The following major steps are included: Step 1. Switch o the system's ac line power Step 2.
  • Page 8: In This Book

    This book describes all of the installation procedures neccessary to install and check the proper operation of either an HP 70909A RF section or an HP 70910A RF section in an HP 70000 Series modular spectrum analyzer system. It also contains information on what to do if problems occur during installation.
  • Page 9: Table Of Contents

    Step 4. Unpack your new RF section ..... Step 5. Check or set the HP-MSIB address on your new RF section ..
  • Page 10 Figures 2-1. Line Voltage Selector ......2-2. Line Fuse Removal and Replacement ....Tables 2-1.
  • Page 11: Installation

    Installation This chapter contains step-by-step instructions that are required for proper installation of either an HP 70909A RF section or an HP 70910A RF section in an HP 70000 Series modular spectrum analyzer system. Tools needed Before installation, assemble these tools:...
  • Page 12 HP 70000 Series modular spectrum analyzer to the HP 70000 Modular Spectrum Analyzer system's display and mainframe.
  • Page 13: Step 2. Disconnect The Existing Rf Section's Rear Panel Cables

    Step 2. Disconnect the existing RF section's rear panel cables Disconnect the SMB cable from the Disconnect the SMB cable from the existing RF section's 21.4 MHz OUT connector. existing RF section's 300 MHz IN connector. Disconnect the SMB cable from the Disconnect the SMA exible cable from existing RF section's TUNE SPAN connector.
  • Page 14: Step 3. Remove The Existing Rf Section

    RF section to make room for your new RF section. An HP 70909A or HP 70910A RF section occupies two adjacent 1/8 module slots in an HP 70000 Series modular spectrum analyzer system mainframe.
  • Page 15: Step 4. Unpack Your New Rf Section

    HP 70910A RF section from its shipping included in the shipping container . container and inspect it thoroughly to ensure that it was not damaged during shipment. 1. HP 70909A RF section or HP 70910A RF section HP 70909A and HP 70910A Installation Guide HP 71209A Quick Start Guide 4.
  • Page 16: Step 5. Check Or Set The Hp-Msib Address On Your New Rf Section

    Step 5. Check or set the HP-MSIB address on your new RF section Locate the HP-MSIB address switches on Check that the HP-MSIB address on your the left-rear side of your new RF section. new RF section is set to row 6 and column 18.
  • Page 17: Step 6. Install Your New Rf Section

    Step 6. Install your new RF section Install your new RF section into the Press against your new RF section's front mainframe. panel while tightening the hex-nut latch with an 8 mm hex-ball driver . This locks your new RF section into position. Close the mainframe front-panel door to cover the hex-lock screw.
  • Page 18: Step 7. Connect The New Rf Section's Rear Panel Cables

    Connect an SMB cable from the LO RF section's 21.4 MHz OUT connector to the section's 300 MHz OUT connector to your new HP 70903A IF section's 21.4 MHz IN RF section's 300 MHz IN connector. connector. Connect an SMB cable from the LO...
  • Page 19: Step 8. Switch On The System's Ac Line Power

    HP 70004A color display and on the bottom of the HP 70001A mainframe. Caution Switch on the ac line power to the HP 70000 Series modular spectrum analyzer system display and mainframe.
  • Page 20: Step 9. Check Proper Operation Of Your New Rf Section

    Step 9. Check proper operation of your new RF section Check that the MEASURE LED is blinking Each time the HP 70000 Series modular at the end of the power-on self test. spectrum analyzer system is turned on, the following actions take place: 1.
  • Page 21: If You Have Problems With Installation

    If you have problems with installation This chapter contains information to help identify and resolve problems that may occur with the installation of your HP 70909A or HP 70910A RF section. Symptoms to various problems are listed at the top of each page.
  • Page 22: If The System's Power-On Self Test Fails

    Check that the line voltage selector switch is set to the correct voltage for the ac line voltage being used. The line voltage selector switch is located on the left side of the HP 70004A display or on the bottom of the HP 70001A mainframe.
  • Page 23 If necessary, obtain HP service. (Refer to \If you require additional technical resources".) If you have problems with installation www.valuetronics.com...
  • Page 24: If Your Rf Section Is Powered On But Not Responding Correctly

    Perform the checks in the procedure \If the system's power-on self test fails". Check that other equipment, cables, and connectors are connected and operating correctly. Check the HP 70000 Series modular spectrum analyzer system display for error messages. If necessary, obtain HP service. (Refer to \If you require additional technical resources".) If you have problems with installation www.valuetronics.com...
  • Page 25: If The Status Err Indicator Led On Your New Rf Section Is Ashing

    If the STATUS ERR indicator LED on your new RF section is ashing The HP 70909A or HP 70910A RF section communicates with the HP 70000 Series modular spectrum analyzer system over the HP-MSIB. If the STATUS ERR indicator LED, located on the RF section's front panel, is ashing at a 1 Hz rate, the module cannot communicate over the HP-MSIB.
  • Page 26: If More Than One Module's Error Indicator Is Ashing

    If more than one module's error indicator is ashing The HP 70909A or HP 70910A RF section communicates with the HP 70000 Series modular spectrum analyzer system over the HP-MSIB. When the error indicator LED on a module is ashing at a 1 Hz rate, then that module is not communicating over the HP-MSIB.
  • Page 27: If Hardware Errors Messages Occur

    If hardware errors messages occur Hardware errors are generated when a module in the HP 70000 Series modular spectrum analyzer system is not working properly. These errors can occur at any time. Hardware errors range from 7000|7999. (Refer to the HP 70909A and HP 70910A Service Guide information about other error messages.)
  • Page 28 This hardware error occurs when the A20 Controller has detected an incorrect drive voltage on the A19 PS/YTF driver assembly. To solve this problem: 1. Refer to the HP 70909A and HP 70910A Service Guide for troubleshooting information. 2. If necessary, obtain HP service. (Refer to \If you require additional technical resources".)
  • Page 29: If You Require Additional Technical Resources

    RF section after warranty expiration (your Hewlett-Packard sales and service oce can provide full details). If your HP 70909A or HP 70910A RF section are covered by a separate maintenance agreement, please be familiar with its terms before calling Hewlett-Packard for service.
  • Page 30 Table 2-1. Hewlett-Packard Sales and Service Oces US FIELD OPERATIONS EUROPEAN OPERATIONS INTERCON OPERATIONS HEADQUARTERS HEADQUARTERS HEADQUARTERS California Australia France Canada Germany Colorado Japan Great Britain Georgia China Illinois New Jersey Singapore Texas Taiwan 2-10 If you have problems with installation www.valuetronics.com...
  • Page 31: Returning Your Equipment For Service

    HP 70000 Series modular spectrum analyzer system display a completed Performance Test record any other speci c data on the performance of the HP 70909A RF section or HP 70910A RF section Caution Damage can result if the original packaging materials are not used. Packaging materials should be anti-static and should cushion the RF section on all sides.
  • Page 32 Table 2-2. Packaging and Contents Item Description HP Part Number Qty Carton-outer 9211-5118 Carton-inner 9211-5119 Carton-sliders 5180-2369 Foam inserts 4208-0493 Foam pads 5180-2370 2-12 If you have problems with installation www.valuetronics.com...