Page 1
No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan Tel: +886 4 2369 2388 GM700SB Emergo Europe Fax: +886 4 2261 7586 Prinsessegracht 20 Email: info@bionime.com User’s Manual 2514 AP The Hague http://www.bionime.com The Netherlands Made in Taiwan Email: EmergoVigilance@ul.com...
Page 3
The RIGHTEST GM700SB is a Bluetooth Low Energy (BLE) device. The GM700SB Blood Glucose Monitoring System is manufactured and supported by Bionime Corporation. If you have any questions or concerns, please contact your local Bionime Customer Service Representative or send an email to info@bionime.com for further assistance.
Page 4
Caution Caution - Before using the RIGHTEST Blood Glucose Monitoring System GM700SB to test your blood - The minimum blood sample size to test using the RIGHTEST Blood Glucose Monitoring glucose, please read all of the instructions and conduct all of the tests including the Quality System GM700SB is 0.75 μL: ( ) Control test (p41).
Page 5
Limitations Limitations - The RIGHTEST Blood Glucose Monitoring System GM700SB is not intended for serum or - To avoid potential electromagnetic interference, keep the meter away from electromagnetic plasma testing. radiation sources such as X-rays or MRI. - Do not use at altitudes greater than 10,000 feet (3,048 meters). - Bluetooth is enabled by default when the meter is on.
Table of Contents Table of Contents About the RIGHTEST Blood Glucose Monitoring System GM700SB View Window Appearance RIGHTEST Blood Glucose Monitoring System GM700SB Understanding Test Results and Messages RIGHTEST Blood Glucose Meter GM700SB RIGHTEST Blood Glucose Test Strip GS700 Quality Control About Quality Control Testing Before Testing Performing a Quality Control Test...
RIGHTEST Blood Glucose Monitoring System GM700SB RIGHTEST Blood Glucose Monitoring System GM700SB Your RIGHTEST Blood Glucose Monitoring System GM700SB consists of several items. Please identify each item, learn its name and how it is used. The items in RIGHTEST Blood Glucose Monitoring System GM700SB are: 1.
RIGHTEST Blood Glucose Meter GM700SB RIGHTEST Blood Glucose Meter GM700SB Test Strip Port Meter Label Display Window Main Button ( Battery Cover Meter Battery Cover Release Up Button ( Down Button ( Point...
Page 9
RIGHTEST Blood Glucose Meter GM700SB RIGHTEST Blood Glucose Meter GM700SB Indicates a test result stored in memory Indicates when to apply the blood sample Appears after you insert a test strip into the Indicates the average result meter Indicates a control solution test result Appears when an error occurs Current date under time mode or testing date Unit of test result...
RIGHTEST Blood Glucose Test Strip GS700 RIGHTEST Blood Glucose Test Strip GS700 The RIGHTEST Blood Glucose Meter GM700SB can only be used with RIGHTEST Blood Glucose Test Strips GS700 and RIGHTEST Control Solution GC700. The use of other test strips or control solutions may lead to incorrect results.
Meter Activation and Battery Change Meter Activation and Battery Change Your RIGHTEST Blood Glucose Meter GM700SB comes with 1 CR2032 3V battery installed. CAUTION 1 new battery provides sufficient power for approximately 1000 tests under normal usage. Please follow the local regulations to properly discard a used battery. Press the main button or insert a strip to activate your meter.
Setting Up Your Meter - Setting the Date, Time, Volume and Test Unit Setting Up Your Meter - Setting the Date, Time, Volume and Test Unit You can enter the Settings Mode two ways. 1. Year setting When the year blinks, press the Up or Down button to select the current 1.
Page 13
Setting Up Your Meter - Setting the Date, Time, Volume and Test Unit Setting Up Your Meter - Setting the Date, Time, Volume and Test Unit 4. 12 or 24-hour Format Setting 7. Volume Setting When the "OFF" blinks, press the Up or Down button to When the time format blinks, press the Up or Down turn the volume on or off and press the main button to button to select the preferred time format and press the...
Turning On / Off the Meter Handling the RIGHTEST Blood Glucose Test Strip GS700 1. How to turn on the RIGHTEST Blood Glucose Meter GM700SB Inserting the RIGHTEST Blood Glucose Test Strip GS700: Press the main button or insert a test strip. 1.
Handling the RIGHTEST Blood Glucose Test Strip GS700 Setting Markers Removing the RIGHTEST Blood Glucose Test Strip GS700: Meal markers add useful information to your test results. You must confirm a marker selection 1. Hold the test strip as shown. before testing.
Performing a Blood Glucose Test Performing a Blood Glucose Test 7. Rotate the clear top part of the cap to adjust the depth. Depth Adjustable Cap Release Button Check the number of lines visible in the View Window. More lines correspond to a greater depth. Try: Plunger "...
Page 17
Performing a Blood Glucose Test Performing a Blood Glucose Test 15. The screen will display a countdown timer. Your test 12. The " " icon will flash on the screen. Press the Up or Down result will be displayed after 5 seconds. buttons to select a status marker.
Performing a Blood Glucose Test Uploading Data Bluetooth is set to on by default. When the mobile app is activated, the meter will CAUTION automaticallyconnect with your mobile device if it has been paired. - Do not apply the blood sample to the strip's sample port until you see " "...
Bluetooth. More app instructions can be found by clicking "Learn and Assist" or "Operation Guide" on the app. For more information, please visit our website http://www.bionime.com . CAUTION - Blood glucose testing is not suggested during flights at any time. This is because the meter, when turning on, will emit electronic signals, which should be restricted at all times aboard aircraft.
Alternative Site Testing Alternative Site Testing Alternative site testing: palm or forearm blood CAUTION Palm sampling Consult your healthcare professional before sampling from your palm or forearm. - Before performing alternative site testing, please replace Forearm Alternate site testing should should be done only during steady-state times (when the Depth adjustable cap with the clear cap for your glucose is not changing rapidly).
View Window Appearance View Window Appearance To obtain an accurate test result, CAUTION ensure that your blood sample Check the expiration date printed on the strip vial every time you use a test strip. Do not covers the entire area of the View use expired test strips.
- If you experience symptoms that are inconsistent with your blood glucose test results Glucose Monitoring System. If you want to purchase additional supplies, please and you have followed all the test instructions in this manual, contact your healthcare contact Bionime customer service. professional immediately.
Page 23
About Quality Control Testing About Quality Control Testing When could a Quality Control Test Be Performed ? CAUTION - To confirm that your RIGHTEST Blood Glucose Meter GM700SB and RIGHTEST Blood Glucose - Each time you open a new bottle of control solution, write the opening date on the label. Test Strip GS700 are working properly.
Performing a Quality Control Test Performing a Quality Control Test 1. Take a test strip from the vial and close the cap immediately. 9. Replace the cap and ensure it is screwed on securely. Close 2. Insert the test strip into the test strip port with the View 10.
RIGHTEST Blood Glucose Meter to test your Maintaining blood glucose. Please contact Bionime customer service. Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying case when not in use.
Page 26
Maintaining the products Maintaining the products Cleaning and disinfection frequency: at least once a week. Cleaning and disinfection frequency: at least once a week. To clean the meter: To clean the lancing device: 1. Wipe the whole surface of the meter using the disinfectant wipes listed above 1.
If you have any questions or concerns, please contact your RIGHTEST Blood Glucose 1. Press the Main button to switch from Time Mode to Memory Mode. You Monitoring System GM700SB authorized distributor or contact Bionime Customer will see the "...
Page 28
Recalling Test Results Recalling Average Test Results 4. Quick Searching: The RIGHTEST Blood Glucose Meter GM700SB provides several average test results. View the To view test results in sequential order automatically, enter the Memory Mode. Press the Up 1-day, 7-days, 14-days, 30-days, 60-days and 90-days average test results for better blood or Down button for 2 seconds.
Recalling Average Test Results Error Messages and Troubleshooting Temperature Error NOTE 1. Please test between 6°C (43°F) and 44°C (111°F) in order to obtain The Average function requires that the correct time and date be entered. Test results accurate test results. must be available during the specific time interval.
Page 30
Error Messages and Troubleshooting Error Messages and Troubleshooting Strip Error Er1 - The test strip was used or moist. Please try again with a new 1. The meter will beep four times while the " " sign flashes if the test strip.
Change" on p18. Test Time 5 seconds After completing the instructions above, the meter should function normally. Please contact Bionime Customer Service, if the meter is still not working. Memory Capacity 500 blood glucose test results with date and time CAUTION Operating Temperature 6 - 44°C (43 - 111°F)
Specification Warranty Bionime Corporation warrants that this product will be free from defects in materials and Meter Dimension 82 mm x 45 mm x 15.5 mm workmanship for five years from the date of purchase. Meter Weight 43 ± 5g with battery This warranty does not apply to the performance of a RIGHTEST Blood Glucose Meter GM700SB that has been altered, misused, tampered with or abused in any way.
Expected Glucose Values Customer Service We aim to provide great service to our customers. Please review these instructions to make sure Expected glucose values without diabetes you know how to use your product correctly. If you have any questions or encounter any issues Fasting Blood Glucose with your product, please call 1-800 781 018 for local distributor Spirit Healthcare for customer GLUCOSE LEVEL...
Need help?
Do you have a question about the RIGHTEST GM700SB and is the answer not in the manual?
Questions and answers