Baker Hughes Druck DPI610E Instruction Manual

Portable pressure calibrator
Hide thumbs Also See for Druck DPI610E:
Table of Contents

Advertisement

Quick Links

DPI610E
Portable Pressure Calibrator
Instruction Manual
All Variants
Druck.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Druck DPI610E and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Baker Hughes Druck DPI610E

  • Page 1 DPI610E Portable Pressure Calibrator Instruction Manual All Variants Druck.com...
  • Page 3 This equipment is safe when you use the procedures in this manual. Do not use this equipment for any other purpose than that specified. This is because the protection given by the equipment can be reduced or canceled. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | i...
  • Page 4 Please write to your local or regional waste administration if you need more information on the collection, reuse, and recycling systems. Please use the link below for take-back instructions and more information about this program. https://druck.com/weee Copyright 2022 Baker Hughes Company. ii | DPI610E Instruction Manual–English...
  • Page 5: Acronyms And Abbreviations

    Relative Humidity Foot RS-232 Serial communications standard Gauge Thermocouple Water Universal Serial Bus Hertz Volts i.e. That is Virtual Communications Port Inch °C Degrees Celsius Intrinsically Safe °F Degrees Fahrenheit kilogram Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | iii...
  • Page 6 Copyright 2022 Baker Hughes Company. iv | DPI610E Instruction Manual–English...
  • Page 7: Table Of Contents

    Pump (Pneumatic Version) 1.13.9 Priming Pump (Hydraulic Version) 1.13.10 Volume Adjuster (Pneumatic Version) 1.13.11 Volume Adjuster Wheel (Hydraulic Version only) 1.13.12 Pressure/Vacuum Selector (Pneumatic Version only) 1.13.13 Auxiliary Ports 1.13.14 External Sensor Port Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual| i...
  • Page 8 How to Add Tasks to the Dashboard How to Remove Tasks from the Dashboard General Settings DATE, TIME and LANGUAGE BACKLIGHT COMMUNICATIONS AUTO POWER DOWN TOUCHSCREEN LOCK ENABLE HOLD ADVANCED Copyright 2022 Baker Hughes Company. ii | DPI610E Instruction Manual–English...
  • Page 9 Millivolts Measure - Setup 6.3.18 HART Process Options 6.4.1 Tare 6.4.2 Min/Max/Mean 6.4.3 Filter 6.4.4 Flow 6.4.5 Alarm 6.4.6 Scaling Utilities Leak Test Switch Test TX (Transmitter Simulation) Simulator Relief Valve Test Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual| iii...
  • Page 10 DATA POINTS 10.4 Setting up Manual Data Logging 10.5 How to Do Periodic Data Logging 10.6 Viewing & Deleting Data Log Files 10.6.1 To View Data Log Files on the Instrument Copyright 2022 Baker Hughes Company. iv | DPI610E Instruction Manual–English...
  • Page 11 How to Start the HART Application (Method 1) 13.1.2 How to Start the HART Application (Method 2) 13.1.3 HART Device Electrical Connection 13.2 HART Device Configuration 13.3 HART Dashboard 13.3.1 Unique ID Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual| v...
  • Page 12 15.8 How to View the File System on a PC 15.9 Favorites, Error Log and Event Log 16. Status Menu 16.1 Status Menu Options 16.2 How to Show the Status Menu Screen 16.3 SOFTWARE BUILD 16.4 CALIBRATION 16.5 BATTERY Copyright 2022 Baker Hughes Company. vi | DPI610E Instruction Manual–English...
  • Page 13 Appendix A. COMPLIANCE STATEMENTS FCC (USA) A.1.1 Federal Communication Commission Interference Statement A.1.2 FCC Radiation Exposure Statement CANADA A.2.1 ISED Canada Statement A.2.2 Radiation Exposure Statement A.2.3 Déclaration d'exposition aux radiations Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual| vii...
  • Page 14 Copyright 2022 Baker Hughes Company. viii | DPI610E Instruction Manual–English...
  • Page 15: Overview

    The instrument uses application firmware. Refer to “SOFTWARE BUILD” on page 259 for how to find the new versions of the firmware applications. Do checks regularly for updates of this firmware and the End User Software Release Note. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 1...
  • Page 16: Dpi610E Variants

    (7 MPa - 100 MPa) DPI610E Aero DPI610E-A Blue DPI610E 2 bar a Aero - (29.6 psi) Pneumatic Pneumatic - Non-IS (0.2 MPa) DPI610E-PC DPI610E-HC DPI610E-A DPI610E-SPC DPI610E-SHC Figure 1-1: DPI610E Variants Copyright 2022 Baker Hughes Company. 2 | DPI610E Instruction Manual–English...
  • Page 17: Equipment In The Package

    Blanking Plug 111M7272-1 Hose sets: 1 m long Safe Area IO620-HOSE-H1 Intrinsically Safe IO620-HOSE-H1-IS Safe Area PV411-115 Reservoir Intrinsically Safe PV411-115-IS - packed separately 250 ml Reservoir Filler Bottle 1S-11-0085 Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 3...
  • Page 18: Specifications And Accessories For Dpi610E

    Note: Before the use of the equipment always read and obey all warnings and cautions given in the DPI610E Quick Start and Safety Manual. Copyright 2022 Baker Hughes Company. 4 | DPI610E Instruction Manual–English...
  • Page 19: Safety

    Do not use the blue case color instruments in locations that have explosive gas, vapor or dust. There is a risk of an explosion. Refer to the Quick Start and Safety Manual that comes with the instrument. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 5...
  • Page 20: General Warnings

    Note: Maximum Operating Pressure (MWP) is given on the label on the bottom of the instrument. Over pressure must be limited to 1.2 × MWP (MWP is based on unit pressure range). Copyright 2022 Baker Hughes Company. 6 | DPI610E Instruction Manual–English...
  • Page 21: Overvoltage Categories

    1.5.2 How to Clean the Instrument Use a damp cloth with water and mild detergent to clean the surface of the instrument. Do not put the instrument into water. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 7...
  • Page 22: Calibration

    Use an approved organization that collects and/or recycles unwanted electrical and electronic equipment. For more information go to our website for Customer Service Department or go to your local government office: https://www.bakerhughes.com/druck Copyright 2022 Baker Hughes Company. 8 | DPI610E Instruction Manual–English...
  • Page 23: Packaging For Storage Or Transportation

    IO610E-CAR CHARGER (optional accessory). The instrument can be charged while energized (Power On) and also when de-energized (Power Off). Battery charge time is approximately two hours from empty to fully charged. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 9...
  • Page 24: Battery Status Indicator

    The instrument has an Auto Power Down function that can be selected or canceled when wanted. See General Settings Section 4 on page 37. On the instrument the function is in the General Settings screen on the Dashboard. Copyright 2022 Baker Hughes Company. 10 | DPI610E Instruction Manual–English...
  • Page 25: Parts

    Pressure/Vacuum Hand Pump Softkeys Instrument Dirt (& Moisture) Trap (IDT) User Interface Clip Fastener for Carrying Strap Power ON/OFF Button Barometer Port Help Button Battery Level Indicator Home Button External Sensor Port Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 11...
  • Page 26: Hydraulic Instruments

    The Test Port is at the top left corner of the instrument. Pressure can be supplied to pressure devices that are connected, either directly or by the use of compatible hose fittings. The Test Port Copyright 2022 Baker Hughes Company. 12 | DPI610E Instruction Manual–English...
  • Page 27: Reservoir (Hydraulic Version)

    On the right side of the instrument is the RS485 communications port for the connection of external remote sensors, such as the PM700E pressure sensor and the RTD-Interface and RTD- probe. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 13...
  • Page 28: Barometric Port (Pneumatic Version Only)

    Dashboard home screen from all locations in the user menu. 1.14.1.4 Enter Button An Enter button is in the middle of the Navigation Pad. 1.14.1.5 Navigation Pad Copyright 2022 Baker Hughes Company. 14 | DPI610E Instruction Manual–English...
  • Page 29: First Use

    Analysis, Documenting, and Data Log icons are set and thus always available. Note: Shortcut icons can be added to the User Interface for applications, (See Chapter 3 on page 33). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 15...
  • Page 30: Dashboard Softkeys

    Touchscreen, tap the wanted time or date parameter to be changed. Tap on the right side of the set value (bold font) to increase or on the left side to decrease the value. Copyright 2022 Baker Hughes Company. 16 | DPI610E Instruction Manual–English...
  • Page 31 Push the Tick Softkey to accept the changes. If changes are not to be saved or to go back to the previous screen, tap the Back button. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 17...
  • Page 32 Chapter 1. Overview Copyright 2022 Baker Hughes Company. 18 | DPI610E Instruction Manual–English...
  • Page 33: Pump Operations

    To vent the system to atmospheric pressure, slowly turn the pressure release valve fully counterclockwise until the sensor reading goes to zero (gauge sensor) or 1 bar (absolute sensor). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 19...
  • Page 34: To Attach The Instrument Dirt Trap To The Test Port

    Note: Make sure that the device under test has a male G ⅜ Quickfit adapter thread or use a suitable adapter rated to 35 bar. If in doubt, please contact our Service Support - see the back of this manual. Copyright 2022 Baker Hughes Company. 20 | DPI610E Instruction Manual–English...
  • Page 35: To Remove The Device Under Test

    Select the necessary pressure function from the Channel Function screen. From the dashboard select: CALIBRATOR > > FUNCTION > Pressure > INT > Normal Select the Tick Softkey and tap the Back icon. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 21...
  • Page 36: To Supply Medium Pressure Or Vacuum

    Turn the Pressure Release Valve firmly fully clockwise to seal the system. CAUTION Do not put the instrument into the pressure medium. If fluid goes into the electrical connection panel, remove the fluid and let the instrument fully dry before use. Copyright 2022 Baker Hughes Company. 22 | DPI610E Instruction Manual–English...
  • Page 37 350 mbar and 1 bar, or use the Volume Adjuster for finer control on pressure. If the pressure is below the necessary value, go back to step 5. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 23...
  • Page 38: Hydraulic System

    Note: We recommend demineralized water or mineral oil as the pressure medium. CAUTION Do not let the instrument go fully into the pressure medium. If fluid goes into the electrical connection panel, let the instrument fully dry before use. Copyright 2022 Baker Hughes Company. 24 | DPI610E Instruction Manual–English...
  • Page 39 Push the reservoir cover into position and turn the locknut clockwise (finger tight) until the reservoir cover seals the reservoir. Then rotate back a quarter turn (counterclockwise). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 25...
  • Page 40: How To Prime The Instrument

    Either tap the screen or use the Navigation Pad arrow keys to select the wanted unit. Push the Tick Softkey and push the Back button to show the previous screen. Copyright 2022 Baker Hughes Company. 26 | DPI610E Instruction Manual–English...
  • Page 41 10. Pull the Priming Pump backwards until it stops. Start turning the Volume Adjuster counterclockwise. While turning the Volume Adjuster, push the Priming Pump slowly and simultaneously inward until the Volume Adjuster stops (a minimum of 29 turns can be Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 27...
  • Page 42 Note: It can take 10-15 full moves of the pump handle to remove trapped air from the system. CAUTION When air bubbles do not come out of the hole in the reservoir center tube, stop operating the pump. Copyright 2022 Baker Hughes Company. 28 | DPI610E Instruction Manual–English...
  • Page 43 5 turns. Now use the priming pump gently (very slowly) otherwise fluid will eject at speed from the hose end. We advise that you hold the hose end in a suitable Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 29...
  • Page 44 20. Use the Priming Pump to prime the system to a maximum fluid pressure of 10-25 bar. Now the system is ready to make the necessary pressure - see the next section. Copyright 2022 Baker Hughes Company. 30 | DPI610E Instruction Manual–English...
  • Page 45: How To Adjust Pressure

    Move the Pump handle seven strokes to move hydraulic fluid into the container. Turn the Volume Adjustment Wheel clockwise until it stops. Move the pump handle seven strokes again. This will flush out old fluid and replace it with new fluid. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 31...
  • Page 46 12. Make sure the Pressure Release Valve is closed (hand-tight) then energize (Power On) the DPI610E. 13. Operate the pump until the pressure reading is at 20 bar. 14. Open the Pressure Release Valve. Copyright 2022 Baker Hughes Company. 32 | DPI610E Instruction Manual–English...
  • Page 47: Basic Tasks

    CH1 will default to show the measured internal pressure. CH2 is set to show the measured Voltage. This task is typically for the calibration of Voltage Output pressure transmitters. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 33...
  • Page 48: I - P (Current To Pressure)

    Tap the option again to start the function that sets the selected task. This is effectively a two- tap action that is done quickly. To use the Navigation Pad buttons: use the Copyright 2022 Baker Hughes Company. 34 | DPI610E Instruction Manual–English...
  • Page 49: How To Add Tasks To The Dashboard

    To use the Navigation Pad buttons, use the UP/DOWN/LEFT/RIGHT buttons to select the wanted task. Then push the Softkey to remove the selected option and its related small HOME icon. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 35...
  • Page 50 Chapter 3. Basic Tasks Copyright 2022 Baker Hughes Company. 36 | DPI610E Instruction Manual–English...
  • Page 51: General Settings

    This function controls the instrument backlight. The DPI610E is usually set to Timed (5 Minutes). To change this value of this function: Select Backlight from the General Settings menu. Select the wanted mode (shown below). Select to make the selection. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 37...
  • Page 52: Communications

    This function controls how the instrument de-energizes. The DPI610E is set to automatically stay on until de-energized by the user. To change this setting: Select Auto Power Down. Select one of the three modes. Select to make the selection. Copyright 2022 Baker Hughes Company. 38 | DPI610E Instruction Manual–English...
  • Page 53: Touchscreen Lock

    4.6 ENABLE HOLD This function lets the user control the mode of the Hold Softkey when on the screen. The DPI610E automatically has Enable Hold selected. To disable this function: Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 39...
  • Page 54: Advanced

    See Section 5.2 on page 41 for more information. Select to continue. The Advanced Menu screen is now unlocked and its options available (see Chapter 5, “Advanced Menu,” on page 41). Copyright 2022 Baker Hughes Company. 40 | DPI610E Instruction Manual–English...
  • Page 55: Advanced Menu

    Refer to Chapter 14, “Instrument Calibration,” on page 229 that describes the options available when this CALIBRATION menu becomes available for use. 5.2 CHANGE PIN This option allows the user to change the instrument PIN number. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 41...
  • Page 56: Software Upgrade

    Dashboard. To select, tap on the icon or push the Softkey to the right of the icon. The General Settings screen appears. Select the COMMUNICATIONS option. Copyright 2022 Baker Hughes Company. 42 | DPI610E Instruction Manual–English...
  • Page 57 Use a micro-USB data cable to connect the DPI610E to the PC. The PC will automatically sense the DPI610E. The screen will show the PC as a Mass Storage drive (default name of this drive is DPI610E). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 43...
  • Page 58: How To Upgrade The Firmware

    QR image: a mobile phone can be used to scan this image. This will show a website on the screen. This screen gives instructions for how to change the application to a new version. Select to show the Status screen again. Copyright 2022 Baker Hughes Company. 44 | DPI610E Instruction Manual–English...
  • Page 59: Factory Reset

    Select the OK button to erase all user files. Select the OK button to start instrument again and complete the change operation. If you select Cancel, the change will be completed on the next start of the system. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 45...
  • Page 60: Format File System

    5.6 SERVICE / ENGINEERING This Advanced function is not for operator use. A special PIN is needed for use only by specialist users. Copyright 2022 Baker Hughes Company. 46 | DPI610E Instruction Manual–English...
  • Page 61: Calibrator Tasks

    Softkey – up to four Softkeys are available for the menu screens. These Softkeys give Touchscreen and button options for the use of different parts of the user interface. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 47...
  • Page 62: Calibrator Tasks Screen Shortcuts

    When in the default 50:50 channel window layout, push the UP button to select the Channel 1 (CH1) window area. INFORMATION Push the DOWN button to select the Channel 2 (CH2) window area. Copyright 2022 Baker Hughes Company. 48 | DPI610E Instruction Manual–English...
  • Page 63: Change Measurement Units

    To change the measurement units on each channel on the calibrator task screen: Select the Unit text on the related channel window. From the dashboard select: CALIBRATOR > > UNITS. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 49...
  • Page 64 SI and Non-SI units. An SI type can only show SI units on the screen. The selected channel window shows the wanted measurement unit. Note: Section 2.1.7 on page 21 gives an alternative method to select units of measurement. Copyright 2022 Baker Hughes Company. 50 | DPI610E Instruction Manual–English...
  • Page 65: 24 V Loop Power Enable/Disable

    Tap on the 10 V/24 V text field on the screen to select the loop power. Tap again on the 10 V/24 V function to set the loop power off - shown by the removal of the voltage text field. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 51...
  • Page 66: Error Indications

    Make sure that the 10 V/24 V function is not in the screen after the function has been canceled. 6.2.5 Error Indications An out-of-range error message occurs when the values of the primary reading measure are more than the full-scale value of the function range. Copyright 2022 Baker Hughes Company. 52 | DPI610E Instruction Manual–English...
  • Page 67 INFORMATION If the screen shows any of the above error indications: look at the sensor/measurement full scale pressure range (shown on screen) against the range of the system being measured. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 53...
  • Page 68: Functions

    Figure 6-3: Channel 1 Functions Figure 6-4: Channel 2 Functions Figure 6-5 is a channel function combination matrix. This shows the combinations of channel selections that are available in the DPI610E product range. Copyright 2022 Baker Hughes Company. 54 | DPI610E Instruction Manual–English...
  • Page 69: None

    100 bar / 1500 psi / 10 MPa G or A 135 bar / 2000 psi / 13.5 MPa G or A 200 bar / 3000 psi / 20 MPa G or A Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 55...
  • Page 70 Select the wanted channel (Channel 1 in this example). Select FUNCTION in the Channel Setup screen. Select Pressure in the Channel Function screen. Select INT (for Internal). Select Normal or Pseudo. Select to make the selections. Copyright 2022 Baker Hughes Company. 56 | DPI610E Instruction Manual–English...
  • Page 71 Pressure Release Valve fully or make sure that the Test Port is open to atmosphere. See Chapter 6.3.3.1 (INT) or Chapter 6.3.3.3 (EXT). Select the wanted pressure function (INT or EXT) from the CH Setup menu on CH1 or CH2. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 57...
  • Page 72 The hydraulic variants do not have the barometer as it is not usual to use gauge readings at the higher pressure at which these hydraulic variants operate. Thus, the hydraulic variants do not support ‘Pseudo-gauge’ or ‘Pseudo-abs’ ranges. Copyright 2022 Baker Hughes Company. 58 | DPI610E Instruction Manual–English...
  • Page 73 Make sure that the pressure port is open to atmosphere and start when the pressure reading is stable. Select to continue. Make sure that Sealed Gauge (SG) is correctly set on the selected channel. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 59...
  • Page 74: Sum

    Pressure function or EXT Pressure function cannot be set on CH2. Note: Be careful when both sensors are not gauge: make sure the effect of atmospheric pressure has been included. Copyright 2022 Baker Hughes Company. 60 | DPI610E Instruction Manual–English...
  • Page 75 CH Setup screen again. Make sure that the screen shows the wanted setup in the CH Setup menu. Select to go back to the Calibrator Main screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 61...
  • Page 76: Difference

    Select Difference from the FUNCTION menu. Select INT-EXT or EXT- INT as wanted. Select to make the selection. Make sure that the screen shows the wanted setup in the CH Setup menu. Select to go back to the Calibrator Main screen. Copyright 2022 Baker Hughes Company. 62 | DPI610E Instruction Manual–English...
  • Page 77: Barometer

    DPI610E. 6.3.6.1 To Set a Pressure Measurement Reading using the Barometer Function: Select the icon as wanted. Select FUNCTION from the Channel Setup menu to show the CHL Function screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 63...
  • Page 78: Observed

    6.3.7.1 To use the Observed function to set a channel: Select the as wanted. Select FUNCTION from the Channel Setup menu. Copyright 2022 Baker Hughes Company. 64 | DPI610E Instruction Manual–English...
  • Page 79: Rtd

    24 V power supply given by DPI610E. You can also use an external power supply for the device under test. 6.3.9.1 Current Measure The DPI610E current measurement range is: +/- 55 mA. You must use the correct terminals to set the Current Measure function: Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 65...
  • Page 80 Select Current > Measure > then select one of the following: 10 V for measure with internal 10 V Loop Power supply (only for non-IS units) 24 V for measure with internal 24 V Loop Power supply. Copyright 2022 Baker Hughes Company. 66 | DPI610E Instruction Manual–English...
  • Page 81 Step 4). Use the diagram on the right to make the necessary connections to the internal 10 V or 24 V supply. This is the source of Current Supply. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 67...
  • Page 82 The DPI610E can supply electrical current accurately within the 0 to 24 mA range. You can directly input the current value to be sourced using the Touchscreen or the Navigation Pad: Copyright 2022 Baker Hughes Company. 68 | DPI610E Instruction Manual–English...
  • Page 83: Current Source Automation Options

    The other methods to set the Current source output are: Nudge (manual) Span Check (manual or automated) Percent Step (manual or automated) Defined Step (manual or automated) Ramp (automated) Use the Current Source function to make these options available. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 69...
  • Page 84: Nudge

    Note: Different automation options will have different parameters. 6.3.11 NUDGE This is the Source Automation option that is automatically selected. It lets the source value be increased or decreased by a set step size value. Copyright 2022 Baker Hughes Company. 70 | DPI610E Instruction Manual–English...
  • Page 85: Span Check

    The SPAN CHECK function gives a 2-point span check. LOW (minimum) and HIGH (maximum) span values can be set. The DWELL time is the interval to wait at each Span Point before the change to the other span point. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 71...
  • Page 86 DOWN nudge buttons to switch between the LOW and HIGH source values. An option is available to automate the Span Check: tap the automation Play Softkey. If using the Navigation Pad: Copyright 2022 Baker Hughes Company. 72 | DPI610E Instruction Manual–English...
  • Page 87: Percent Step

    • AUTO-REPEAT - this tick box option lets the automation cycle be repeated continuously until manually stopped. SCALING - Refer to Section 6.4.6 on page 90. • Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 73...
  • Page 88: Defined Step

    The DEFINED STEP automation lets a defined step size within the span limits be set.The LOW, HIGH and STEP SIZE value set the spam limits. There is the option to auto-repeat the automation process. Copyright 2022 Baker Hughes Company. 74 | DPI610E Instruction Manual–English...
  • Page 89 Softkey to use the selected automation option and its automation settings. When the Defined Step function is selected, the screen shows its icon in the top right of the CH2 window in the Calibrator screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 75...
  • Page 90: Ramp

    Tap or select the parameter value to be changed. Use the Touchscreen or Navigation Pad to change the value. Select the Tick icon to set the new values. Select the Proceed icon to operate the option and go back to the Calibrator screen. Copyright 2022 Baker Hughes Company. 76 | DPI610E Instruction Manual–English...
  • Page 91: Voltage

    Using the Voltage function makes available the option to use the internal 10 V or 24 V power supply from the DPI610E. The measurement of the voltage input from an external power source, is also available. The diagram shows terminal connections for measurement of voltage. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 77...
  • Page 92 10 V for measurement with internal 10 V loop power supply. Select the Tick icon to make the selection and set the instrument. Make sure that the Voltage mode is active. Select the Back icon to show the Calibration screen. Copyright 2022 Baker Hughes Company. 78 | DPI610E Instruction Manual–English...
  • Page 93: Millivolts Measure - Setup

    DPI610E internal 10 V or 24 V power supply or to measure the millivolts input from an external power source. The diagram shows terminal connections for measurement of millivolts. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 79...
  • Page 94: Hart

    The DPI610E can use the HART (Highway Addressable Remote Transducer) communication protocol. Basic HART operation and setup on HART supported devices can be done. The HART bi-directional communications technology operates as a master/slave protocol. When the Copyright 2022 Baker Hughes Company. 80 | DPI610E Instruction Manual–English...
  • Page 95 Select None in the FUNCTION column and tap the page down Softkey to look at the second page of CH2 functions. Tap HART on the second screen to see two more selection columns. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 81...
  • Page 96: Process Options

    Tare symbol in the related channel window. Note: Tare is only available for most Measure functions but not for Barometer and HART. To select the Tare function: Copyright 2022 Baker Hughes Company. 82 | DPI610E Instruction Manual–English...
  • Page 97: Min/Max/Mean

    Its value is continuously shown as the reading values change, in addition to the live primary reading. When the function is on, the screen shows the Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 83...
  • Page 98 Make sure that the screen shows the Min/Max/Mean icon as the PROCESS option. This shows that the Min/Max/Mean is in operation in the Channel Setup screen. Select the Back icon. Copyright 2022 Baker Hughes Company. 84 | DPI610E Instruction Manual–English...
  • Page 99: Filter

    The Filter function can be used on most functions but not HART. When the Filter process option is on, the screen shows the Filter status icon in the related channel. To make available the Filter function: Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 85...
  • Page 100 PROCESS row. This shows that FILTER is on in the Channel Setup menu (as the PROCESS Option). Select the Back icon to go back to the Calibration Main screen. Copyright 2022 Baker Hughes Company. 86 | DPI610E Instruction Manual–English...
  • Page 101: Flow

    PROCESS row. This shows that FLOW is in operation in the Channel Setup menu (as the PROCESS Option.) Select the Back icon to go back to the Calibration Main screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 87...
  • Page 102: Alarm

    The user Alarm option is available with all Measure functions except for Barometer and HART. To select and do the setup of the user Alarm option: Select the icon as the wanted channel. Select PROCESS. Copyright 2022 Baker Hughes Company. 88 | DPI610E Instruction Manual–English...
  • Page 103 Push the Tick Softkey to set the alarm settings and go back to the PROCESS options screen. Note: To cancel the Alarm, select the ALARM checkbox to remove the Tick mark. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 89...
  • Page 104: Scaling

    INFORMATION The Scaling option is available with most measure and source functions but not for Observed and HART. Copyright 2022 Baker Hughes Company. 90 | DPI610E Instruction Manual–English...
  • Page 105 UNIT LABEL - a free text field where the special unit can be named. It is limited to a maximum of six characters. This special unit is automatically given the value of ‘%’. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 91...
  • Page 106 The screen shows the Scaling icon in the related channel window. The UNITS field shows the set special label. In addition, the full-scale value will be shown as its equivalent in the special unit. Copyright 2022 Baker Hughes Company. 92 | DPI610E Instruction Manual–English...
  • Page 107 Select the Setup Softkey to see or change the Scaling parameters which set the special unit. For information for how to set the parameters, see Section 6.4.6.1 for Method 1. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 93...
  • Page 108 Chapter 6. Calibrator Tasks Copyright 2022 Baker Hughes Company. 94 | DPI610E Instruction Manual–English...
  • Page 109: Utilities

    Touchscreen or push the Navigation Pad OK button to start the Leak Test utility. Note: If no compatible function is set on CH1, the Internal Pressure function will be selected for the Leak Test. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 95...
  • Page 110 Note: the screen will only show the icon for the use of an absolute gauge sensor. After the Leak Test times are set, use the DPI610E pump to pressurize the system to the necessary pressure. Copyright 2022 Baker Hughes Company. 96 | DPI610E Instruction Manual–English...
  • Page 111 System,” on page 253.) Only the list of test result files can be looked at on the device. Data about the files can only be seen when the files are opened on a PC. See Section 10.6.2, “To View Data Log files on a PC,” on page 162. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 97...
  • Page 112: Switch Test

    Tap on the Tasks icon on the Dashboard to select the menu. Tap the Switch Test two times on the Touchscreen or tap the OK button to start the utility. Copyright 2022 Baker Hughes Company. 98 | DPI610E Instruction Manual–English...
  • Page 113 (NO) switch in the test window. If a closed circuit is sensed, the switch will be identified as Normally closed (NC). Fully close the Pressure Release Valve. Make sure that there are no leaks. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 99...
  • Page 114: Tx (Transmitter Simulation) Simulator

    From the Tasks menu, select Tx Sim from the Task list. Tap the Tx Sim option again on the Touchscreen or push the Navigation Pad OK button to start the utility. Copyright 2022 Baker Hughes Company. 100 | DPI610E Instruction Manual–English...
  • Page 115 There is also the option for the DPI610E to supply loop power (10 V or 24 V). The usual setting is Off (no power supplied by the DPI610E). Push the Back Softkey. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 101...
  • Page 116: Relief Valve Test

    Select the Tick Softkey to set and go back to the Calibrator screen. The screen will show the related Relief Valve type icon in the channel window under the Function name field. Copyright 2022 Baker Hughes Company. 102 | DPI610E Instruction Manual–English...
  • Page 117 The automatic result file name will be the DPI610E current date and time stamp. This can be changed if necessary. Select the Tick Softkey to set the Result file name and complete the Save operation. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 103...
  • Page 118 Section 15.8 on page 256.) Only the list of test result files can be looked at on the device. Data about the files can only be seen when the files are opened on a PC (See Section 10.6.2, “To View Data Log files on a PC,” on page 162). Copyright 2022 Baker Hughes Company. 104 | DPI610E Instruction Manual–English...
  • Page 119: The Dpi610E-A Instrument

    The DPI610E-A menu has three options: None, Altitude and Airspeed. Select None when the instrument only has to show readings on its display. This chapter gives the procedures for the Altitude and Airspeed functions. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 105...
  • Page 120: Controls And Connection

    (refer to Section 8.1 on page 105). Make sure that all the necessary connections between the instrument and the system to be tested are safe. Copyright 2022 Baker Hughes Company. 106 | DPI610E Instruction Manual–English...
  • Page 121 Select the CH1 Channel, tap on the screen icon or use the related Softkey. Note: If the Aero application has already been used, the last saved channel configuration will be shown. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 107...
  • Page 122 10. Tap in the field that has the different measurement unit and then tap the Tick icon. This will change the measurement unit and show the CH1 Setup screen again. Copyright 2022 Baker Hughes Company. 108 | DPI610E Instruction Manual–English...
  • Page 123 The screen shows the Setup icon. Select this icon to show the Leak Test Setup screen. The format for the Wait time and Test time is: HH:MM:SS (Hours, Minutes, Seconds). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 109...
  • Page 124 16. Tap the Play icon to start the Leak Test. (This icon will change to a STOP icon after selection). Copyright 2022 Baker Hughes Company. 110 | DPI610E Instruction Manual–English...
  • Page 125: Altitude Switch Test

    Pressure switches can use two types of contact: Normally Open (NO) or Normally Closed (NC). Different procedures are necessary for switches whose electrical contacts are accessible and those that are not. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 111...
  • Page 126: How To Do An Altitude Switch Test (Accessible Switch Contacts)

    Make sure that the Let-down Valve is closed (turn the valve fully clockwise). Turn the Pressure/Vacuum Selector fully clockwise to the vacuum setting. Turn the Release Valve fully clockwise to close the valve. Copyright 2022 Baker Hughes Company. 112 | DPI610E Instruction Manual–English...
  • Page 127 CH1 Setup screen for the selected test. If necessary, make changes to the channel options on this setup screen. Tap the Back icon to continue. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 113...
  • Page 128 11. At the end of the test the Actuation and De-actuation and Hysteresis value are shown. Select the Save Softkey to save the test data if wanted. There is an option to do the test again: select the Restart icon. Copyright 2022 Baker Hughes Company. 114 | DPI610E Instruction Manual–English...
  • Page 129: How To Do An Altitude Switch Test (Non-Accessible Switch Contacts)

    Make sure that the Let-down Valve and the Release Valve are closed (turn both valves fully clockwise). Turn the Pressure/Vacuum Selector fully clockwise to the vacuum setting. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 115...
  • Page 130 Operate the pump until the switch operates, then tap the Valve Open icon (Actuation) for a Manual (NC) switch test. If a Manual (NO) switch test is selected, tap the Switch Close icon (Actuation). Copyright 2022 Baker Hughes Company. 116 | DPI610E Instruction Manual–English...
  • Page 131 Softkey to save the test data if wanted. An option is available to do the test again: tap the Restart icon. 10. Slowly turn the Let-down Valve fully counterclockwise to open the valve. The system goes to ground level pressure. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 117...
  • Page 132: Airspeed Leak Test

    Make sure that all the necessary connections between the instrument and the system to be tested are safe. Copyright 2022 Baker Hughes Company. 118 | DPI610E Instruction Manual–English...
  • Page 133 From the new *screen, select the CH1 Channel. In this example, by a tap on the screen icon or the use of the related Softkey. This shows the CH1 Setup screen*. *The screen can be different from that shown, if the instrument has been used. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 119...
  • Page 134 Function name field. Tap in the CH1 window to maximize the window and show the Setup icon and Play icon. Tap the Setup icon to show the Leak Test Setup screen. Copyright 2022 Baker Hughes Company. 120 | DPI610E Instruction Manual–English...
  • Page 135 Note: If a WAIT TIME has been set, a ‘Waiting’ countdown will start from the WAIT TIME value down to zero. This gives enough time for the pressure to become stable.The test starts Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 121...
  • Page 136: Airspeed Switch Test

    Electrical connections are shown by Figure 8.6.1 on page 118. The switch contacts must be voltage free. Copyright 2022 Baker Hughes Company. 122 | DPI610E Instruction Manual–English...
  • Page 137 To increase or decrease the shown accuracy of a measurement, tap on the RESOLUTION row. Select the wanted number of digits (4 to 7 digits) in the shown selection screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 123...
  • Page 138 To stop the test completely, tap the Stop icon. Slowly open the Release Valve and let the airspeed pressure decrease until the switch operates again. Copyright 2022 Baker Hughes Company. 124 | DPI610E Instruction Manual–English...
  • Page 139: Method (Non-Accessible Switch Contacts)

    When the airspeed switch contacts cannot connect to the instrument, use an external indicator or annunciator to show the start of the airspeed switch operation. The user, in response to this switch operation, makes the user interface react. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 125...
  • Page 140 Manual (NO) - Select to test a Normally Open switch. • The screen will have a different icon for each type of switch. Tap the Tick icon to show the next screen. Copyright 2022 Baker Hughes Company. 126 | DPI610E Instruction Manual–English...
  • Page 141 Slowly operate the pump handle until the switch operates, then tap the Valve Open icon (Actuation) for a Manual (NC) valve test. If a Manual (NO) valve test is selected, tap the Valve Close icon (Actuation). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 127...
  • Page 142 If the results are saved, the test details will be erased, and the test will be available to be done again. If the results are not saved, a new test cycle can be set up, ready to operate. Copyright 2022 Baker Hughes Company. 128 | DPI610E Instruction Manual–English...
  • Page 143 Airspeed Switch Test 10. Slowly turn the Let-down Valve counterclockwise to make the system pressure go to ground level. Continue to turn the Let-down Valve fully counterclockwise to open the valve. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 129...
  • Page 144 Chapter 8. The DPI610E-A Instrument Copyright 2022 Baker Hughes Company. 130 | DPI610E Instruction Manual–English...
  • Page 145: External Sensors

    To complete the connection, rotate the locking collar until hand-tight. Align the male end of the cable to the port on the DPI610E and use the same method to lock the cable in position. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 131...
  • Page 146: The Zero Function

    Function” on page 57 for more information about the use of this function. 9.1.5 Available External Pressure Sensors Only the external pressure sensors in Table 9-2 are compatible with the DPI610E type of instrument. Copyright 2022 Baker Hughes Company. 132 | DPI610E Instruction Manual–English...
  • Page 147: How To Set An External Pressure Sensor

    1000 bar / 15 000 psi / 100 MPa 1400 bar / 20 000 psi / 140 MPa 9.1.6 How to set an External Pressure Sensor Select the wanted channel (CH1 in this example). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 133...
  • Page 148 DPI610E. All PM700E sensors etched with a part number ending in “-3” have the correct firmware installed for full supported use on the DPI610E. Copyright 2022 Baker Hughes Company. 134 | DPI610E Instruction Manual–English...
  • Page 149 DPI610E is already in operation. Also, a message text “CAL DUE” will permanently be in the status bar until the sensor is disconnected or the sensor has been calibrated. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 135...
  • Page 150: Rtd Probe And Interface

    Use the supplied 2.9m (9.5 feet) cable to connect the RTD-PROBE/RTD-INTERFACE to the DPI610E. The RTD-INTERFACE is supplied with a field-rewireable M12 connector, that lets connections be done by the use 2,3 or 4-wire RTDs. Copyright 2022 Baker Hughes Company. 136 | DPI610E Instruction Manual–English...
  • Page 151: Temperature Considerations

    To complete the connection, turn the swivel connector fully until hand-tight. Align the male end of the cable to the port on the DPI610E and use the same method to tighten into position. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 137...
  • Page 152 “- - - - -“. If the external sensor is disconnected while the function is in use, the screen can show an error message. After successful connection, the full-scale value of the RTD probe is shown and the RTD probe type. Copyright 2022 Baker Hughes Company. 138 | DPI610E Instruction Manual–English...
  • Page 153: Setup Of A Channel For A Rtd Sensor

    Select RTD and select the Tick icon to make the selections and show the Channel Setup screen. To set the RTD measurement as resistance or temperature, select the RTD Mode option from the CH Setup screen and push the Enter key in the Navigation Pad. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 139...
  • Page 154: Rtd Profiles

    Tap on the wanted profile to select it. Tap on the View icon to look at or change the coefficients and parameters in the profile. An alternative method is to use the UP/DOWN Copyright 2022 Baker Hughes Company. 140 | DPI610E Instruction Manual–English...
  • Page 155 Coefficient 3 (c, C, Delta) Third coefficient value Min. Temperature Minimum temperature Max. Temperature Maximum temperature When all parameter fields are filled, use the Back Softkey to go back to the RTD Profile screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 141...
  • Page 156: The Adroit Sensor

    Voltage or Current (4 to 20 mA). To calibrate ADROIT sensors without the DPI610E, it is necessary to use the Druck ADROIT PC application, together with the ADROIT interface box and an applicable calibrated pressure source. Copyright 2022 Baker Hughes Company. 142 | DPI610E Instruction Manual–English...
  • Page 157: Setup Of An Adroit Sensor

    CH2 setup menu. The CONFIGURATION option will then appear on the CH2 Setup screen. Select this option to show the CH2 Sensor Configuration screen: the default selection Configuration option is Standard. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 143...
  • Page 158 Select the Back button on the CH2 Setup screen to show the Calibrator screen. Make sure that the ADROIT sensor is correctly connected to the pressure port on the DPI610E. Copyright 2022 Baker Hughes Company. 144 | DPI610E Instruction Manual–English...
  • Page 159: Adroit Calibration

    On the Dashboard push the Settings Softkey and then select the ADVANCED menu option. Enter the correct PIN to get access to the ADVANCED menu (for more information, refer to “ADVANCED” on page 40). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 145...
  • Page 160: Perform Calibration

    Select the PERFORM CALIBRATION option to get to the calibration options. There are two types of adjustments available: Full Adjustment - this is a 2-point calibration adjustment of the sensor through the zero • (or minimum full-scale) and the positive full-scale pressure range. Copyright 2022 Baker Hughes Company. 146 | DPI610E Instruction Manual–English...
  • Page 161: Full Adjustment Calibration

    At Calibration point 2, apply the full-scale pressure of the sensor. Use the DPI610E pump and/or Volume Adjuster to apply pressure. When the pressure reading is stable, push the Next Softkey to continue. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 147...
  • Page 162 Calibration Point 2. The system calculates the Zero Offset and Span Adjustment values from these shown values. If these values are not acceptable, select Cancel to go back to the calibration procedure. Copyright 2022 Baker Hughes Company. 148 | DPI610E Instruction Manual–English...
  • Page 163: Zero Adjustment

    Refer to page 157 for more information about the contents of this screen. Change the sensor data on the screen if necessary. Select the Tick icon to go to the next step of the calibration procedure. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 149...
  • Page 164 RESTART Softkey lets the calibration procedure be started again if a repeat calibration is necessary. BACK Softkey stops the calibration procedure and takes the user to the Perform • Calibration menu screen. Copyright 2022 Baker Hughes Company. 150 | DPI610E Instruction Manual–English...
  • Page 165: View Sensor Status

    Select RESTORE FACTORY CALIBRATION from the Adroit Calibration screen. Select OK on the popup box to use the restore operation: this operation can take several seconds before it shows the screen as shown. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 151...
  • Page 166 Chapter 9. External Sensors Copyright 2022 Baker Hughes Company. 152 | DPI610E Instruction Manual–English...
  • Page 167: Data Log

    TIME INTERVAL, LOG DURATION and DATA POINTS are only listed in the Data Log Setup menu when TRIGGER is set to PERIODIC (See Section 10.2 on page 155). 10.1.1 How to Set the Data Log Filename To set the Data Log filename: Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 153...
  • Page 168 Note: Only a maximum of 20 characters and symbols are available. Select to set the new file name. Make sure that the new file name is in the Data Log Setup screen. Copyright 2022 Baker Hughes Company. 154 | DPI610E Instruction Manual–English...
  • Page 169: Trigger Menu

    Functions set in the Calibrator channels have an effect on the minimum interval available. See table below. Function Minimum Interval (HH:MM:SS.S) Internal Barometer 00:00:05.0 Internal Pressure 00:00:00.2 External Sensor (Pressure and RTD) 00:00:00.2 Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 155...
  • Page 170: Log Duration

    Its format is HH:MM:SS. The supported time interval range is between 00:00:01 to 999:59:59. Note: The time interval value must always be less than the LOG DURATION value. Copyright 2022 Baker Hughes Company. 156 | DPI610E Instruction Manual–English...
  • Page 171: Data Points

    6 to 5, the log duration will automatically adjust to 50 seconds (000:00:50) based on the unchanged time interval of 10 seconds and the new number of selected data points. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 157...
  • Page 172: Setting Up Manual Data Logging

    The KEY PRESS option (see Section 10.2 on page 155) is a manual data trigger. Use the following steps to continue with a KEY PRESS triggered Data Log. Select the Proceed icon in the Data Log Setup screen. Copyright 2022 Baker Hughes Company. 158 | DPI610E Instruction Manual–English...
  • Page 173 The screen shows a popup window. Select the OK button to fully stop the Data Logging. The screen will then show a message that the data log file has been saved. Select the Cancel button to continue the data log logging. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 159...
  • Page 174: How To Do Periodic Data Logging

    After the selection of the PERIODIC Data Log mode and the values of the PERIODIC options, select the Proceed icon in the Data Log Setup screen. The display then shows the Calibrator primary screen. Copyright 2022 Baker Hughes Company. 160 | DPI610E Instruction Manual–English...
  • Page 175: Viewing & Deleting Data Log Files

    Dashboard. 10.6 Viewing & Deleting Data Log Files 10.6.1 To View Data Log Files on the Instrument Select FILES in the Data Log screen. Select the wanted Data Log file. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 161...
  • Page 176 DPI610E device to the PC. (See Section 10.7 on page 164). On the PC, the Windows File Explorer shows the DPI610E memory as a mass storage device or drive. Select the DPI610E drive and select the DataLog folder from the root directory. Copyright 2022 Baker Hughes Company. 162 | DPI610E Instruction Manual–English...
  • Page 177 10.6.3.1 To Erase a Single Data Log file Select FILES from the Data Log screen. Select the Data Log file. Select the ‘single’ trash icon on the Data Log File screen. Select OK to erase the file. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 163...
  • Page 178 This device can either be a micro-USB memory stick or an external PC. Select the Settings icon on the Dashboard. Push the Home button if necessary, to show the Dashboard. Copyright 2022 Baker Hughes Company. 164 | DPI610E Instruction Manual–English...
  • Page 179 If a copy of the Data Log file is to be put into the memory of an external PC, connect a micro- USB data cable to the USB port. Select MASS STORAGE, then select to make the selection. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 165...
  • Page 180 DPI610E, which will be shown as a Mass Storage drive (default name is DPI610E). Note: USB access must be available for the PC to use this function. When the copy operation is complete, remove the cable from the DPI610E. Copyright 2022 Baker Hughes Company. 166 | DPI610E Instruction Manual–English...
  • Page 181 4 to 20 mA output, CH1 can be set to INT Pressure and CH2 to Current Measure (with 10/24 V power if wanted). Select the wanted types of measurement units. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 167...
  • Page 182: Error Type

    This is reliant on the Error type selected. The default tolerance value is 0.1%. 11.4 Analysis Function Set the Input and Output channel and error analysis parameters. Refer to “Analysis Application” on page 167 for more information. Copyright 2022 Baker Hughes Company. 168 | DPI610E Instruction Manual–English...
  • Page 183 When the test is complete, tap the Back icon to go from the Datalog and Analysis application. Use the Data Log application to get access to Datalog files (Refer to “Data Log” on page 153). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 169...
  • Page 184 Chapter 11. Analysis Copyright 2022 Baker Hughes Company. 170 | DPI610E Instruction Manual–English...
  • Page 185 A calibration certificate template wizard is supplied by Druck. This puts the data into an applicable form for use in printing or filing. The related Excel Macro file is available on: Druck.com/DPI610E as “Druck DPI610E Calibration Template”. 12.1.1 How to Start the Documenting Application Select the Documenting Softkey from the Dashboard.
  • Page 186 Note: A maximum of 20 characters is available for the file name. Select the Tick Softkey to set and show the Create Procedure screen. Copyright 2022 Baker Hughes Company. 172 | DPI610E Instruction Manual–English...
  • Page 187 The input range is automatically filled in with the full sensor range related to the Function. Only the range can be changed and must be in the full sensor range. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 173...
  • Page 188 Softkey can be used to interchange the Input and Output functions: to use the original Input function as the Output function and the original Output function as the Input function. Copyright 2022 Baker Hughes Company. 174 | DPI610E Instruction Manual–English...
  • Page 189 Note: There must be a minimum of 2 points UP and a maximum of 25 test points in total (all UP and DOWN points). For each Points UP and DOWN setting, the test point values are calculated and can be seen by the selection of the View Softkey. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 175...
  • Page 190 TOLERANCE – this is the maximum deviation or permitted error margin for each input test point in the calibration. It is specified as a percentage of the Input range. The default test point tolerance value is 5% and can be changed if necessary. Copyright 2022 Baker Hughes Company. 176 | DPI610E Instruction Manual–English...
  • Page 191 (FAIL). It is measured as a percentage of the output. It can also be in fixed measurement units dependent on the tolerance type. The default value is 0.1% (based on ‘% FS’). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 177...
  • Page 192 Options include: • %Fullscale (Percent of Full-scale) • %Span (Percent of Span) • % Reading (Percent of Reading) • %Fixed (Measurement units). See Section 11.3.3 on page 168 for more details. Copyright 2022 Baker Hughes Company. 178 | DPI610E Instruction Manual–English...
  • Page 193 The test procedure can also be selected from the Internal Procedure Files screen. To select a test procedure, for example, File 2, tap in the row or use the Navigation Pad. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 179...
  • Page 194 SERIAL NUMBER (Optional) – the serial number of the asset or DUT. Leave blank if not known. Maximum number of characters: 50. • MANUFACTURER (Optional) – the manufacturer of the asset or DUT. Maximum number of characters: 30. Copyright 2022 Baker Hughes Company. 180 | DPI610E Instruction Manual–English...
  • Page 195 23 to 36 in. Hg). AMBIENT HUMIDITY – Enter the ambient humidity value where the test is done. The • default value is 70%. Values entered must be between 0 and 100%bar. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 181...
  • Page 196 Select the OK button to remove the message. This gives a checkpoint where the user can go back and check all the data of the test procedure are correct. Copyright 2022 Baker Hughes Company. 182 | DPI610E Instruction Manual–English...
  • Page 197 12.3 The Documenting Main Screen This section gives information about the different parts of the Documenting Main screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 183...
  • Page 198 Note: If the marker is not shown, look at the error value, as it is likely that the error is too large to be shown by the error and tolerance value. Copyright 2022 Baker Hughes Company. 184 | DPI610E Instruction Manual–English...
  • Page 199 To use PSEUDO RANGE, select the related pseudo-range option from the settings screen. The tick-box has the tick mark when in operation and has no tick mark when not in operation. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 185...
  • Page 200 The screen shows a popup message window that shows the Pass/Fail status of both the input (test point) and the output (result). Select OK to accept and save the reading or Cancel to reject the reading. Copyright 2022 Baker Hughes Company. 186 | DPI610E Instruction Manual–English...
  • Page 201 50. Select the Save Softkey to save the CALIBRATION REMARKS and continue. The next screen is the procedure result Summary screen. This screen gives data about the completed test procedure. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 187...
  • Page 202 Softkey, to do a check on the adjustment, to make sure it is correct. Make the necessary adjustment to the DEVICE UNDER TEST (DUT). Examine the output signal through its full calibration range. Do this to make sure it is in specified limits. Copyright 2022 Baker Hughes Company. 188 | DPI610E Instruction Manual–English...
  • Page 203 Softkey from this menu screen. To look at As-Found or/and As-Left calibration results, select the wanted Asset file name. Tap the file name to select it and tap a second time to open. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 189...
  • Page 204 To see Calibration remarks that relate to the test, select the Clipboard Softkey. Results can also be looked at when, the test procedure is selected, the device under test is selected and then the related test result (As-Found or As-Left). Copyright 2022 Baker Hughes Company. 190 | DPI610E Instruction Manual–English...
  • Page 205 This template uses the calibration procedure result data as the contents for a formatted calibration certificate. Make a copy of the certificate template and save it by the use of the given URL or QR code. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 191...
  • Page 206 To download test procedures from 4Sight2 on to the DPI610E, use the supplied USB data cable to connect the instrument to the PC. Note: This function is not available for use with the DPI610E-A (Aero). Copyright 2022 Baker Hughes Company. 192 | DPI610E Instruction Manual–English...
  • Page 207 Assets or Work List items, please refer to the 4Sight2 Calibration Management Software User Manual 123M3138). From the 4Sight2 work list, select the calibration procedure(s) to be done. Select the Portables calibration type option and tap the Send button. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 193...
  • Page 208 (shown by its model and serial number). Make the Test Equipment profile: enter data into the data fields in the popup screen and tap the Create button to complete the procedure. Copyright 2022 Baker Hughes Company. 194 | DPI610E Instruction Manual–English...
  • Page 209 Select the Calibration Test Procedure(s) to be sent to the DPI610E and tap Send to Test Equipment. When the procedure is completed, a Success message will be in the Operation Status tab. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 195...
  • Page 210 Select the wanted Remote test procedures from the Remote Procedure Files screen. Use the icon to show the next page of files. Use the icon to set the order in which the files are listed. Copyright 2022 Baker Hughes Company. 196 | DPI610E Instruction Manual–English...
  • Page 211 See Section 12.2.2, “How to Make an Internal Procedure,” on page 172. Examine and change, if necessary, the environment data and user ID. Push the Next Softkey to go to the next step. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 197...
  • Page 212 Note: See Section 12.3, “The Documenting Main Screen,” on page 183 for more information about the test screen and Section 12.4, “Documenting Settings,” on page 185 for more information about available settings. Copyright 2022 Baker Hughes Company. 198 | DPI610E Instruction Manual–English...
  • Page 213 Note: When a test procedure is done on a new Asset for the first time, there will be no As- Found or As-Left results. When no As-Found or As-Left results are sensed for an asset, the calibration data will be automatically saved as, As-Found. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 199...
  • Page 214 END - the end-pressure value of the switch test. • REPEAT CYCLE - the number of switch test cycles to be completed in the test • procedure in a set sequence. Copyright 2022 Baker Hughes Company. 200 | DPI610E Instruction Manual–English...
  • Page 215 See Section 12.2.2, “How to Make an Internal Procedure,” on page 172. Examine and change, if necessary, the environment data and USER ID. Push the Next Softkey to go to the next step. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 201...
  • Page 216 Test data are correct and to remove this popup message. Use the Back Softkey to go back to the test information or use the Proceed Softkey to show the test screen. Copyright 2022 Baker Hughes Company. 202 | DPI610E Instruction Manual–English...
  • Page 217 Softkey will replace the Add Test Point Softkey. The title bar will show Procedure complete with the overall test result status. Select the Tick Softkey to fully complete the calibration procedure. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 203...
  • Page 218 Softkey to do this. The procedure will use the initial test data and device under test (DUT) data. To stop the use of the screen, use the Back Softkey to go back to the Remote Procedure menu. Copyright 2022 Baker Hughes Company. 204 | DPI610E Instruction Manual–English...
  • Page 219: How To View Test Results

    Results can also be looked at when the test procedure is selected and also the related test result, for example, As-Found or As-Left. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 205...
  • Page 220 If no icons are shown, it means no results are available for that test procedure. The first icon shows the As-Found result and the second icon shows the As- Left result. - FAIL INDICATION - ADJUST INDICATION Copyright 2022 Baker Hughes Company. 206 | DPI610E Instruction Manual–English...
  • Page 221 • Pass or Fail result for each test cycle. Note: For Switch Test with multiple switch cycles, use the Page Down Softkey to see the test results of each cycle. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 207...
  • Page 222 From the related Calibration screen in 4Sight2, select the wanted calibration and tap Receive from Test Equipment. The success of the copy operation will be shown by the “Success...” text in the Operation Status area. Copyright 2022 Baker Hughes Company. 208 | DPI610E Instruction Manual–English...
  • Page 223 To see the results, tap on the toggle icon in the left-hand section of the screen to change from Send/Receive to Review mode. Select the calibration file to show data of the test and its results. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 209...
  • Page 224 Chapter 12. Documenting For more information on how to analyze or use test data in 4Sight2, please refer to the 4Sight2 User Manual 123M3138. Copyright 2022 Baker Hughes Company. 210 | DPI610E Instruction Manual–English...
  • Page 225 From the Calibrator screen: select the channel 2 Setup (either tap the screen icon or push the Softkey). Note: The HART function is only available on CH2. Select FUNCTION from the CH2 Setup screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 211...
  • Page 226 Note: the 250 ohm HART resistor is important for data movement with the HART device to occur. It gives sufficient impedance in the signal loop for the HART signal to occur. Select the Back Softkey to go back to the Calibrator screen. Copyright 2022 Baker Hughes Company. 212 | DPI610E Instruction Manual–English...
  • Page 227 DPI610E shows information about the different electrical connections shown. See section 13.1.1 for setting the loop power. HART with internal 24 V loop power available. This is when the menu selection is: HART > Measure > 24V. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 213...
  • Page 228 HART > DIRECTION > Measure > Off. HART when connected to an external current loop. This is when the menu selection is: HART > DIRECTION > Master. 13.2 HART Device Configuration Copyright 2022 Baker Hughes Company. 214 | DPI610E Instruction Manual–English...
  • Page 229 Softkey and enter the wanted value. Push the Tick Softkey to accept and the Back Softkey to go back to the Device Poll screen. Note: An applicable Poll Address value must be between 0 and 63. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 215...
  • Page 230 It can show a maximum of four found devices. The scan operation can be stopped immediately if necessary, by the use of the Stop Softkey. Select the Search Softkey to start the scan again. Copyright 2022 Baker Hughes Company. 216 | DPI610E Instruction Manual–English...
  • Page 231 Note: The HART device must connect successfully to the DPI610E, for the next screen to show the HART Softkey. Refer to the previous section for instructions. From the Calibrator screen: select the HART channel icon (either tap the screen icon or push the Softkey.) Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 217...
  • Page 232 The PREAMBLES data relates to the code in a file header: this identifies the start and end of a data packet. Note: this is a read-only screen. Copyright 2022 Baker Hughes Company. 218 | DPI610E Instruction Manual–English...
  • Page 233 To change the device data, select the option and enter the value or text. Select the Tick Softkey to make this new change to the device. The options include: • Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 219...
  • Page 234 CH1 Primary - Main reading on CH1 • CH2 MA - Loop current measured by the DPI610E 13.3.4 Signal Condition Figure 13-4: Signal Condition screen This screen shows the following variables: Copyright 2022 Baker Hughes Company. 220 | DPI610E Instruction Manual–English...
  • Page 235 URL (Upper Range Limit) • Minimum Span Note: This is a read-only screen. 13.4 HART Service Methods Figure 13-6: Service Methods Screen The Service Methods available on the DPI610E are: Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 221...
  • Page 236 Enter the wanted output value and select the Tick Softkey. This screen shows the current measured by the DPI610E as the result of the forced current output from the HART device. Copyright 2022 Baker Hughes Company. 222 | DPI610E Instruction Manual–English...
  • Page 237 Softkey (Step 2) to change this value if wanted. If no change is wanted, select the Next Softkey to proceed (Step 3). Enter the wanted Low Trim value and select the Tick Softkey to enter this value. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 223...
  • Page 238 The default High Trim value is 20 mA. This value can be edited if wanted by the selection of the Edit Softkey (Step 6). If no change is necessary, select the Next Softkey to continue (Step 7). Copyright 2022 Baker Hughes Company. 224 | DPI610E Instruction Manual–English...
  • Page 239 The High Trim is complete. This completes this step of the D/A trim procedure. The final step of the procedure is to make sure that the D/A trim adjustment is sufficient. Select the Verification Softkey to test if the adjustment has been sufficient. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 225...
  • Page 240 HART device is shown by Select the Trim Softkey. Select OK in the popup message window to accept the Trim, or select Cancel to go back to the previous step. Copyright 2022 Baker Hughes Company. 226 | DPI610E Instruction Manual–English...
  • Page 241 Buffers Available Device Variable Priority too low Index Not Allowed Invalid Extended Command Not Command Number Implemented Invalid I/O Card Multiple Meanings 65-72 Number Error Invalid Channel Reserved Warning 96-111 Number Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 227...
  • Page 242 Chapter 13. HART Copyright 2022 Baker Hughes Company. 228 | DPI610E Instruction Manual–English...
  • Page 243: Instrument Calibration

    PIN is selected) and then select the Tick icon. In the Advanced Menu screen, select CALIBRATION. Refer to Chapter 5, “Advanced Menu” on page 41, for instructions for how to use the other Advanced Menu options. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 229...
  • Page 244: How To Do Calibration

    Source 24 mA Voltage Measure 20 V 30 V Source 10 V Millivolts Measure 200 mV 2000 mV Pressure Measure Dependent on sensor fitted Barometer Measure 750 to 1150 mbar Copyright 2022 Baker Hughes Company. 230 | DPI610E Instruction Manual–English...
  • Page 245 Section 6.3.9 on page 65 for Current Measure electrical connection diagram). Select PERFORM CALIBRATION from the Instrument Calibration menu. Select a FUNCTION option (for example, Current). Select a DIRECTION option if applicable (for example, Measure). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 231...
  • Page 246 Note: • All electrical measure functions must have a 3-point calibration. • Current Source function must have a 2-point calibration. • Voltage source function must have a 1-point calibration. Copyright 2022 Baker Hughes Company. 232 | DPI610E Instruction Manual–English...
  • Page 247 Select OK to apply the calibration adjustment. If wanted, to stop the procedure, select the Cancel button to go back to the Calibration screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 233...
  • Page 248 Test Port to the external calibrated pressure source. This sensor calibration must only be done by Service Centers and personnel that have the necessary approval. Select PERFORM CALIBRATION from the Instrument Calibration menu. Select Pressure function. Copyright 2022 Baker Hughes Company. 234 | DPI610E Instruction Manual–English...
  • Page 249 Enter REFERENCE VALUE 1 as shown on the external calibrated equipment. Push the Tick Softkey to enter the value and go back to the Calibration screen. Push the Next Softkey to complete Calibration point 1 and go to Calibration point 2. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 235...
  • Page 250 When verification is complete, select the Back icon to stop the calibration procedure. An option is to select the Restart Softkey to do the calibration again. Copyright 2022 Baker Hughes Company. 236 | DPI610E Instruction Manual–English...
  • Page 251 To start the calibration, apply nominal negative full-scale or zero pressure as instructed by the screen message. This is Calibration point 1. When the measured Pressure reading is stable, select the Next icon to continue. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 237...
  • Page 252 There are three methods available to continue, each with an icon. These are: VERIFICATION Softkey shows the Verification screen (see Step 9). RESTART Softkey lets the calibration procedure to be done again if a repeat calibration is necessary. Copyright 2022 Baker Hughes Company. 238 | DPI610E Instruction Manual–English...
  • Page 253 Figure 14-2: Instrument Status screen To get access to the Internal Pressure Sensor Status screen (Figure 14-2), select INTERNAL PRESSURE SENSOR STATUS from the Instrument Calibration screen as shown: Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 239...
  • Page 254 Figure 14-3: Instrument Calibration Date & Interval screen To get access to the Instrument Calibration Date & Interval screen (Figure 14-3), select SET CALIBRATION DATE & INTERVAL from the Instrument Calibration menu as shown: Copyright 2022 Baker Hughes Company. 240 | DPI610E Instruction Manual–English...
  • Page 255 Tap either the left or right value again and again to decrease or increase the selected value. Select to enter the changes. 14.4.2 How to change the Calibration Interval Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 241...
  • Page 256: Backup Calibration

    Select OK to make a backup. Make sure that the screen shows the “Calibration backup successful” message. If the screen does not show this message, do steps 1 and 2 again. Copyright 2022 Baker Hughes Company. 242 | DPI610E Instruction Manual–English...
  • Page 257: Restore Factory Calibration

    Make sure that the screen shows the FACTORY CALIBRATION RESTORE successful message. 14.8 EXTERNAL PRESSURE SENSOR CALIBRATION MENU See Section 14.1 on page 229 for how to get access to the Calibration menu from the Dashboard. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 243...
  • Page 258 Select PERFORM CALIBRATION from the External Sensor Calibration menu screen. The calibration procedure of the external PM700E sensor is like that of the DPI610 internal pressure sensor. See Section 14.2.2, “Calibration - Internal Pressure Sensor,” on page 234. Copyright 2022 Baker Hughes Company. 244 | DPI610E Instruction Manual–English...
  • Page 259 Make the user notification available. Set a date for the next calibration. The date that is automatically used, is calculated by the use of the last calibration date and the specified calibration interval. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 245...
  • Page 260 Tap either the left or right value again and again to decrease or increase the selected value. Copyright 2022 Baker Hughes Company. 246 | DPI610E Instruction Manual–English...
  • Page 261 Refer to Section 9 on page 131 for how to set the DPI610E to recognize and use an external RTD sensor. This is necessary to make the DPI610E user interface show the calibration options for the RTD sensor. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 247...
  • Page 262 User instructions for each step of the procedure. A screen message will give an instruction to start the calibration. Apply nominal negative full- scale or zero pressure. This is Calibration point 1. Copyright 2022 Baker Hughes Company. 248 | DPI610E Instruction Manual–English...
  • Page 263 This adjustment uses the same calibration points used in the procedure. Select OK to do the calibration adjustment. If the procedure is to be stopped, select the Cancel button to go back to the Calibration screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 249...
  • Page 264 (default date uses the date of the last calibration plus the specified calibration interval). This is Read Only - this variable cannot be changed on this screen. Copyright 2022 Baker Hughes Company. 250 | DPI610E Instruction Manual–English...
  • Page 265 Tap either the left or right value again and again to decrease or increase the selected value. Select to confirm the changes. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 251...
  • Page 266 Select to confirm the changes. Tap either the left or right value again and again to decrease or increase the selected value. Select to confirm the changes. Copyright 2022 Baker Hughes Company. 252 | DPI610E Instruction Manual–English...
  • Page 267 Factory Calibration – this is the default calibration done on the instrument in the factory • before it is sent. Its values saved in permanent storage and cannot be changed or removed by the user. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 253...
  • Page 268 Documenting function (see Chapter 12 on page 171). To look at the list of available internal procedures, select the Local (Procedures) sub-folder. Select the Remote sub-folder to look at a list of 4Sight2 procedures in memory. Copyright 2022 Baker Hughes Company. 254 | DPI610E Instruction Manual–English...
  • Page 269 These files are read-only: open the file on a PC to look at test result data. Use either the Single Delete Softkey (to erase individual files) or the Delete All Softkey (to erase all files in the sub-folder). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 255...
  • Page 270: Relief Valve

    Softkey, then select USB: change the USB setting to ‘Storage’ if not set. This lets the DPI610E device be available as a mass storage drive, identified as ‘DPI610E’ in File Explorer on the PC. Copyright 2022 Baker Hughes Company. 256 | DPI610E Instruction Manual–English...
  • Page 271 Refer to Section 16.8 on page 261 for information about how to use the Error Log function. Refer to Section 16.9 on page 262 for information about how to use the Event Log function. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 257...
  • Page 272 Chapter 15. Files System Copyright 2022 Baker Hughes Company. 258 | DPI610E Instruction Manual–English...
  • Page 273: Status Menu Options

    The screen shows Status menu options. To select the row that has the wanted option, tap on the row or use the Navigation Pad. 16.3 SOFTWARE BUILD In the Status menu screen select the SOFTWARE BUILD row, to show the Status - SOFTWARE BUILD screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 259...
  • Page 274 In the Status menu screen select the BATTERY row, to show the Status - Battery screen. The Status - Battery screen shows the instrument battery level and if the battery gives power (discharging) or receives power (being charged). Copyright 2022 Baker Hughes Company. 260 | DPI610E Instruction Manual–English...
  • Page 275: Error Log

    CSV file. The Error Log screen shows this export file but this file can only be opened if moved and looked at from a PC: (refer to Section 10.6.1 on page 161 for how to find and read these files). Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 261...
  • Page 276: Event Log

    Select the EVENT LOG from the Status menu screen. Note: The screen shows a ‘Please Wait’ popup message while the log file is made. It can take up to 1 minute to complete. Copyright 2022 Baker Hughes Company. 262 | DPI610E Instruction Manual–English...
  • Page 277 The screen shows a message window with the text, “Delete all files?”. Select OK or Cancel. Tap the down icon to show one more page of available files. Tap the Back icon to go back to the Status screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 263...
  • Page 278 Chapter 16. Status Menu Copyright 2022 Baker Hughes Company. 264 | DPI610E Instruction Manual–English...
  • Page 279: Favorites Menu

    Calibrator application. This example screen relates to a Calibrator task for a Leak Test and Current Measure. Select the Back icon on this screen to show the Dashboard screen. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 265...
  • Page 280 Softkey from the Dashboard. (For this screen image see Step 1 in Section 17.1 on page 265). Select the Add New Softkey from the Favorites screen. Select the CREATE NEW option. Copyright 2022 Baker Hughes Company. 266 | DPI610E Instruction Manual–English...
  • Page 281 17.3 Edit an Existing Favorite File From the Favorites menu, go to the wanted saved Favorites file: tap the file name to select the file or use the Up/Down button on the Navigation Pad. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 267...
  • Page 282 Saved Favorites setup files can be looked at in the File System menu by the selection of the File System Softkey from the Dashboard and the selection of the Favorites folder. Refer to Section 15.8 on page 256 for more information. Copyright 2022 Baker Hughes Company. 268 | DPI610E Instruction Manual–English...
  • Page 283: General Specification

    *Recommended wait time is 5 minutes. Longer or shorter wait times will affect the leak rates. 18.2 Open Source Software Licenses Two files are available with the software installation: 1568-notices-report-08_08_2022_17_04.txt covers DK0491 DPI610E Bootloader. 1563-notices-report-08_08_2022_16_01.txt covers DK0492 DPI610E Main Application. Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 269...
  • Page 284 Chapter 18. General Specification Copyright 2022 Baker Hughes Company. 270 | DPI610E Instruction Manual–English...
  • Page 285: Contact Details

    Contact details 19. Manufacturer 19.1 Contact details Druck Limited 2 Fir Tree Lane Groby Leicester LE6 0FH United Kingdom Tel: +44 (0)116 231 7100 www.Druck.com Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | 271...
  • Page 286 Chapter 19. Manufacturer Copyright 2022 Baker Hughes Company. 272 | DPI610E Instruction Manual–English...
  • Page 287: Appendix A. Compliance Statements

    Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes: l'appareil ne doit pas produire de brouillage; Copyright 2022 Baker Hughes Company. English–DPI610E Instruction Manual | i...
  • Page 288 RF peut être augmentée si l'appareil peut être conservé aussi loin que possible du corps de l'utilisateur ou que le dispositif est réglé sur la puissance de sortie la plus faible si une telle fonction est disponible. Copyright 2022 Baker Hughes Company. ii | DPI610E Instruction Manual–English...
  • Page 290: Office Locations