Cardiac resynchronization therapy pacemaker

Advertisement

CAUTION: After all leads have been implanted and before they are connected to the device, establish and
document the baseline morphology for capture and sensing thresholds for each lead using a suitable
recording system, such as a 12‑lead electrocardiogram (ECG) or an intracardiac electrogram (IEGM).
To connect the device to the leads:
1.
1. Remove blood and body fluids from the terminal pins of the implanted leads.
2. Check the markings on the device case and verify proper atrial and ventricular connections.
2.
CAUTION: Exercise caution when turning the setscrew, which may be backed out of the connector if
turned counterclockwise for more than two rotations.
NOTE: In CRT‑Ps: For proper sensing and pacing, it is important to ensure that left and right ventricular
signals are correctly detected and that pacing pulses are delivered in the desired chamber.
3.
3. Use the #2 torque wrench packaged with the device to retract the setscrews in the device connector so
that the pacing lead terminal pins can be fully inserted.
NOTE: The same #2 torque wrench is also supplied in a separate package, which can be used for this
purpose.
4.
4. Insert a lead firmly into the connector until the lead pin is immobile and visible in the viewport at the
opposite end of the connector.
5. Grip the torque wrench by the large part of the handle as shown in the figure on the left below. On
5.
applicable wrench models, do not grip the torque wrench by the smaller, narrower part of the handle as
shown on the right.
32

Advertisement

loading
Need help?

Need help?

Do you have a question about the Endurity Core and is the answer not in the manual?

Questions and answers

Tori Garvin
January 25, 2025

My mother has model number PM2272 cardiac pacemaker. Can she have MRI Done. She has been in hospital for 3 days. Doctors want do a MRI until they hear from manufactory. Makes no since. They have her card info.

User image 67951fc7942f7
3 comments:
HK
February 8, 2025

@Tori Garvin, of couse 1.5T and 3T

Mr. Anderson
February 10, 2025

The Abbott Endurity Core PM2272 is not listed among the MR Conditional devices in the provided context. Before performing an MRI scan on a patient with this pacemaker, it is necessary to contact Abbott Medical or consult regulatory authorities to determine if it is approved for MRI use.

This answer is automatically generated

HK
February 11, 2025

@Mr. Anderson PM2272 isn't Endurity Core, It's Assurity MRI; of couse 1.5T and 3T MRI Scan

Kristina Gardner
April 15, 2025

Abbott pacemaker PM2272, is it safe for Hyperbaric oxygen therapy? If so, at what ATA- 2 or 2.5 safe? Please email the information to me if possible. Thank you!

Aldo De Giuli
May 14, 2025

Is Abbott Endurity 2162 Pace Maker MRI compatible ?

James Sean
January 20, 2025

can batteries be replaced in your model no. PM2272 pacemaker

1 comments:
Mr. Anderson
May 14, 2025

No, the battery in the Abbott Core PM2272 pacemaker cannot be replaced separately. When the battery reaches the end of its lifespan, the entire pacemaker must be replaced.