Turning It Off; Sampling Precautions - Decagon Devices pawkit Operator's Manual

Portable water activity measurement system
Hide thumbs Also See for pawkit:
Table of Contents

Advertisement

Pawkit User Manual
4. Operation
you can either restart the measurement by pressing but-
ton I again, or you can record the shown value and end
the measurement procedure.
6. Remove the sample cup by lifting the Pawkit. Lift the
Pawkit straight up as shown to avoid spilling the sample
cup. The sample may now be discarded or covered with
a lid if it is to be re-measured at a later time.

Turning it off

To turn off the Pawkit, leave it idle for more than 5 minutes,
and it will shut off automatically. If the Pawkit has automati-
cally shut itself off, pressing button (I) will wake up the
instrument and display the last water activity measurement.

Sampling Precautions

Long exposure to a variety of volatile substances or to sam-
ples with water activities near 1.00 can shift the sensor cali-
bration. Therefore, always remove samples as soon as the
Pawkit is finished sampling (beeps) to avoid damage to the
sensor. If a sample is accidentally left in the chamber for an
extended period of time, be sure to check the calibration
when the instrument is next used.
22

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the pawkit and is the answer not in the manual?

Table of Contents