6 Operability
6.2
Climatic test conditions
The following results have been verified for the TarisIR® mini (M version):
Cold, passive
Test norm
Temperature
Duration
Operating mode
Cold, active
Test norm
Temperature
Duration
Operating mode
NUC interval
Dry heat, passive
Test norm
Temperature
Duration
Operating mode
Dry heat, active
Test norm
Temperature
Duration
Operating mode
NUC interval
Constant humid heat
Test norm
Temperature
Humidity
Duration
Operating mode
© InfraTec GmbH 2024
DIN ISO 9022-10-08-1
- 40°C
16 h
Unpacked, passive
DIN ISO 9022-10-05-2
See chapter 3 "Product specifications"
16 h
Unpacked, active
2 s
DIN ISO 9022-11-05-1
+ 85°C
16 h
Unpacked, passive
DIN ISO 9022-11-05-1
See chapter 3 "Product specifications"
16 h
Unpacked, active
2 s
DIN ISO 9022-12-02-1
+ 40°C
90-95% rel. humidity
96 h
Unpacked, passive
13