G2S BPAP Series User Manual 1. Symbols 1.1 Control Buttons Ramp Button Mute Button Knob 1.2 Device Symbols Follow Instructions for Use Consult Instructions for Use Type BF Applied Part (mask and SpO probe) Class II (Double Insulated) For indoor use only...
Page 43
G2S BPAP Series User Manual CE Marking 0123 Single Patient Multiple Use Batch Code Non-Ionizing Radiation SD Card Air Inlet Air Outlet Caution Marking of Electrical and Electronic Equipment Logo of BMC Medical Co., Ltd. Medical Device Unique Device Identifier...
Warnings, Cautions, and Important Tips appear throughout this manual as they apply. 3. Intended Use BPAP System (G2S BPAP Series) is a Bi-level PAP (Bi-level Positive Airway Pressure) device, which is intended to provide non-invasive ventilation for patients with obstructive Sleep Apnea (OSA) and Respiratory Insufficiency.
G2S BPAP Series User Manual • The optional SpO Kit used with the device together is intended to be used to obtain readings of SpO value as part of routine checkups. • Any serious incidents in relation to this device should be reported to the manufacturer and the competent authority in your country.
G2S BPAP Series User Manual CAUTION! • Contact your health care professional if symptoms of sleep apnea recur. Contact your health care professional if you have any questions concerning your therapy. IMPORTANT TIPS! • An irregular sleep schedule, alcohol consumption, obesity, sleeping pills, or sedatives may aggravate your symptoms.
Page 47
G2S BPAP Series User Manual Main Device offer to Communications Port 2.0 A Type of Protection against Electric Shock Class II Equipment Degree of Protection against Electric Shock Type BF Applied Part Degree of Protection against Ingress of Water IP22 Pressure Range IPAP: 4.0 ~...
Page 48
G2S BPAP Series User Manual For models with a maximum pressure of 25 hPa, the flow table is as follows. Test Pressure (hPa) Measured Pressure at the Patient Connection Port (hPa) Average Flow at the Patient Connection Port (L/min) When the working pressure is set to the values listed in the table, the average flow rate at the patient end should be greater than 80% of the corresponding flow value in the table.
G2S BPAP Series User Manual 7. Available Therapies The device delivers the following therapies: CPAP – Delivers Continuous Positive Airway Pressure; CPAP maintains a constant level of pressure throughout the breathing cycle. AutoCPAP – Delivers CPAP therapy and provides an air pressure no less than the prescribed one based on the patient’s needs.
G2S BPAP Series User Manual 8. Glossary Apnea A condition marked by the cessation of spontaneous breathing. AutoCPAP Adjust CPAP pressure automatically to improve patient comfort based on monitoring of apnea and snoring events. Auto Off When this feature is enabled, the device automatically discontinues therapy whenever the mask is removed.
Page 51
G2S BPAP Series User Manual Res Rate Respiratory Rate. Number of breaths per minute. Reslex A therapy feature that is enabled by your home care provider to provide pressure relief during exhalation. Standby State The state of the device when power is applied but the airflow is turned off.
G2S BPAP Series User Manual 9. Model Product Contents Maximum Work Model Work Mode Optional Pressure Main Device Accessories (hPa) Main device G2S B20S CPAP, S (2.4-inch TFT) Main device G2S B20A CPAP, S, AutoS (2.4-inch TFT) Main device G2S B20T CPAP, S, S/T, T (2.4-inch TFT)
G2S BPAP Series User Manual 10. Package Contents After unpacking the system, make sure you have everything shown here (Different models of the product may contain different components): Articles Qty. Notes Device Air Filter Power Adapter Power Cord Mask Optional...
G2S BPAP Series User Manual • Please contact the manufacturer for an SD card if needed. • The manufacturer cannot guarantee the normal function of the device, nor the safety and effectiveness of the device when it exceeding the expected service life.
G2S BPAP Series User Manual Name Function SD Card Slot Insert the SD card into this slot Air Outlet Deliver pressurized air; connects to the tube Connected to external equipment (only for connection of KS-CM01 Communications Kit, WL-200 Cellular Module and SG-200 SpO &GPRS Kit to the...
G2S BPAP Series User Manual 12.2 Installing the Air Filter and Filter Cap (1) Attach the air filter to the filter cap, as shown in Fig. 12-1. Air Filter Filter Cap Fig. 12-1 (2) Install the filter cap containing the air filter to the device, as shown in Fig. 12-2.
G2S BPAP Series User Manual WARNINGS! • The device is powered on for use when the power cord and power adapter are connected. Use the Knob to turn the blower On / Off. • Using the device at an AC voltage outside the specified range (see Section 6 “AC Power Consumption”) may damage the device or cause device failure.
G2S BPAP Series User Manual 12.5 Assembling the Tube and Mask (1) Connect one end of the tube to the air outlet of the main device, as shown in Fig. 12-6. Air Outlet Tube Fig. 12-6 (2) Connect the other end of the tube to the mask according to the user manual for the mask.
G2S BPAP Series User Manual WARNINGS! • Connect the oxygen tube to the oxygen inlet of the mask. • The oxygen supply must comply with the local regulations for medical oxygen. • Turn on the device before turning on the oxygen. Turn off the oxygen supply before turning off the device.
G2S BPAP Series User Manual If the SD card is inserted incorrectly, a symbol indicating incorrect insertion will appear on the main screen of the device, as shown in Fig. 12-9. Fig. 12-9 CAUTIONS! • If no SD card is inserted, neither of the symbols will appear on the main screen of the device.
G2S BPAP Series User Manual 13. Routine Use 13.1 Connecting the Tube Connect the power cord, power adapter, and tube properly in accordance with the instructions in the First Time Setup. Connect the mask and headgear according to the user manual of the mask.
G2S BPAP Series User Manual CAUTIONS! • Do not position the device where it is difficult to operate the disconnection device. • To isolate the device from the supply mains, disconnect the plug. 14. Heated Humidifier The humidifier can be obtained from your home care provider. The humidifier can reduce nasal dryness and irritation by adding moisture (and heat if applicable) to the airflow.
G2S BPAP Series User Manual Fig. 14-3 WARNING! • Change water before every use and do not surpass the maximum water level line. Otherwise, the liquid may enter the tube. CAUTIONS! • Empty the water chamber when the heated humidifier is not in use.
G2S BPAP Series User Manual • Take precautions to protect furniture from water damage. 14.2 Emptying the Water Chamber (1) Removing the water chamber according to instructions in 14.1.1. (2) Emptying the water chamber: Open the cap, as shown in Fig. 14-6, and pour any remaining water out of the water chamber.
G2S BPAP Series User Manual • The greater the difference between the temperature inside the air tubing and the room temperature, the more likely condensation will occur in the tubing. • If there is only a small amout of condensed water droplets in the tubing in the morning after treatment, the humidity level is appropriate;...
G2S BPAP Series User Manual Fig. 15-4 Note: The above interface only applies to G2S B25T, G2S B25VT, G2S B30T, G2S B30VT, G2S B30AT and G2S LAB. 15.1.2 Bringing up the Initial Setup Interface From the Main Interface shown in Fig. 15-1 or Fig. 15-2 or Fig. 15-3 or Fig. 15-4, press and hold the Ramp Button for three seconds.
G2S BPAP Series User Manual certain option, press the Knob , and the option is then displayed in yellow, meaning that the option is now ajustable, as shown by the Humidifier option in Fig. 15-7. Fig. 15-7 15.1.5 Adjusting Options Adjust the option by turning the Knob .
G2S BPAP Series User Manual Fig. 15-10 Note: are page turning symbols. 15.1.8 Exiting the Patient Menu (1) Returning to the Initial Setup Interface Move the cursor to the Back option by turning the Knob , as shown in Fig. 15-11.
G2S BPAP Series User Manual 15.2 Options in the Patient Menu and Corresponding Descriptions Option Range Description There are five humidity levels available. As the number Off / 1 ~ 5 Humidifier increases, the humidity rises accordingly. “Off” means the...
G2S BPAP Series User Manual 16. Alarm This chapter describes device the alarms of the device and proper responses the operators should make to different alarms. After the device is running, disconnect its power supply by unplugging the power cord, an audible alert will sound like “beep beep beep, beep-beep, beep beep beep, beep-beep”,...
G2S BPAP Series User Manual In accordance with the requirements of relevant standards, the volume of the audible alarm signal meets the requirements. The sound pressure range of the measured auditory alarm signal is described as follows: Measured A-weighted sound...
Page 72
G2S BPAP Series User Manual When the airflow is on, an audible alert will sound if the airway pressure exceeds the alarm limit; the screen will display “High Pressure!!!”. Note: The thresholds for different models: Off, 5 ~ 21 hPa applies to G2S B20T, in 0.5 hPa increments, the default...
Page 73
G2S BPAP Series User Manual Mask Blocked!! (only applies to G2S When the airflow is on, an audible alert B20T, G2S B25T, Middle Function will sound if the vents of the mask are B25VT, G2S Priority Alarm blocked; the screen will display “Mask B30T, G2S B30VT, Blocked!!”.
G2S BPAP Series User Manual 16.6 Reposition of Alarming After the alarm faults are cleared, the residual alarm messages still remain (alarming messages are shown on the top of the screen without any visual and auditory alarm). Turn the button leftwards or rightwards to reduce the residual alarming messages.
G2S BPAP Series User Manual 17.1 Cleaning the Mask and Headgear For details, refer to the cleaning instructions in the user manual of the mask. 17.2 Cleaning the Optional Kits For details, refer to the cleaning instructions in the user manual of the corresponding user manual.
G2S BPAP Series User Manual residue is left; then wipe it dry with a lint-free cloth, so as to prevent calcareous accumulations. • Check the water chamber for any leak or damage. Replace the water chamber if there is any damage.
BSF to ensure delivering normal treatment pressure. • If you use ozone or other cleaning and disinfection methods that are not recommended by BMC, BMC will not be able to verify the safety or performance of the equipment. 36 / 45...
18. Traveling with the Device (1) Use the BMC carrying case to carry the device and accessories along with you. Do not put them in your checked baggage. (2) This device operates on a power supply of 100 V - 240 V and 50 Hz / 60 Hz, and is suitable for use in any country in the world.
21. Technical Support Please contact BMC directly if you need the circuit diagram of the device or the list of components for certain purposes such as maintenance or connection to other equipment. BMC will provide the technical documents in whole or in part according to your needs.
G2S BPAP Series User Manual 23. Troubleshooting The table below lists common problems you may encounter with the device and possible solutions to resolve them. If none of the corrective actions solve the problem, please contact your home care provider.
Page 81
G2S BPAP Series User Manual Problem Possible Cause Solution(s) Contact your physician and home care provider. The patient is allergic to the Facial reddening Use a latex-free mask. materials of the mask Place a lining between the skin and mask Turn the humidity setting down, or raise the room temperature.
G2S BPAP Series User Manual 23.2 Common Problems in the Device and Corresponding Solutions Problem Possible Cause Solution (s) Take a few deep breaths with the The Auto On / Off feature is mask on, and the device will start...
G2S BPAP Series User Manual 24. EMC Requirements The cables must be provided by BMC. The information of each cable is as follows: Power Cord: 1800 mm ± 30 mm, unshielded. Guidance and manufacturer's declaration - electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The user of the device should ensure that it is used in such an environment.
G2S BPAP Series User Manual Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment.
Page 85
G2S BPAP Series User Manual Recommended separation distances between RF wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent...
(1) year for main device and three (3) months for all accessories from the date of sale by BMC Medical Co., Ltd. to the dealer. If the product fails to perform in accordance with the product specifications, BMC Medical Co., Ltd. will repair or replace, at its option, the defective material or part.